Main Output
The main output on the unit has four taps, allowing the selection of
output voltages up to 240V and output currents up to 100A.
*Off time of 15 minutes. On times based on an ambient temperature of 25°C.
**6 second intermittent ratings available with 230V supply.
I Limit Mode
The main output has a current limit mode that gives very fine control
of output currents up to 10A. It also allows fine current control into
very low impedance loads such as digital relays.
Auxiliary AC and DC Outputs
A switched, isolated auxiliary DC supply with current limit protection
is available to power the relay under test, and a 110Vac auxiliary
output is available for tests.
DC stabilised output
24V, 48V, 60V
1.0A
110V, 220V
0.23A
Fixed AC output
110Vac
300mA
Metering
The output is metered by a digital true RMS system with a single
cycle capture memory ammeter—whenever the timer stops and the
output is switched off, the reading is held on the display. A current
trip is set to 110% of full scale of the selected metering range.
Supply Requirements
115/230V±10% auto-selecting 50/60Hz 1ph 1200VA max
Accessories
Operating manual, output lead set, mains lead, spare fuses and
carry strap.
Timing System
Range
0-999.999s/9999.99s/99999.9s autoranging
Resolution
1/10/100ms
Accuracy
0.01%rdg+2d (+4d current operated mode)
Contact o/c 24V
Contact s/c
20mA
Vdc
24—240V
Each contact circuit will auto-select for normally open or normally
closed contacts. A DC voltage of 24—240VDC may also be used to
trigger either timer channel. Contact state is shown by an LED. The
output automatically switches off at the end of the test to safeguard
the relay under test.
Pulse mode is used for setting the current level in devices sensitive
to heating. Current is injected for 200ms.
RS232 and T&R Link
An RS232 port is provided to allow connection to a PC or a printer.
The T&R Link output allows a T&R DVS3 mk2 voltage source to
phase lock to the 100ADM mk4 current..
Protection and Safety
The unit is designed to comply with BSEN61010 and is CE marked.
An earth terminal is provided for connection to a local earth for
testing in sub-station environments.
Dimensions
Weight
380mm x 314mm x 221mm
20.0kg
Including handle and corner protectors
Temperature Range
Storage
-20°C to 60°C
Operating
0°C to 45°C
Lead Set specifications
The 100ADM mk4 is supplied with a lead set consisting of:
2 x 5m 25mm
2
100A leads terminated in M10 ring crimps
2 x 5m, 2 x 0.5m 2.5mm
2
25A leads terminated in 4mm plugs
1 x 5m 2 core 0.75mm
2
lead terminated in 4mm plugs
The lead set is supplied in a plastic carrying case.
Lead set weight 5.3kg including case.
Optional Accessories
Filter unit, RB10 resistor box, Printer, pushbutton lead for runback
timing on disc induction relays.
The standard 100ADM mk4 is
supplied in an aluminium case. It is
also optionally available mounted
in a Pelicase with space for leads.
100ADM mk4 Specification
100ADM mk4 Data Sheet Rev 1 03/10/12
8
Range
Current
Output
@230V
Cont
5 min
*
1 min
*
O/C
Load V
10V
33A
67A
100A
10.5V
8.7V
@
100A
35V
10A
20A
30A
36V
32V
@30A
100V
3A
6A
10A
108V
99V
@10A
240V
1A
2A
3A
276V
259V
@3A
240Vdc
1A
2A
3A
Range
Current (A)
Output V
@230V
Short
circuit
Cont.
5 min
2 min
O/C
Load V
10V
10A
3A
6A
10A
8.6V
5V
@5A
35V
3A
1A
2A
3A
29V
13V
@2A
100V
1A
0.3A
0.6A
1A
88V
40V
@0.6A
240V
0.3A
0.1A
0.2A
0.3A
224V 130V
@0.2A
Range
Resolution
Trip
current
Accuracy
Acquisition
time
2.000A
0.001A
2.2A
±0.5%rdg±5d
20ms
10.00A
0.01A
11A
±0.5%rdg±5d
20ms
20.00A
0.01A
22A
±0.5%rdg±5d
20ms
100.0A
0.1A
110A
±0.5%rdg±5d
20ms
Mode
Timer Start
Timer Stop
Internal start
Press ‘ON’
C1 or C2 change
1 contact
C1 1st change
C1 1 2nd change
2 contacts
C1 change
C2 change
Current operated
Current > 10% of
Current < 10% of
Pulse
Press ‘ON’
200ms