beautypg.com

Portability, Specifications, Ordering information – Atec Fluke-3130 User Manual

Page 2: Models

background image

2 Fluke Calibration 3130 Portable Pressure Calibrator

Fluke Calibration

PO Box 9090,

Everett, WA 98206 U.S.A.

Fluke Europe B.V.

PO Box 1186, 5602 BD

Eindhoven, The Netherlands

For more information call:

In the U.S.A. (877) 355-3225 or Fax (425) 446-5116

In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222

In Canada (800)-36-FLUKE or Fax (905) 890-6866

From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116

Web access: http://www.flukecal.com

©2014 Fluke Calibration. This information is preliminary and subject to change.

Printed in U.S.A. 2/2014 4265091C_EN

Pub_ID: 12033-eng

Modification of this document is not permitted without written permission

from Fluke Calibration.

Fluke Calibration.

Precision, performance, confidence.

Portability

The 3130 is built into a ruggedized case with internal storage for power supply,

test leads, and fittings. In addition, it includes a rechargeable, NiMH battery,

allowing for approximately 50 hours of operation. When using the internal

pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi.

The battery is recharged by simply plugging the unit in using the included

universal mains adapter.

Specifications

Environmental
Operating temperature

-10 °C to +50 °C

Storage temperature

-20 °C to +60 °C

Power requirements

12 V dc (Universal ac adapter/charger supplied)

Battery

Internal 3800 mAh advanced NiMH pack

Operating time
On full charge

Approx 50 hours (Measure only or external air; no

pump)

Using internal pump

Approx 100 calibration cycles to 300 psi

Physical
Dimensions

15.25 in L x 12 in W x 7 in D

Weight

~7 kg (15 lb)

EMI/RFI conformance

EN61326:2006 Annex A

Connectors/ports

1/8 in NPT (External supply port and test port)

Included accessories

Manual, NIST-traceable certificate, test leads, univer-

sal ac adapter/charger

Ranges
Pressure (internal pump)

-80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)

Pressure (external air)

0 to 2 MPa (0 to 300 psi, 0 to 20 bar)

mA

0 to 24.000 mA

Volts

0 to 30.000 V dc

Engineering units

psi, bar, mbar, kPa, MPa, kgf/cm

2

, mmH

2

O @ 4 °C,

mmH

2

O @ 20 °C, cmH

2

O @ 4 °C, cmH

2

O @ 20 °C,

inH

2

O @ 4 °C, inH

2

O @ 20 °C, inH

2

O @ 60 °F, mmHg

@ 0 °C, inHg @ 0 °C

Instrumental measurement uncertainty
Pressure

0.025 % of reading ± 0.01 % FS

mA

0.015 % of reading ± 0.002 mA

Volts

0.015 % of reading ± 0.002 V

Temperature effect (all functions)
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C

Ordering Information

Models

3130-G2M

Portable Pressure Calibrator

with US, EUR, UK, and China/

Australia line cords (standard

calibration)

3130-G2M/C

Portable Pressure Calibrator

with US, EUR, UK, and China/

Australia line cords (ISO 17025

accredited calibration)

3130-G2M-2

Portable Pressure Calibrator

with Brazil, Italy, Switzerland,

and India line cords (standard

calibration)

3130-G2M-2/C Portable Pressure Calibrator

with Brazil, Italy, Switzerland,

and India line cords (ISO 17025

accredited calibration)