TM

Technical Specifications

KeyTek ECAT

®

Model E103 Series Control Center

Computer-driven control center and power units required for the opera-
tion of any KeyTek ECAT test system

Virtual Front Panel

control

8 x 40 character keyboard entry & LCD display - allows operator to see
vital test parameters without list scrolling

FiberCom

fiber-optic

Uses KeyTek ECAT software for full computer control of ECAT test equip-

interface & control system

ment (user-supplied PC running Windows, 8 MB RAM, one serial port)

Module Bay

For one full-width plug-in module or two half-width modules. Up to five
additional bays (for a total of six) and/or S-ECAT for floor-standing
system available (required for more than three docking bays)

Surge V & I monitor ports

For waveform monitoring with an external (user-supplied) oscilloscope at
1kV/V and 200A/V;V & I signals supplied from optional AC coupler/
decoupler or KeyTek ECAT modules/options

Input voltage

100V to 240VAC, 50/60 Hz

Typical input current

3.5A @ 100V; 1.5A @ 240V

Operating temperature

15° to 35° C

Operating humidity

10% to 75% non-condensing

Options

E103-S - adds oscilloscope trigger for installed surge modules

AC input connectors are available for most national and international standards.

Technical Specifications

KeyTek ECAT

®

Expert Computer-Aided Testing
for pulsed EMI immunity

The KeyTek ECAT

®

is a modular, full capability

EMC test system for measuring and analyzing the
vulnerability of telecom, electronic, and electrical
equipment/ components to pulsed EMI hazards,
including EFT, Surge & PQF

(Power Quality Failure).

A powerful design and production tool, the KeyTek
ECAT is the premier EMC test system to meet
Telcordia, UL, FCC, and IEC standards. It features
a totally integrated modular architecture that
enables manufacturers and designers to easily
and rapidly test for all pulsed EMI threats
including pre-compliance, production sampling
and final compliance.

It’s the only EMC test system you’ll ever need.

Tests for pulsed EMI hazards:
EFT, Surge, PQF

Ideal test system to address
most applicable EMC &
Telecom standards, including
CE Mark/IEC standards

Easy-to-use Windows-based
application software for quick
implementation of international
& national test routines

Virtual Front Panel

retains key

operating parameters during
set up & testing

Multi-level system interlock
architecture provides
maximum safety

Single output port/instant
mode switching

AC Mains current monitoring

Accurate automatic report
generation

Flexible, economically
upgradable architecture