MODEL 9354-1 & 9354-2 TRANSIENT GENERATORS
for susceptibility tests MIL-STD461C/D/E and other specifications
APPLICATION
The Model 9354-1 & 9354-2 Transient
Generators were especially designed for the per-
formance of a variety of pulse susceptibility tests
on sub-systems and/or equipment, in accordance
with MIL-STD-461D and E, method CS116; also
capable of testing for RTCA DO160D, section 22;
MIL-STD-461C, methods CS10 and CS11.
Through the use of many Solar accessories,
including various reactive networks and coupling
devices, as well as other commercially available
items such as loop antennas, parallel plates,
and TEM cells, the generated output may be
modified and applied to other specifications.
(Contact Solar customer service for details.)
DESCRIPTION
The Model 9354-1 & 9354-2 Transient
Generators provide nine selectable waveforms,
including six damped sinusoidal pulses (10 KHz,
100 KHz, 1 MHz, 10 MHz, 30 MHz, and 100 MHz)
and three double exponential pulses (6.4
ȖS, 70
ȖS, 500 ȖS on 9354-1 and 120 ȖS on 9354-2).
Auto pulsing of the sinusoidal repetition rate
is internally adjusted from 0.5 to 1.0 pulse per
second. A front panel-mounted push button can
be used to manually trigger single pulses. The
peak amplitude of the selected output pulse is
adjustable as a percentage of the charge voltage.
The six damped sinusoidal waveforms were
designed to meet the requirements of
MIL-STD-461D and E, method CS116, when
applied in accordance with the test method of
MIL-STD-462D. Continuous tunable frequencies
can be obtained by the use of the optional
variable frequency modules.
These same waveforms are applicable to the
requirements of MIL-STD-461C, methods CS10
and CS11, when applied in accordance with the
test method of MIL-STD-462, Notice 5.
The 1MHz and 10 MHz damped sinusoidal
waveforms have been extended to a peak open
circuit voltage of 3200 volts and a calculated
short circuit current of 128 amperes to meet
the requirements of RTCA DO-160E, Section 22,
Table 22-2 and 22-3, waveform 3.
The three double exponential pulses were
designed to meet the requirements of RTCA
DO-160E, Section 22, Tables 22-2 and 22-3.
Table 1 lists the test level that can be achieved
from the Model 9354-1 & 9354-2.
FEATURES
I
Panel-mounted digital voltmeter. Monitors the
adjusted open circuit discharge voltage.
I
Pulse rates up to one pulse per second
maximum (factory adjusted).
I
Single pulse feature enables controlled
isolation of transient effects.
I
Output voltage adjustable from 0.1% to 100%
of selected discharge voltage.
Table 1: DO-160E Test Levels possible
from Model 9354-1 & 9354-2
Waveform
Pin Injection
Cable Bundle
Injection
1 (70
ȖS)
no requirement
level 1- 4
2 (6.4
ȖS)
no requirement
level 1 - 4
3 (1MHz)
level 1 - 5
level 1 - 4
3 (10MHz)
no requirement
level 1 - 4
4 (70
ȖS)
level 1 - 5
level 1 - 4
5B (500
ȖS)
level 1 - 4
level 1 - 3
5A (120
ȖS)
level 1 - 4
level 1 - 3
9