The V60 Series rear panel layout includes three standard remote interfaces: A fully
programmable RS232 port, a PLC Interface port and a High Voltage Scanner control port.
Available options include GPIB and rear terminal outputs.
outputs of Testing, Pass, Fail and inputs of Start,
Reset. Output isolation is 400V with 130mA
continuous load current
Scanner Interface: Control port for V16 and
V17 HV Scanners
Timer Accuracy: 0.25% ± 50mS
Power Source: AC 100V, 120V, 220V, 230V at
± 10% 50/60Hz
Operating Environment: Indoor use, Altitude
up to 6500ft (2000m), Installation Category II,
Pollution Degree 2
Operating Temperature: 0ºC to 40ºC, 80% RH
max
Storage Temperature: -10ºC to 70ºC, 70% RH
max
Dimensions: 17.6”L x 13”W x 5.9”H
446L x 330W x 149H mm
Weight: 30 lbs (14kg) net, 34 lbs (16kg) ship
Accessories: NIST certification, TL-20-60
Alligator Test Leads, serial interface cable,
instruction manual and power cord
Accuracy: Stated specifications apply for 1
year at 23ºC ± 5ºC
Calibration: ANSI/NCSL Z540 Certificate of
Calibration, with data, traceable to NIST
provided
Compliance: CE mark certified to EN61010
and EMC directive EN50081-1 & EN50082-2
Warranty: 3-year extended warranty with
registration and annual factory calibration, 1 year
parts and labor standard
Represented By:
QuickTest™ Software
Option QT-60: Test automation software. Easy to
load graphical test sequence development and
operating environment. Provides time/date,
operator, test parameters, test results and readings
for review, printout or export to database program
Ordering Information
V60 AC Electrical Safety Tester
V61 AC/DC Electrical Safety Tester
V63 AC/DC/IR Electrical Safety Tester
QT-60 QuickTest™ Software (Serial and GPIB)
GP-V60 GPIB Interface
GP-1 1 Meter GPIB Cable
GP-2 2 Meter GPIB Cable
RPO-60 Rear Panel Output
TL-115-60 Nema 5-15 Test Adaptor
TL-10-60 High Voltage Probe Test Lead Set
TL-20-60 High Voltage Alligator Lead Set
TL-TP-60 HV Test Pistol Probe Set
TL-IEC-60 IEC Socket Lead Set
V16 16 Channel HV Scanner
V17 8 Channel HV/8 Ch. High Current Scanner
OM-60 Additional Operator’s Manual
Vitrek Corporation
9880A Via Pasar
San Diego, CA 92126
V: 858-689-2755
F: 858-689-2760
E-mail:
Website:
www.vitrek.com
AC Hipot
Voltage Range: 0.100kV to 5.000kV
Voltage Setting: 5 Volt Increments
Voltage Regulation: 1% + 5V
(Line & Load)
Voltage Accuracy: 1.5% of reading + 10V
Current Sourcing:
30 to 40mA (above 500V, 180sec. max.)
0 to 29.99mA (above 500V, continuous)
0 to 10mA (below 500V, continuous)
Current Limit Range: 0.10 to 40mA,
0.02mA/step
Current Accuracy: 1.5% of reading + 50µA
DC Hipot (V61, V63 only)
Voltage Range: 0.100kV to 6.000kV
Voltage Setting: 5 Volt Increments
Voltage Regulation: 1% + 5V
(Line & Load)
Voltage Accuracy: 1.5% of reading + 10V
Current Sourcing:
0 to 10mA (above 500V, continuous)
0 to 2mA (below 500V, continuous)
Current Limit Range: 0.10 to 10mA,
0.01mA/step
Current Accuracy: 1.5% of reading + 50µA
Insulation Resistance (V63 only)
DC Output Voltage: 50V / 100V 500V / 1000V
Resistance Accuracy 50V / 100V:
1 to 50M : 5% of reading
51 to 200M : 10% of reading
201 to 1990M : 20% of reading
Resistance Accuracy 500V / 1000V:
1 to 500M : 5% of reading
501 to 2000M : 10% of reading
2001 to 9990M : 20% of reading
Continuity Measurement
Test Current: 100mA
Resistance Accuracy: ±0.075
Resistance Range: 0.001 to 1.500
General
Test Database: 10 Test Sequences with up to
16 steps per test sequence
Front Panel Lockout: Prevents database edit
while in lockout mode
ARC Detect: 40 level (1 to 40mA)
Serial Port: 9600 Baud, fully bi-directional
RS232 interface with PC cable included
GPIB: Option GP-V60, IEEE-488 interface
PLC Interface: 9-pin D-sub connector with
WORLD LEADERS IN COMPLIANCE AUTOMATION
SM
Prices and specifications subject to change without notice.
©Copyright 2004. Vitrek Corporation, all rights reserved
.