Radiated immunity test system 6 ghz, Its 6006, Amp in amp out amp in amp out – Atec Teseq-ITS-6006 User Manual
Page 5: Power, Amp in amp out amp in amp out amp in amp out
5
82-254005 E02 March 2011
5
Radiated immunity test system 6 GHz
its 6006
Application with 2 power amplifiers and 1 antenna
Software
Amp in
Amp out
Amp in
Amp out
EUT monitoring: analog, digital, optical to ITS 6006
Power amplifier 1 and 2
ITS 6006
Switch (part of ITS 6006)
EUT
1
B
A
ITS 6006
Power
RADIATED IMMUNITY TEST SYSTEM 80 MHz - 6 GHz
Power Meter
Generator
4
3
2
1
Interlock
Switch
1
2
A
B
A
B
Analog
0..24 V
Digital 2
0..24 V
Optical
User Port
Monitoring
Digital 1
0..24 V
4
3
2
1
RF on
Generator out 1 to 4
PM channel 1 to 3
2x Switch 1/2
EUT monitoring
DIRECTIONAL COUPLER
Power Meter
DIRECTIONAL COUPLER
Power Meter
Application with 3 power amplifiers and 1 antenna
Software
Amp in
Amp out
Amp in
Amp out
Amp in
Amp out
EUT monitoring: analog, digital, optical to ITS 6006
Power amplifier 1 to 3
ITS 6006
Switch (part of ITS 6006)
EUT
A
2
1
B
B
A
ITS 6006
Power
RADIATED IMMUNITY TEST SYSTEM 80 MHz - 6 GHz
Power Meter
Generator
4
3
2
1
Interlock
Switch
1
2
A
B
A
B
Analog
0..24 V
Digital 2
0..24 V
Optical
User Port
Monitoring
Digital 1
0..24 V
4
3
2
1
RF on
Generator out 1 to 4
PM channel 1 to 3
2x Switch 1/2
EUT monitoring
DIRECTIONAL COUPLER
Power Meter
DIRECTIONAL COUPLER
Power Meter
DIRECTIONAL COUPLER
Power Meter
Software
The included software allows configuration and application of the ITS 6006 as an signal generator with many additional features like scalar quadripole
measurements for measuring, for example, the insertion loss of the RF cable. For EMC testing, the ITS 6006 can be used in combination with Teseq’s