Atec Vishay-2311 User Manual

Page 4

background image

microstrain (0.0012 mV/V).
+25 000 microstrain (5% bridge unbalance or 12.5 mV/V), resolution 12.5
microstrain (0.006 mV/V).

Balance Time:
2 seconds, typical.

Manual Vernier Balance:
+50 microstrain (+0.025 mV/V).

Interaction:
Essentially independent of excitation and amplifier gain.

Storage:
Digital; up to 2 years without line power.

Shunt Calibration:
Circuit (2-level, dual polarity):
Single-shunt (for stress analysis) across any bridge arm, including
dummy gage.
Double-shunt (for transducer) across opposite bridge arms.

Provision for 4 dedicated leads to shunt external arms.

Cal circuit selected by switches on PC board.

Standard Factory-Installed Resistors (+0.l%) Simulate:
+200 and +1000 microstrain at GF=2 across dummy half bridge;
+1000 microstrain at GF=2 across dummy gage (120 and 350 ohm);
+1 mV/V (double-shunt) for 350 ohm transducer.

Remote-Operation Relays (Option Y):
4 relays (plus remote-reset relay for bridge balance and relay for excitation
on/off). Each requires 10 mA at 5 Vdc, except excitation on/off 25 mA.

Amplifier:
Gain:
1 to 11 000 continuously variable. Direct-reading.
+1% max of reading.

Ten-turn counting knob (X1 to X11) plus decade multiplier (X1 to X1000).

Frequency Response (all gains full output):
dc coupled: dc to 50 kHz, -0.5 dB max.
dc to 125 kHz, -3 dB max.
ac coupled: 1.7 Hz to 125 kHz, -3 dB max.

Model 2311 (Signal Conditioning Amplifier): Strain Gage Instruments

http://www.vishay.com/brands/measurements_group/guide/inst/2300/2311.htm (4 of 7) [12/9/2002 9:47:40 AM]