Main Output

The units have three independently controlled outputs, one for

each phase. The output is switched on and off by illuminated

pushbuttons, and the current is set using the three control knobs

on the front panel. All ranges are continuously variable from

zero.

Range Continuous 5 minutes 1 minute

5V 50A 100A 200A

The above intermittent on times must be followed by an off time

of 15 minutes, and are based on an ambient temperature of

25°C.

Auxiliary AC Outputs

A switched, isolated auxiliary 240V 50mA AC supply is available

to power auxiliary control circuits. The output is protected by a

fuse.

Metering

The output on the 200A-3PH/ED is metered by a digital true

RMS system with a 3.5 digit LED display. The range for each

channel is independently selected for each channel by a switch

on the front panel.

Range Resolution Trip Current Accuracy

10A 0.01A 11.5A ±0.6% rdg+6d

20A 0.01A 23A ±0.6% rdg+6d

100A 0.1A 115A ±0.6% rdg+6d

200A 0.1A 230A ±0.6% rdg+6d

An electronic current trip is automatically set to 115% of full scale

of the selected metering range to protect the device under test.

Supply Requirements

400V+14%-6% 50/60Hz 3ph 4 wire 3300VA max

A delta-star transformer is available as an optional accessory for

operation from a 3 wire supply.

Temperature Range

Storage -20°C to 60°C

Operating 0°C to 45°C

Dimensions Weight

200A-3PH/ED 315mm x 550mm x 300mm 35kg

Accessories Supplied With Unit

Operating manual, mains lead, spare fuse set, 5m 200AL/3PH

output lead set.

Lead Set specifications

The 200AL/3PH lead set includes the following items:

6 x 5m 25mm

2

200A leads terminated in M10 fork crimps

1 x 5m 2 core auxiliary lead terminated in 4mm plugs

1 x 5m 2 core auxiliary lead terminated bare ends

The lead set is supplied in a reinforced plastic wallet.

Optional Accessories

200A-3PH delta-star transformer.

Timing System

The timing system on the 200A-3PH/ED is very flexible, allowing

a wide range of protective devices to be timed.

Range 0-999.999s

Resolution 1ms

Accuracy 0.01% rdg +1d

(all except current operated mode)

0.01% rdg +3d (

current operated mode)

The contact circuit has an open circuit voltage of 24Vdc and a

short circuit current of 300mA. The contacts may be connected

to volt-free contacts or contacts connected to a DC voltage

system (220Vdc maximum).

The following functions are provided:

Mode Timer Start Timer Stop

Normally Open Press ‘ON’ Contact closes

Normally Closed Press ‘ON’ Contact opens

Current operated Current > 20% Current < 20%

of range of range

Off Timer inoperative Timer inoperative

The output is automatically switched off at the end of the test to

safeguard the relay under test.

Protection and Safety

The 200A-3PH/ED is protected by electronic over current trips on

the outputs, fuses on the contact input, the mains input and

regulators.

The auxiliary AC output is fuse protected. An earth terminal is

provided for connection to a local earth.

The unit is designed to comply with BSEN61010, and is CE

marked.

Optional Delta-Star Supply Transformer

The optional delta-star supply transformer allows operation from

115V, 230V, 400V, and 440V 3 wire (delta) supplies. The input

voltage is selected by a switch on the unit. An auxiliary single

phase output is also provided to supply power to a DVS3 voltage

source.

Input: 115V, 230V, 400V, 440V ±10%
3 wire, 4000VA max

3

φ

output: 400V 4 wire 3300VA 1 min on/15 min off

1100VA continuous

1

φ

output: 230V 300VA 5 min on/15 min off

150VA continuous

200A-3PH/ED Specification