Atec T-R-200A-3PH-ED User Manual
Page 2
Note: Due to the company’s continuous research programme, the information above may change at any time without prior notificatio n. Please check that you have the most recent data on the product.
T&R Test Equipment Ltd, 15-16 Woodbridge Meadows, Guildford, Surrey, GU1 1BJ, UK
Tel: +44 (0)1483 207428 Fax: +44 (0)1483 511229 email: [email protected]
www.trtest.com
Main Output
The units have three independently controlled outputs, one for
each phase. The output is switched on and off by illuminated
pushbuttons, and the current is set using the three control knobs
on the front panel. All ranges are continuously variable from
zero.
Range Continuous 5 minutes 1 minute
5V 50A 100A 200A
The above intermittent on times must be followed by an off time
of 15 minutes, and are based on an ambient temperature of
25°C.
Auxiliary AC Outputs
A switched, isolated auxiliary 240V 50mA AC supply is available
to power auxiliary control circuits. The output is protected by a
fuse.
Metering
The output on the 200A-3PH/ED is metered by a digital true
RMS system with a 3.5 digit LED display. The range for each
channel is independently selected for each channel by a switch
on the front panel.
Range Resolution Trip Current Accuracy
10A 0.01A 11.5A ±0.6% rdg+6d
20A 0.01A 23A ±0.6% rdg+6d
100A 0.1A 115A ±0.6% rdg+6d
200A 0.1A 230A ±0.6% rdg+6d
An electronic current trip is automatically set to 115% of full scale
of the selected metering range to protect the device under test.
Supply Requirements
400V+14%-6% 50/60Hz 3ph 4 wire 3300VA max
A delta-star transformer is available as an optional accessory for
operation from a 3 wire supply.
Temperature Range
Storage -20°C to 60°C
Operating 0°C to 45°C
Dimensions Weight
200A-3PH/ED 315mm x 550mm x 300mm 35kg
Accessories Supplied With Unit
Operating manual, mains lead, spare fuse set, 5m 200AL/3PH
output lead set.
Lead Set specifications
The 200AL/3PH lead set includes the following items:
6 x 5m 25mm
2
200A leads terminated in M10 fork crimps
1 x 5m 2 core auxiliary lead terminated in 4mm plugs
1 x 5m 2 core auxiliary lead terminated bare ends
The lead set is supplied in a reinforced plastic wallet.
Optional Accessories
200A-3PH delta-star transformer.
Timing System
The timing system on the 200A-3PH/ED is very flexible, allowing
a wide range of protective devices to be timed.
Range 0-999.999s
Resolution 1ms
Accuracy 0.01% rdg +1d
(all except current operated mode)
0.01% rdg +3d (
current operated mode)
The contact circuit has an open circuit voltage of 24Vdc and a
short circuit current of 300mA. The contacts may be connected
to volt-free contacts or contacts connected to a DC voltage
system (220Vdc maximum).
The following functions are provided:
Mode Timer Start Timer Stop
Normally Open Press ‘ON’ Contact closes
Normally Closed Press ‘ON’ Contact opens
Current operated Current > 20% Current < 20%
of range of range
Off Timer inoperative Timer inoperative
The output is automatically switched off at the end of the test to
safeguard the relay under test.
Protection and Safety
The 200A-3PH/ED is protected by electronic over current trips on
the outputs, fuses on the contact input, the mains input and
regulators.
The auxiliary AC output is fuse protected. An earth terminal is
provided for connection to a local earth.
The unit is designed to comply with BSEN61010, and is CE
marked.
Optional Delta-Star Supply Transformer
The optional delta-star supply transformer allows operation from
115V, 230V, 400V, and 440V 3 wire (delta) supplies. The input
voltage is selected by a switch on the unit. An auxiliary single
phase output is also provided to supply power to a DVS3 voltage
source.
Input: 115V, 230V, 400V, 440V ±10%
3 wire, 4000VA max
3
φ
output: 400V 4 wire 3300VA 1 min on/15 min off
1100VA continuous
1
φ
output: 230V 300VA 5 min on/15 min off
150VA continuous
200A-3PH/ED Specification
200A-3PHED Data sheet rev 4 15/05/06