Atec Teseq-CDN_3061 User Manual
Automated 1-phase coupling/decoupling network, Cdn 3061
A d v a n c e d Te s t S o l u t i o n s f o r E M C
The new, modular CDN 3061 is an easily upgradeable test instruments that maximize the user’s
initial investment. Users can select a CDN confi guration that fulfi lls their basic testing needs with the
assurance that they can upgrade it to a version to fi t different generator compatibility as their testing
requirements change. The CDN 3061 is designed for maximum reliability in a wide range of test setups.
Over temperature protection, which allows short term operation at currents exceeding the nominal
rating, prevents damage to internal components.
The automated single-phase coupling network incorporates surge and burst coupling for continu-
ous EUT currents of up to 16 A and complies to the requirements for the surge standard IEC/EN 61000-4-
5:2005, as well the coupling modes given in the ANSI C62.45 standard and for burst standard IEC/EN 61000-
4-4:2005. This network features a module for the power quality test required by IEC/EN 61000-4-11 and
IEC/EN 61000-4-29. Automated variac accessories are available for connection to the CDN to perform
the variation test described in IEC/EN 61000-4-11. Voltage variation test parameters are set via the
generator’s user interface.
The CDN 3061 utilizes the latest electronic component technology which guaranties remarkable phase
coupling accuracy which exceeds the existing standard’s requirements, and represents a signifi cant step
forward in higher test results reproducibility.
Single-phase model with exceed-
ing current range
Fully automated Surge and Burst
coupling
IEC and ANSI coupling
High accuracy switching tech-
nology
Modular and upgradeable to fit
new generator architecture
AUTOMATED 1-PHASE
COUPLING/DECOUPLING NETWORK
CDN 3061
Parameter Value
Dimensions CDN 3061-C16:
W: 449 mm (17.7”), H: 221.5 mm (8.75”, 5 HU), D: 565 mm (22.2”)
Weight CDN 3061-C16:
20 kg (44 lb) approx.
Coupling network specifications
Dimensions/weight
Electrical parameter
Parameter
Value
Instrument supply:
85 – 265 V
AC
Decoupling attenuation:
Residual pulse voltage on EUT power supply inputs 15% max.
Residual voltage on non-pulsed power supply inputs 15% max.
Standard-compliant pulses
Electric fast transient EFT (burst)
Combination wave
Power quality test (CDN 3061 only)
Surge decoupling inductance:
1.5 mH
Connections: Front panel
HV-surge pulse input from generator
(Fischer connector)
EFT connector from generator
(SHV connector)
Ground/PE connection
Rear panel
Cable connector for EUT supply input and output
EUT supply: 1-phase (P/N/PE)
EUT V
AC
30 to 270 V
AC
rms, line to neutral, 50/60 Hz (below 30 V
synchronization not guaranteed, asynchronous mode only)
EUT V
DC
0 to 270 V
DC
DC - 65 Hz with no EUT loss, 400 Hz max. with EUT power loss
EUT current:
1 x 16 A rms continuous (over temperature protection)
1 x 25 A rms for 30 min (over temparature protection)
© February 2011 Teseq
®
Specifications subject to change without notice.
Teseq
®
is an ISO-registered company. Its products
are designed and manufactured under the strict
quality and environmental requirements of the ISO
9001. This document has been carefully checked.
However, Teseq
®
does not assume any liability for
errors or inaccuracies.
Teseq AG
Nordstrasse 11F 4542 Luterbach Switzerland
T + 41 32 681 40 40 F + 41 32 681 40 48
[email protected]
www.teseq.com
691-260B February 2011