Technical data sheet, Order codes and options, Technical specifications – Atec Wayne-Kerr-3265B User Manual
Page 2
www.waynekerrtest.com [email protected]
October 2005 Issue A3
Wayne Kerr’s policy is one of continuous development and consequently the product may vary in detail from the description and specification in this publication.
Technical data sheet
UK
Wayne Kerr Electronics
Vinnetrow Business Park
Vinnetrow Road, Runcton
Chichester, West Sussex
PO20 1QH, UK
Tel: +44 1243 792200
Fax: +44 1243 792201
USA
Wayne Kerr Electronics
165L New Boston Street
Woburn
MA 01801-6201 USA
Tel: +1 781 938 8390
Sales: (800) 933 9319
Fax: +1 781 933 9523
1009 Fixture enables 50 A to be applied to an SMD inductor
Stable component fixturing ensures high accuracy and
repeatable measurements. Enclosed fixtures, with safety
interlocks, minimises risk to operators.
Technical specifications
The 3265B series of DC bias units must be used in conjunction
with either a 3260B, 3255BL, 3255B or 3255BQ.
Compliance voltages
Maximum compliance voltage (<12 kHz) 11 V DC at 0.25 V AC
drive level.
10 V compliance at 1 V AC drive level. For f>12 kHz deduct 0.5 V.
Applications
Permits measurement analysis of wound components with levels of
DC bias current higher than the standard 1A
Variable measured
In impedance mode: L, Z, R, Q, D.
Not applicable to Rdc, or transformer measurements
Measured frequency range
3255BL 20 Hz to 200 kHz (3265B/5A or 3265B10A only)
3255B: 20 Hz to 500 kHz
3255BQ: 20 Hz to 1 MHz
3260B: 20 Hz to 3 MHz (with 3265BQ/25A)
Basic accuracy
± 1%, Varies with measurement speed, frequency and options
chosen.
Measurement terminals
2-terminals measurement via M8 studs.
4-terminal measurement via Kelvin leads and M8 studs.
Measurement terminals internally protected by 1.6A fuses against
normal inductor back-EMF or accidental disconnection of inductor.
Front panel fuses easily replaced.
Control connections
I
2
C bus link controls application of DC current and monitors
status of analyzer. Status data includes excessive voltage drop
and over temperature
Optional facilities
3265B bias units may be connected in parallel to give five
times the rated current.
Interlock
Bias safety interlock socket on rear panel of analyzer provides
door lock and closed control lines
Temperature range
Storage: -40°C to 70°C.
Operating: 0°C to 40°C 20 A (3265BQ/25A and 3265B/25A )
Full accuracy: 15°C to 30°C (at maximum rated current)
Power supply
Universal 90 to 255 V AC, 47 to 63 Hz
Input current 9 A rms max.
Power factor >0.9
Unit powers up automatically when connected to a powered
analyzer. Isolating switch provided
Dimensions
Height 190 mm (7
1
/
2
″)
Width 440 mm (17
3
/
8
″)
Depth 520 mm (20
1
/
2
″)
Weight 15 kg (33 lb)
Cooling
Fan cooled. Intake front, exhaust rear. Fan filter accessible on
front panel. Over temperature trip provided
Order codes and options
Description
Order code
25A DC bias unit 3265BQ/25A (3 MHz) 1J3265BQ/25A
25A DC bias unit 3265B/25A
1J3265B/25A
10A DC bias unit 3265B/10A
1J3265B/10A
5A DC bias unit 3265B/5A
1J3265B/5A
All units are supplied with user manual, power cable,
spare fuses, 4 x BNC to BNC links and daisy chain
control link.
Accessories
Description
Order code
Kelvin clips leads (fine jaws)
1EVA40100
Kelvin clips leads (large jaws)
1EVA40180
Power transfer bus bars
4-324-6009-PAIR
Filter pad (washable)
HMDEC10
1009 DC bias fixture
1J1009
1009 50A lead set
5-328-2005