Atec Teseq-Schaffner-INA6502 User Manual
Automatic step transformer, Ina 6502
A d v a n c e d Te s t S o l u t i o n s f o r E M C
AUTOMATIC STEP TRANSFORMER
INA 6502
L
U
IN
U
VAR
N
PE
16AT
N.C
N
To power
network
PE
Instr.
supply
Optional
remote
test + EUT
power off
EUT power
r
e
k
a
e
r
b
ti
u
c
r
i
C
f
f
o
/
n
o
To Modula
EUT
power in
Power supply
X
3
Modula peripheral controller PC 6501
Plug X1 – System cable IN
25 way sub D – male
Plug X2 – System cable OUT
25 way sub D – female
0% 40% 70% 80%
System cable to Modula system
Termination with Modula
interlock plug
The automatic step transformer type INA 6502 is a standard accessory for the Teseq Modula 6000
instrumentation series. It enables the ac power supply voltage, which is fed to the EUT, to be set to 0%,
40%, 70% or 80% of UINunder program control.
The INA 6502 complies with the requirements of IEC/EN 61000-4-11(2004) for voltage dips and under-
voltages. A large number of product standard specifi cations can be fully met as they do not ask for
continuous and arbitrary voltage variations.
The device is connected to the Modula system via the system cable and is fully embedded in the
confi guration management concept, the software control and the report mechanism.
The voltage steps, UVAR, are programmed as a percentage of the power line voltage using either the
mobile handheld controller or by means of WinModula. The updated version of WinModula includes
preprogrammed tests using the INA 6502 for the basic standard, the generic standards and a consider-
able number of product standards.
A circuit breaker is incorporated into the housing of the INA 6502. This allows the on/off switching of the
power to the EUT at any moment via the user interface or automatically under program control. A remote
EUT-off hardware switch to cut power and high voltage pulse generation is available as an option.
The step transformer is designed into a sturdy aluminum housing fi tted with convenient handles. In
applications the INA 6502 can be used in any position and may even be placed under the test bench if
the space on the set-up becomes limited.
Block diagram
Integrated function unit to
Modula system
Power line dips to 0%, 40%, 70%
and 80%
Complies with IEC/EN 61000-4-11
Includes EUT power control
switch