Specifications, Fluke calibration, Precision, performance, confidence – Atec Fluke-7250 Series User Manual
Page 4
4 Fluke Calibration Series 7250 Ruska Pressure Controller/Calibrator
General
Temperature
operating: 18 °C to 36 °C (64 °F to 96 °F)
Storage: -20 °C to 70 °C (-4 °F to 158 °F)
Humidity
5 % to 95 % RH, non-condensing
Weight
Model 7250/7250i: 7.7 kg (17 lb)
Model 7250xi: 9 kg (20 lb)
Dimensions (H x W x D)
All versions: 178 mm x 419 mm x 483 mm (7 in x 16.5 in x 19 in)
Pressure medium
Nitrogen or clean dry air
Display
TFT, VGA, active matrix, 162.5 mm (6.4 in) 640 x 480 resolution, 65,000 colors
Test port and supply
connection
1/4 in NPT female
Warm up time
2 to 3 hours; may be left on indefinitely
Pressure ranges
Model 7250xi
Any Full Scale pressure range from 20 psig to 2 500 psig (140 kPa to
17.2 MPa).
Model 7250i
Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa).
or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa
absolute).
Model 7250
Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa).
or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa
absolute). For ranges below 35 kPa (5 psi), see the 7250LP.
optional modes
Absolute using barometric reference sensor for ranges from 15 psig to
2 500 psig (100 kPa to 17.2 MPa)
Vacuum (negative gauge)
Vacuum reference (requires external vacuum pump)
Model 7250HP
3000 psi absolute (21 MPa absolute)
Performance
Precision
1
Model 7250xi: From 5 % to 100 % full scale (FS): 0.005 % of reading
Below 5 % FS: 0.005% of 5% FS
Model 7250i: From 25 % to 100 % FS: 0.005 % of reading
Below 25 % FS: 0.005 % of 25 % FS
Model 7250: Ranges to 2 500 psi (17.2 MPa): 0.003 % of FS
Model 7250HP: 0.012% of reading or 0.0036% FS, whichever is greater
Stability
Ranges 2 500 psi (17.2 MPa): 0.0075 % of reading per year
3 000 psia (21 MPa) range: 0.005% of reading
Control stability
Active mode: 0.001 % FS (10 ppm)
Passive mode: no additional uncertainty
Display resolution
User selectable to 1:1,000,000
Control response
20 seconds or less with zero overshoot into a 245 cm
(15 in
3
) volume for ranges
to 2 500 psi (17.2 MPa)
60 seconds for 3 000 psia (21 MPa) range
Negative gauge
precision (optional)
Model 7250xi: 0.005 % of 5 % FS or 0.00075 psi (0.05 mbar), whichever is
greater
Model 7250i: 0.005 % of 25 % FS or 0.0075 psi (0.5 mbar), whichever is
greater
Model 7250: 0.003 % of FS
Model 7250HP: Not applicable
Barometric reference
(optional)
0.002 psi (0.014 kPa) maximum error per year
Vacuum reference
(optional)
0.0002 psi (0.0014 kPa) maximum error per year (only available on 7250i
and 7250)
Calibration
An accredited calibration report is provided standard. For information on the calibration uncertainty,
please see the appropriate Fluke Calibration Scope of Accreditation.
Specifications
Total Uncertainty (k=2)
2
Model 7250xi
5 % to 100 % FS
90 day: 0.006 % reading
1 year: 0.009 % reading
Model 7250i
25 % to 100 % FS
90 day: 0.006 % reading
1 year:0.009 % reading
Model 7250
90 day: 0.003 % FS + 0.002 % reading
1 year: 0.003 % FS + 0.0075 % reading
Model 7250HP
30 % to 100 % FS
90 day: 0.012% reading
1 year: 0.013% reading
Control parameters
Volume
82 cm to 980 cm (5 in
3
to 60 in
3
)
Low Control (ab-
solute)
0.15 psia (1 kPa)
Communications
RS-232 and IEEE-488, SCPI syntax. Series 7215, Model 7010, Series
6000 and DPI 510 emulation are standard.
MET/CAL
®
driver
optional
LabView
®
driver
optional
Firmware updates are performed via RS-232 interface
Languages
The Series 7050 is capable of displaying menus and functions in:
English, French, Chinese, German, Japanese, Spanish and Italian
Options
Barometric reference (absolute and vacuum)
Vacuum mode (negative gauge)
Vacuum reference (requires external vacuum pump)
Accessories
3911197
Transportation Case with Wheels and Handles
3879677
Rack mount kit for 19 in. EIA cabinets, 24 in. deep
3879689
Rack mount kit for 19 in. EIA cabinets, 30 in. deep
3876298
Lines and Fittings Kit
4014713
Vacuum Pump with auto-vent, 85 liters/minute
3876271
Liquid Trap
3891089
MET/CAL Pseudo-Driver
3070175
CoMPASS for Pressure Enhanced, Single License
Model 7250i
Uncertainty (2 sigma)
Precision
0.005 % of reading
Stability (one year)
0.0075 % of reading
Calibration standard
0.0010 % of reading
Environmental temperature
(included in precision)
0.000 % of reading
Head pressure
0.001 % of reading
Expanded uncertainty (2
sigma)
0.009 % of reading
Fluke Calibration
Po Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
Po Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke Calibration.
Precision, performance, confidence.
™
For more information call:
In the U.S.A. (877) 355-3225 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40
2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2010 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 9/2011 3833773B D-EN-N
Modification of this document is not permitted
without written permission from Fluke Corporation.
1
Precision is defined as the combined effects of linearity, repeatability
and hysteresis throughout the operating temperature range.
2
Total uncertainty is the maximum deviation from the true value of
pressure including precision, stability, temperature effects and the
calibration standard. This assumes routine zeroing of the instrument.
Expression of uncertainty conforms with the recommendations of the
ISO Guide to the Expression of Uncertainty in Measurement.