4650a, Model 4650a accuracy specifications

background image

Valhalla Scientific, Inc.

4650A Specifications

V

ALHALLA

S

CIENTIFIC

,

I

NC

.

4650A

P

ROGRAMMABLE

D

IGITAL

I

GNITER

T

ESTER

S

PECIFICATIONS

_______________________________________________________________

Accuracy

MODEL 4650A ACCURACY SPECIFICATIONS

Rng # Maximum

Input

Maximum

1

Resolution

Accuracy

2

Accuracy

3

Test

4

Current

1

2.49999Ω 10µΩ

±0.015% of rng

±0.1% of range

5mA

2

24.9999Ω 100µΩ

±0.015% of rng

±0.1% of range

500µA

3

249.999Ω 1mΩ

±0.015% of rng

±0.1% of range

50µA

4

2.49999kΩ 10mΩ

±0.015% of rng

±0.1% of range

5µA

5

24.9999kΩ 100mΩ

±0.015% of rng

±0.1% of range

500nA

6

249.999kΩ

1Ω

±0.015% of rng

±0.1% of range

500nA

7

2.49999MΩ 10Ω

±0.5% of rng

±0.2% of range

50nA

8

24.99MΩ 10kΩ

±2% of rng

±0.2% of range

5nA

Notes:
[1] Resolution is user-selectable. The maximum available resolution for each range is shown.
[2] Valid for inputs above 10% of the maximum input.
[3] Valid for inputs from 1% to 10% of the maximum input. Inputs less than 1% are not specified.
[4] Fail-safe current is 7.5mA on all ranges if range 1 is installed, otherwise 750µA all ranges.

Operating Specifications


Compliance Voltage................................... Clamped at 50mV±5%(rngs 1-5) or 200mV±5%(rngs 6-8)

True Ohms Mode ...........................................................Pulsed test current available in ranges 1-4 only

CMRR......................................................................................................................... > 70dB @ 50-60Hz

NMRR......................................................................................................................... > 35dB @ 50-60Hz

Overload Protection................................................................................................. 100V

peak

all terminals