Portability, Specifications, Ordering information – Atec Fluke-3130 User Manual
Page 2: Models

2 Fluke Calibration 3130 Portable Pressure Calibrator
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2014 Fluke Calibration. This information is preliminary and subject to change.
Printed in U.S.A. 2/2014 4265091C_EN
Pub_ID: 12033-eng
Modification of this document is not permitted without written permission
from Fluke Calibration.
Fluke Calibration.
Precision, performance, confidence.
™
Portability
The 3130 is built into a ruggedized case with internal storage for power supply,
test leads, and fittings. In addition, it includes a rechargeable, NiMH battery,
allowing for approximately 50 hours of operation. When using the internal
pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi.
The battery is recharged by simply plugging the unit in using the included
universal mains adapter.
Specifications
Environmental
Operating temperature
-10 °C to +50 °C
Storage temperature
-20 °C to +60 °C
Power requirements
12 V dc (Universal ac adapter/charger supplied)
Battery
Internal 3800 mAh advanced NiMH pack
Operating time
On full charge
Approx 50 hours (Measure only or external air; no
pump)
Using internal pump
Approx 100 calibration cycles to 300 psi
Physical
Dimensions
15.25 in L x 12 in W x 7 in D
Weight
~7 kg (15 lb)
EMI/RFI conformance
EN61326:2006 Annex A
Connectors/ports
1/8 in NPT (External supply port and test port)
Included accessories
Manual, NIST-traceable certificate, test leads, univer-
sal ac adapter/charger
Ranges
Pressure (internal pump)
-80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)
Pressure (external air)
0 to 2 MPa (0 to 300 psi, 0 to 20 bar)
mA
0 to 24.000 mA
Volts
0 to 30.000 V dc
Engineering units
psi, bar, mbar, kPa, MPa, kgf/cm
2
, mmH
2
O @ 4 °C,
mmH
2
O @ 20 °C, cmH
2
O @ 4 °C, cmH
2
O @ 20 °C,
inH
2
O @ 4 °C, inH
2
O @ 20 °C, inH
2
O @ 60 °F, mmHg
@ 0 °C, inHg @ 0 °C
Instrumental measurement uncertainty
Pressure
0.025 % of reading ± 0.01 % FS
mA
0.015 % of reading ± 0.002 mA
Volts
0.015 % of reading ± 0.002 V
Temperature effect (all functions)
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C
Ordering Information
Models
3130-G2M
Portable Pressure Calibrator
with US, EUR, UK, and China/
Australia line cords (standard
calibration)
3130-G2M/C
Portable Pressure Calibrator
with US, EUR, UK, and China/
Australia line cords (ISO 17025
accredited calibration)
3130-G2M-2
Portable Pressure Calibrator
with Brazil, Italy, Switzerland,
and India line cords (standard
calibration)
3130-G2M-2/C Portable Pressure Calibrator
with Brazil, Italy, Switzerland,
and India line cords (ISO 17025
accredited calibration)