12
Model 4165
Range Maximum
Input Resolution
Test
Current
[1]
Accuracy
.2Ω .19999
Ω 10µΩ
100 mA
±0.03% of rdg ±3 digits
2Ω 1.9999
Ω 100µΩ
10 mA
±0.03% of rdg ±3 digits
20Ω 19.999
Ω 1mΩ
1 mA
±0.03% of rdg ±3 digits
200Ω 199.99
Ω 10mΩ
100 µA
±0.03% of rdg ±3 digits
2kΩ 1.9999
kΩ 100mΩ
10 µA
±0.03% of rdg ±3 digits
20kΩ 19.999
kΩ
1Ω
1 µA
±0.03% of rdg ±3 digits
Full Scale Test Voltage: 20mV
[1] The Model 4165 uses reduced test currents and incorporates a user-selectable clamp on the
current source for making "dry" measurements. The clamp may be set to 50mV or 2V.
Model 4165-1344
Range Maximum
Input
[2]
Resolution
Test Current
[1]
Accuracy
.2Ω .19999
Ω 10µΩ
100 mA
±0.03% of rdg ±3 digits
2Ω 1.9999
Ω 100µΩ
10 mA
±0.03% of rdg ±3 digits
20Ω 19.999
Ω 1mΩ
1 mA
±0.03% of rdg ±3 digits
200Ω 199.99
Ω 10mΩ
100 µA
±0.03% of rdg ±3 digits
2kΩ 1.9999
kΩ 100mΩ
10 µA
±0.03% of rdg ±3 digits
20kΩ 19.999
kΩ
1Ω
1 µA
±0.03% of rdg ±3 digits
Full Scale Test Voltage: 20mV
[1] The Model 4165-1344 uses reduced test currents and incorporates a user-selectable clamp on
the current source for making "dry" measurements. The clamp may be set to 20mV or 2V. The
4165-1344 also has a reverse current mode that may be used to ensure a dry, true-ohms
measurement.
[2] If the clamp is set to "20mV", the Maximum Input specification is changed to: 20% of range
for the .2Ω range or 50% of range for all other ranges.