beautypg.com

726 specifications, General specifications, Ordering information – Atec Fluke-726 User Manual

Page 2

background image

726 Specifications

2 Fluke Corporation 726 Precision Multifunction Process Calibrator

Fluke Corporation
PO Box 9090, Everett, WA USA 98206

Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands

For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com

©2005 Fluke Corporation. All rights reserved.
Printed in U.S.A. 12/2005 2527590 D-EN-N Rev C

Fluke. Keeping your world

up and running.

General Specifications

Storage temperature: -20 °C to 71 °C
Operating temperature: -10 °C to 55 °C
Relative humidity: 90 % (10 °C to 30 °C);
75 % (30 °C to 40 °C); 45 % (40 °C to 50
°C); 35 % (50 °C to 55 °C)
Shock: 1 meter drop test
Safety: CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and
EN55022:1994 Class B
Size (HxWxD): 200 mm x 96 mm x 47 mm
(7.9 in x 3.8 in x 1.9 in)
Weight: 650 g (23 oz)
Battery: Four AA alkaline batteries
Battery life: 25 hours typical
Warranty: Three-years

Ordering Information

Fluke-726

Precision Multifunction
Process Calibrator

Included

TL75 Test Leads, AC72 Test Clips, one pair
of stackable test leads, product overview
manual (print) and user’s manual (CD-ROM)
in14 languages.

Simultaneous Function Capability

Channel A

Channel B

24.000 mA DC

M

M or S

24.000 mA DC with 24 V loop supply

M

100.00 mV DC

M or S

30.000 V DC measure

M

20.000 V DC measure, 20.000 VDC Source

M or S

5 to 4000 Ohms

M or S

Thermocouple J, K, T, E, R, S, B, M, L, U, N, C, BP, XK

M or S

RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200,

M or S

500, 1000 (385), Cu 10

Pressure (requires Fluke 700PXX Modules)

M

M used as S

Frequency; Squarewave, 1 CPM to10 kHz;

M or S

fixed amplitude 5 V p-p

M = Measure, S = Source/Simulate

Function

Measure or Source

Range or Type

Resolution

Accuracy

Notes

Voltage

0 to 100 mV

0.01 mV

0.01 % Rdg + 2 LSD

Max load, 1 mA

0 to 20 V (source)

0.01 V

0 to 30 V (measure)

0.01 V

mA

0 to 24

0.001 mA

0.01 % Rdg + 2 LSD

Max load, 1000

mV (TC terminals)

-10.00 mV to +75.00 mV

.01 mV

0.01 % of range + 1 LSD

Resistance 5

Ω to 4000 Ω

≤ 0.01 Ω to 0.1 Ω

0.015 %

Frequency

2.0 to 1,000 CPM

0.1 CPM

± 0.05 %

For frequency source;

1 to 1100 Hz

1 Hz

± 0.05 %

1 V to 20 V p-p

1.0 to 10.0 kHz

0.1 kHz

± 0.25 %

squarewave, -0.1 V offset

10.0 to 15.0 kHz

0.1 kHz

± 0.5 %

Loop Supply

24 V dc

N/A

10 %

Thermocouples

J, K, T, E, L, N, U, C, BP, XK

0.1 ºC, 0.1 ºF

to 0.2 ºC

Thermocouples

B, R, S

1 ºC, 1 ºF

to 1.2 ºC

RTDs

Cu 10 Ni 120 (672)

0.1 °C, 0.1 °F

to 1.8 °C

Pt 100, 200, 500,

1000 (385)

Pt 100 (3916),

Pt 100 (3926) Cu 10

0.01 ºC, 0.01 °F

to 0.15 ºC