Fiber, Tools, 555b – Atec Rifocs-555B User Manual
Page 2: Specifications, Ordering information, Rifocs corporation, 1mm ingaas optical power meter
fiber
TOOLS
RIFOCS Corporation
Fiber Optic Instruments & Components
1340 Flynn Rd. Camarillo, CA 93012
Phone: (805) 389-9800 Fax: (805) 389-9808
http://www.rifocs.com
™
555B
1mm InGaAs Optical Power Meter
Specifications
1
Detector type
1mm indium-gallium-arsenide (InGaAs)
Calibration wavelengths
850nm, 1300nm, 1310nm, and 1550nm
Power range
+3 to -60dBm
Linearity at 1310nm and 1550nm:
±0.5dB
+3dBm to -3dBm
±0.05dB
-3dBm to -50dBm
±0.5dB
-50dBm to -60dBm
Absolute accuracy
±0.25dB at calibration conditions
Typical wavelength dependence:
820 to 880nm
0.033dB/nm
975 to 985nm
0.02dB/nm
1270 to 1330nm
0.005dB/nm
1500 to 1625nm
0.0024dB/nm
Polarization dependence
< 0.1dB
Resolution
±0.01dB
Power requirements
Two AA-size 1.5V alkaline batteries provide approx. 100 hours of
continuous operation
Connector interface
Snap-On Connector (SOC) interface
Environmental:
Operating temp.
-15°C to +55°C
Storage temp.
-35°C to +70°C
Humidity
0 to 95% RH, non-condensing
Dimensions
7.2 x 14.2 x 3.5 cm (2.8 x 5.6 x 1.4 in.)
Weight
250g (8.9 oz)
ISO 9001
Ordering Information
One Snap-On Connector (SOC) adapter is included
with the 555B optical power meter. Please specify the
desired connector adapter type when ordering using
the SOC Adapter Table, below. Additional SOC
adapters may also be ordered separately.
Part No.
Description
555B
555B optical power meter
90AC
AC power converter
DS-HH555B Rev. B
SOC Adapter Table
Part No.
Description
1001
Blank
1010
DIN 47256
1020
NTT/FC-PC
1030
AT&T/ST-PC
1038
MIL-T-29504 optical termini
1040
HMS-10 (2.5mm)
1047
Mini-BNC
1050
Diamond HMS-0 (3.5mm)
1057
Stratos 430/Holtek 38000
1062
NTT/SC-PC
1081
Radiall VFO
1086
Diamond HMS-10A (SMA-2.5)
1087
SMA-905/906
10E0
Radiall EC
10E2
Diamond E-2000
10TB
Simplex TOSLINK/Spectran J-pin
10TD
TR/TX set, duplex TOSLINK/Spectran J-pin
10TR
Duplex TOSLINK TX
10TX
Duplex TOSLINK TR
10ZP
H-P Versalink/Spectran V/Z-pin
Subject to change without notice
1 Within specified ambient environment of +20°C to +25°C.