Main Output

The main output on the unit has two taps, allowing the selection
of output voltages up to 16V and output currents up to 750A.
The unit operates at slightly reduced ratings when operating from
a 115V supply.

750ADM

750ADM-H

115V

230V

115V

230V

Open circuit voltage

2.5V

3.5V

3.5V

5.0V

Voltage at 500A

1.9V

2.8V

2.8V

4V

Continuous current

125A

125A

125A

125A

5 min on

250A

250A

250A

250A

1 min on

500A

500A

440A

500A

Max current

750A

750A

500A

750A

Max current on time

10s

30s

10s

20s

Open circuit voltage

10V

16V

10V 16V

Full load voltage

7.5V

10V

7.5V 10V

Continuous current

10A

10A

10A 10A

1 min on

40A

40A

40A 40A

Metering

The output is metered by a digital true RMS system with a
memory ammeter - whenever the output is switched off, the
current reading is held on the display.

Range

Resolution

Trip current Accuracy

Capture time

20.00A

0.01A

21A

±0.5%rdg+5d

20ms

50.00A

0.01A

53A

±0.5%rdg+5d

20ms

200.0A

0.1A

210A

±0.5%rdg+5d

20ms

750A

1A

788A

±0.5%rdg+2d

20ms

A current trip is automatically set to 105% of full scale of the
selected metering range to protect the device under test.

Timing System

Range

0-999.999s/9999.99s/99999.9s auto-ranging

Resolution

1ms/10ms/100ms

Accuracy

±0.01%rdg+2d (all modes except current operated)

±0.01%rdg+4d (current operated mode)

The contact circuits have an open circuit voltage of 24Vdc and a
short circuit current of 20mA. Each contact circuit will auto-
select for normally open or normally closed contacts. A DC
voltage of 24-240Vdc may also be used to trigger either timer
channel.

The following functions are provided:

Mode

Timer Start

Timer Stop

Off

Timer inactive

Timer inactive

Internal start

Press ‘ON’

Contact 1

Single contact

Contact 1

Contact 1

Dual contact

Contact 1

Contact 2

Current operated

Current > 20%

Current < 20%

of range

of range

The output is automatically switched off at the end of the test to
safeguard the relay under test.

RS232

An RS232 port is provided to allow connection to a PC or a
printer. ADMlog software is available as an optional extra to log
results from the 750ADM and transfer them to a spreadsheet.

T&R Link

The T&R Link allows a T&R DVS3 mk2 voltage source to phase
lock to the 750ADM current.

Supply Requirements

Auto-selecting

750ADM

750ADM-H

115V±10% 50/60Hz 1ph

2000VA max

1900VA max

230V±10% 50/60Hz 1ph

3000VA max

3900VA max

Temperature Range

Storage

-20°C to 60°C

Operating

0°C to 45°C

Dimensions

Weight

380mm x 314mm x 221mm 21.5kg (750ADM) 23.5kg (750ADM-H)

Accessories

Operating manual, mains lead, spare fuse set and carrying strap.

Optional 750ADM-AL Lead Set specifications

A range of output leads are available to complement the
750ADM. The standard 750ADM-AL lead set is 3m long,
recommended for use with a 230V Supply. A 1.5m lead set is
also available, and is recommended when operating from a 115V
supply. The leads consist of double insulated 95mm

2

welding

cable terminated in Dinse high

current connectors

at the 750ADM end and
high current welding
clamps at the load
end.

Low current timer
leads are also included
with the lead set:

The 3m lead set weighs 9.8kg including high current leads, timer
leads and case.

Protection and Safety

The unit is protected by electronic over current and thermal trips
on the outputs, and circuit breakers on the input and power
circuit. An earth terminal is provided for connection to a local
earth when testing in a substation environment. The unit is

designed to comply with BSEN61010, and is CE

marked.

750ADM mk2/750ADM-H mk2 Specification

750ADM mk2 Data sheet rev 0 16/09/09

75

0A

O

ut

pu

t

40

A

O

ut

pu

t

26

The standard 750ADM is supplied
in an aluminium case. It is also
optionally available mounted in a
Pelicase with space for leads.