4 Fluke Calibration Series 7250 Ruska Pressure Controller/Calibrator

General
Temperature

operating: 18 °C to 36 °C (64 °F to 96 °F)

Storage: -20 °C to 70 °C (-4 °F to 158 °F)

Humidity

5 % to 95 % RH, non-condensing

Weight

Model 7250/7250i: 7.7 kg (17 lb)

Model 7250xi: 9 kg (20 lb)

Dimensions (H x W x D)

All versions: 178 mm x 419 mm x 483 mm (7 in x 16.5 in x 19 in)

Pressure medium

Nitrogen or clean dry air

Display

TFT, VGA, active matrix, 162.5 mm (6.4 in) 640 x 480 resolution, 65,000 colors

Test port and supply

connection

1/4 in NPT female

Warm up time

2 to 3 hours; may be left on indefinitely

Pressure ranges
Model 7250xi

Any Full Scale pressure range from 20 psig to 2 500 psig (140 kPa to

17.2 MPa).

Model 7250i

Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa).

or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa

absolute).

Model 7250

Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa).

or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa

absolute). For ranges below 35 kPa (5 psi), see the 7250LP.

optional modes

Absolute using barometric reference sensor for ranges from 15 psig to

2 500 psig (100 kPa to 17.2 MPa)
Vacuum (negative gauge)
Vacuum reference (requires external vacuum pump)

Model 7250HP

3000 psi absolute (21 MPa absolute)

Performance
Precision

1

Model 7250xi: From 5 % to 100 % full scale (FS): 0.005 % of reading

Below 5 % FS: 0.005% of 5% FS
Model 7250i: From 25 % to 100 % FS: 0.005 % of reading

Below 25 % FS: 0.005 % of 25 % FS
Model 7250: Ranges to 2 500 psi (17.2 MPa): 0.003 % of FS
Model 7250HP: 0.012% of reading or 0.0036% FS, whichever is greater

Stability

Ranges 2 500 psi (17.2 MPa): 0.0075 % of reading per year

3 000 psia (21 MPa) range: 0.005% of reading

Control stability

Active mode: 0.001 % FS (10 ppm)

Passive mode: no additional uncertainty

Display resolution

User selectable to 1:1,000,000

Control response

20 seconds or less with zero overshoot into a 245 cm

(15 in

3

) volume for ranges

to 2 500 psi (17.2 MPa)
60 seconds for 3 000 psia (21 MPa) range

Negative gauge

precision (optional)

Model 7250xi: 0.005 % of 5 % FS or 0.00075 psi (0.05 mbar), whichever is

greater
Model 7250i: 0.005 % of 25 % FS or 0.0075 psi (0.5 mbar), whichever is

greater
Model 7250: 0.003 % of FS
Model 7250HP: Not applicable

Barometric reference

(optional)

0.002 psi (0.014 kPa) maximum error per year

Vacuum reference

(optional)

0.0002 psi (0.0014 kPa) maximum error per year (only available on 7250i

and 7250)

Calibration
An accredited calibration report is provided standard. For information on the calibration uncertainty,

please see the appropriate Fluke Calibration Scope of Accreditation.

Specifications

Total Uncertainty (k=2)

2

Model 7250xi

5 % to 100 % FS

90 day: 0.006 % reading

1 year: 0.009 % reading

Model 7250i

25 % to 100 % FS

90 day: 0.006 % reading

1 year:0.009 % reading

Model 7250

90 day: 0.003 % FS + 0.002 % reading

1 year: 0.003 % FS + 0.0075 % reading

Model 7250HP

30 % to 100 % FS

90 day: 0.012% reading

1 year: 0.013% reading

Control parameters
Volume

82 cm to 980 cm (5 in

3

to 60 in

3

)

Low Control (ab-

solute)

0.15 psia (1 kPa)

Communications
RS-232 and IEEE-488, SCPI syntax. Series 7215, Model 7010, Series

6000 and DPI 510 emulation are standard.
MET/CAL

®

driver

optional

LabView

®

driver

optional

Firmware updates are performed via RS-232 interface
Languages
The Series 7050 is capable of displaying menus and functions in:

English, French, Chinese, German, Japanese, Spanish and Italian
Options
Barometric reference (absolute and vacuum)

Vacuum mode (negative gauge)

Vacuum reference (requires external vacuum pump)
Accessories
3911197

Transportation Case with Wheels and Handles

3879677

Rack mount kit for 19 in. EIA cabinets, 24 in. deep

3879689

Rack mount kit for 19 in. EIA cabinets, 30 in. deep

3876298

Lines and Fittings Kit

4014713

Vacuum Pump with auto-vent, 85 liters/minute

3876271

Liquid Trap

3891089

MET/CAL Pseudo-Driver

3070175

CoMPASS for Pressure Enhanced, Single License

Model 7250i

Uncertainty (2 sigma)

Precision

0.005 % of reading

Stability (one year)

0.0075 % of reading

Calibration standard

0.0010 % of reading

Environmental temperature

(included in precision)

0.000 % of reading

Head pressure

0.001 % of reading

Expanded uncertainty (2

sigma)

0.009 % of reading

Fluke Calibration

Po Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.

Po Box 1186, 5602 BD

Eindhoven, The Netherlands

Fluke Calibration.

Precision, performance, confidence.

For more information call:

In the U.S.A. (877) 355-3225 or

Fax (425) 446-5116

In Europe/M-East/Africa +31 (0) 40

2675 200 or Fax +31 (0) 40 2675 222

In Canada (800)-36-FLUKE or

Fax (905) 890-6866

From other countries +1 (425) 446-5500 or

Fax +1 (425) 446-5116

Web access: http://www.flukecal.com

©2010 Fluke Corporation.

Specifications subject to change without notice.

Printed in U.S.A. 9/2011 3833773B D-EN-N

Modification of this document is not permitted

without written permission from Fluke Corporation.

1

Precision is defined as the combined effects of linearity, repeatability

and hysteresis throughout the operating temperature range.

2

Total uncertainty is the maximum deviation from the true value of

pressure including precision, stability, temperature effects and the

calibration standard. This assumes routine zeroing of the instrument.

Expression of uncertainty conforms with the recommendations of the

ISO Guide to the Expression of Uncertainty in Measurement.