beautypg.com

Technical data sheet, Order codes and options, Technical specifications – Atec Wayne-Kerr-3265B User Manual

Page 2

background image








www.waynekerrtest.com [email protected]

October 2005 Issue A3

Wayne Kerr’s policy is one of continuous development and consequently the product may vary in detail from the description and specification in this publication.

Technical data sheet

UK

Wayne Kerr Electronics

Vinnetrow Business Park

Vinnetrow Road, Runcton

Chichester, West Sussex

PO20 1QH, UK

Tel: +44 1243 792200

Fax: +44 1243 792201

USA

Wayne Kerr Electronics

165L New Boston Street

Woburn

MA 01801-6201 USA
Tel: +1 781 938 8390

Sales: (800) 933 9319

Fax: +1 781 933 9523

1009 Fixture enables 50 A to be applied to an SMD inductor

Stable component fixturing ensures high accuracy and
repeatable measurements. Enclosed fixtures, with safety
interlocks, minimises risk to operators.

Technical specifications

The 3265B series of DC bias units must be used in conjunction
with either a 3260B, 3255BL, 3255B or 3255BQ.

Compliance voltages

Maximum compliance voltage (<12 kHz) 11 V DC at 0.25 V AC
drive level.
10 V compliance at 1 V AC drive level. For f>12 kHz deduct 0.5 V.

Applications

Permits measurement analysis of wound components with levels of
DC bias current higher than the standard 1A

Variable measured

In impedance mode: L, Z, R, Q, D.
Not applicable to Rdc, or transformer measurements

Measured frequency range

3255BL 20 Hz to 200 kHz (3265B/5A or 3265B10A only)
3255B: 20 Hz to 500 kHz
3255BQ: 20 Hz to 1 MHz
3260B: 20 Hz to 3 MHz (with 3265BQ/25A)

Basic accuracy

± 1%, Varies with measurement speed, frequency and options
chosen.

Measurement terminals

2-terminals measurement via M8 studs.
4-terminal measurement via Kelvin leads and M8 studs.
Measurement terminals internally protected by 1.6A fuses against
normal inductor back-EMF or accidental disconnection of inductor.
Front panel fuses easily replaced.

Control connections

I

2

C bus link controls application of DC current and monitors

status of analyzer. Status data includes excessive voltage drop
and over temperature

Optional facilities

3265B bias units may be connected in parallel to give five
times the rated current.

Interlock

Bias safety interlock socket on rear panel of analyzer provides
door lock and closed control lines

Temperature range

Storage: -40°C to 70°C.
Operating: 0°C to 40°C 20 A (3265BQ/25A and 3265B/25A )
Full accuracy: 15°C to 30°C (at maximum rated current)

Power supply

Universal 90 to 255 V AC, 47 to 63 Hz
Input current 9 A rms max.
Power factor >0.9
Unit powers up automatically when connected to a powered
analyzer. Isolating switch provided

Dimensions

Height 190 mm (7

1

/

2

″)

Width 440 mm (17

3

/

8

″)

Depth 520 mm (20

1

/

2

″)

Weight 15 kg (33 lb)

Cooling

Fan cooled. Intake front, exhaust rear. Fan filter accessible on
front panel. Over temperature trip provided

Order codes and options

Description

Order code

25A DC bias unit 3265BQ/25A (3 MHz) 1J3265BQ/25A
25A DC bias unit 3265B/25A

1J3265B/25A

10A DC bias unit 3265B/10A

1J3265B/10A

5A DC bias unit 3265B/5A

1J3265B/5A

All units are supplied with user manual, power cable,
spare fuses, 4 x BNC to BNC links and daisy chain
control link.

Accessories
Description

Order code

Kelvin clips leads (fine jaws)

1EVA40100

Kelvin clips leads (large jaws)

1EVA40180

Power transfer bus bars

4-324-6009-PAIR

Filter pad (washable)

HMDEC10

1009 DC bias fixture

1J1009

1009 50A lead set

5-328-2005