COMPLETE AUTOMATIC TEST PROCEDURE

The 2293 performs resistance measurements on all

windings without any reconnection. The test procedure is

simple and efficient. One end of the measurement cable set

is connected to each bushing of the transformer using the

special Kelvin clamp, while the other end is connected to

the 2293.

Once the cable set is connected to the test object, the

instrument will automatically measure selected resistances

of both windings and all phases. In addition, the motorized

tap changer can be controlled to perform fully automated

measurements on all taps. A complete transformer can be

tested by a single person in a fraction of the time compared

to traditional instruments.

GRAPHICAL INTERFACE

The 7” touch screen full graphical interface guides the

operator through the test procedure.

Select the test object by touching the appropriate icon and

press start – The unit then visualizes each test cycle and

displays the results graphically or in list format.

DEMAGNETIZATION FUNCTION

Applying a DC current to an inductive test object, like a

power transformer, magnetizes the core. The resulting

magnetic remanence will have an adverse effect on other

measurements.

The 2293 includes a fully automatic demagnetization

feature which eliminates the magnetic remanence. Select

the test object by touching the appropriate icon and press

start – The unit visualizes the whole demagnetization cycle

and performs the correct core demagnetization.

HEAT RUN TEST

The 2293, which can measure HV and LV side resistances

simultaneously and accurately, is the perfect tool for
resistance measurements during a heat run test.
The instrument provides efficient and accurate acquisition

of the required data points to allow drawing
the necessary cooling curve.

Results are available as CSV and can be easily exported.

DATA HANDLING AND TEST REPORTS

The 2293 allows easy data handling. Results can be saved

on a USB memory stick or transferred to any computer. In

addition the instrument can be connected via Ethernet to a

LAN. It also includes a thermal printer for immediate report

printing.

Select

Start

Results can also be saved in the instrument’s memory

while continuing testing, to be downloaded later after

having returned to the office.

Results

Haefely has a policy of continuous product improvement. Therefore we reserve the right to change design and specification without notice.

LL_2293_0311_SG - 2/4

a brand of

Haefely is a subsidiary

of Hubbell Incorporated.