Automated 3-phase coupling/decoupling network, Cdn 3063 – Atec Teseq-Schaffner-CDN3063 User Manual
Page 2
A d v a n c e d Te s t S o l u t i o n s f o r E M C
AUTOMATED 3-PHASE
COUPLING/DECOUPLING NETWORK
CDN 3063
Parameter
Value
CDN 3063-B32:
Automated 3-phase coupling network for burst (EFT)
CDN 3063-S32:
Automated 3-phase coupling network for surge (combination
wave, ring wave)
CDN 3063-C32:
Automated 3-phase coupling network for burst (EFT) and surge
(combination wave, ring wave)
Dimensions CDN 3063-C32:
W: 449 mm (17.7”), H: 310.5 mm (12.25”, 7 HU), D: 565 mm (22.2”)
Weight CDN 3063-C32:
43 kg (94.8 lb) approx.
Technical data coupling network
Dimensions/Weight
Electrical parameter
Parameter
Value
Standards-compliant pulses:
Electric fast transient EFT (burst) up to 4.8 kV
Combination wave up to 6.6 kV
Ring wave up to 6.6. kV
Decoupling attenuation:
Residual pulse voltage on EUT power supply inputs; conforms to
IEC/EN 61000-4-5
Residual voltage on non-pulsed power supply inputs; conforms to
IEC/EN 61000-4-5
Instrument supply:
85 – 265 V
AC
AC/DC EUT current range:
3 x 32 A continuous (over temperature protection)
3 x 40 A for approx. 35 min (over temparature protection)
3 x 50 A for approx. 10 min (over temperature protection)
EUT voltage range:
3-phase (5 wire) usable for 2-phase (4 wire) and 1-phase (3 wire)
EUT V
AC
30* to 480 V
AC
rms, line to line
EUT V
AC
30* to 280 V
AC
rms, line to neutral, 50/60 Hz
EUT V
DC
0 to 480 V
DC
DC - 65 Hz with no EUT power loss, 400 Hz max. with EUT power
loss *(below 30 V synchronization not guaranteed, asynchronous
mode only)
Connections front panel:
HV-surge pulse input from generator
(Fischer connector)
EFT connector from generator
(SHV connector)
Ground/PE connection
Connections rear panel:
Connector for EUT supply, Phoenix screw terminals, phase rotation
detection