Corrected absolute ac/dc uncertainty, Performance specifications, Fully traceable performance – Atec Fluke-792A User Manual
Page 4: Description, 1 year, t, 12°c, Volt-hertz product: 1x10, Volt-hertz product: 1 x 10
Corrected absolute ac/dc uncertainty
1 year, T
CAL
± 12°C
AC/DC uncertainty
±
ppm input
Voltage
Frequency
Voltage
Voltage
Range
Input
10 Hz
20 Hz
40 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
800 kHz
1 MHz
22 mV
2 mV
2300
1700
1700
1700
1700
1700
1700
1700
2000
2700
5000
5600
5600
6 mV
780
430
380
380
380
380
380
500
1200
1700
2100
2400
2400
10 mV
500
300
300
300
300
300
300
300
600
800
1400
2000
2000
20 mV
400
200
200
180
180
180
180
250
600
800
1200
1300
1300
220 mV
20 mv
450
250
220
220
220
220
220
280
700
900
1400
1200
1200
60 mV
260
130
90
70
70
70
70
130
270
560
660
650
650
100 mV
240
100
55
55
55
55
55
72
160
300
330
380
380
200 mV
220
90
45
34
34
34
34
72
160
240
270
330
330
700 mV
200 mV
230
90
55
55
55
55
55
83
160
240
270
330
330
600 mV
210
75
35
26
26
26
26
50
65
130
150
170
170
2.2V
600 mV
210
65
30
25
25
25
25
45
60
120
140
140
140
1V
200
60
30
15
15
15
15
40
50
120
130
130
130
2V
200
60
25
10
10
10
10
40
50
120
130
130
130
7V
2V
200
65
30
22
22
22
22
45
55
120
140
160
160
6V
200
60
25
10
10
10
10
40
50
120
130
140
140
22V
6V
200
65
30
22
22
22
22
45
55
120
130
170
170
10V
200
60
30
15
15
15
15
40
50
120
130
150
150
20V
200
60
25
15
15
15
15
40
50
120
130
150
150
70V
20V
200
70
30
25
25
25
25
55
70
130
150
150
150
60V
200
65
25
20
20
20
20
50
65
130
220V
60V
200
70
33
33
33
33
33
65
70
130
100V
200
70
30
20
20
20
20
60
70
200V
200
65
30
18
18
18
18
60
70
1000V
200V
200
95
40
33
33
33
33
65
70
600V
200
95
39
27
27
30
30
60
70
1000V
200
95
30
25
25
27
27
60
70
Volt-hertz product: 1x10
8
@ 100 kHz, 2.2x10
7
@ 1 MHz
Waveform requirements: Sinusoidal, distortion less than 1%
792A AC/DC Transfer Standard Fluke Corporation 4
Corrected ac/dc uncertainties relative to calibration standards
1 year, T
CAL
± 5°C
AC/DC uncertainty
±
ppm input
Voltage
Frequency
Voltage
Voltage
Range
Input
10 Hz
20 Hz
40 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
800 kHz
1 MHz
22 mV
2 mV
1300
1200
1100
1100
1100
1100
1100
1300
1500
2000
3000
4000
4000
6 mV
400
325
275
275
275
275
275
400
750
1000
1300
1500
1500
10 mV
300
260
225
225
225
225
225
250
400
650
900
1000
1000
20 mV
250
150
125
125
125
125
125
175
350
500
650
700
700
220 mV
20 mV
300
160
150
130
130
130
130
180
400
550
700
800
800
60 mV
200
50
25
25
25
25
25
50
250
450
450
500
500
100 mV
150
50
12
12
12
12
12
50
150
200
250
300
300
200 mV
80
50
10
10
10
10
10
50
50
200
200
275
275
700 mV
200 mV
70
50
25
25
25
25
25
50
50
200
200
275
275
600 mV
70
19
7
7
7
7
7
30
35
70
125
130
130
2.2V
600 mV
70
19
10
10
10
10
10
30
35
70
125
130
130
1V
70
19
7
7
7
7
7
30
35
70
100
100
100
2V
70
19
7
6
6
7
7
30
35
70
10
100
100
7V
2V
70
19
15
15
15
15
15
30
35
70
125
130
130
6V
70
19
7
6
6
7
7
30
35
70
100
100
100
22V
6V
70
19
15
15
15
15
15
30
35
70
125
130
130
10V
70
19
7
7
7
7
7
30
35
70
100
100
100
20V
70
19
7
7
7
7
7
30
35
70
100
100
100
70V
20V
70
19
15
15
15
15
15
30
35
70
115
115
115
60V
70
19
7
7
7
7
7
30
35
70
220V
60V
70
19
15
15
15
15
15
30
35
100
100V
70
19
7
7
7
7
7
30
50
200V
70
19
7
7
7
7
7
30
50
1000V
200V
70
19
17
16
16
18
18
30
50
600V
100
35
17
16
16
18
18
30
50
1000V
100
35
10
10
10
10
10
30
50
Volt-hertz product: 1 x 10
8
@ 100 kHz, 2.2 x 10
7
@ 1 MHz
Waveform requirements: Sinusoidal, distortion less than 1%.
Corrected absolute ac/dc uncertainty
1 year, T
CAL
± 5°C
AC/DC uncertainty
±
ppm input
Voltage
Frequency
Voltage
Voltage
Range
Input
10 Hz
20 Hz
40 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
800 kHz
1 MHz
22 mV
2 mV
2200
1600
1600
1600
1600
1600
1600
1600
1900
2500
4500
5000
5000
6 mV
500
400
350
350
350
350
350
500
1000
1500
1800
1800
1800
10 mV
320
280
280
280
280
280
280
300
500
700
1200
1500
1500
20 mV
280
180
140
125
125
125
125
200
450
650
1000
1000
1000
220 mV
20 mV
350
230
190
190
190
190
190
220
450
650
1000
1100
1100
60 mV
230
120
80
60
60
60
60
120
250
500
600
600
600
100 mV
210
90
50
50
50
50
50
65
150
270
300
350
350
200 mV
200
80
40
30
30
30
30
65
150
220
250
300
300
700 mV
200 mV
220
80
50
50
50
50
50
75
150
220
250
300
300
600 mV
200
70
30
22
22
22
22
45
60
130
140
140
140
2.2V
600 mV
200
65
30
25
25
25
25
45
60
120
140
140
140
1V
190
60
30
15
15
15
15
40
50
115
125
125
125
2V
190
60
25
10
10
10
10
40
50
115
125
125
125
7V
2V
190
65
30
22
22
22
22
45
55
120
140
140
140
6V
190
60
25
10
10
10
10
40
50
120
125
125
125
22V
6V
190
65
30
22
22
22
22
45
55
120
125
140
140
10V
190
60
30
15
15
15
15
40
50
120
125
125
125
20V
190
60
25
15
15
15
15
40
50
120
125
125
125
70V
20V
190
65
30
25
25
25
25
55
70
130
140
140
140
60V
190
60
25
20
20
20
20
50
65
130
220V
60V
190
65
33
33
33
33
33
65
70
130
100V
190
65
30
20
20
20
20
60
70
200V
190
60
27
18
18
18
18
60
70
1000V
200V
190
90
40
33
33
33
33
65
70
600V
190
90
35
27
27
30
30
60
70
1000V
190
90
27
25
25
27
27
60
70
Volt-hertz product: 1x10
8
@ 100 kHz, 2.2x10
7
@1 MHz
Waveform requirements: Sinusoidal, distortion less than 1%
792A AC/DC Transfer Standard Fluke Corporation 3
Performance
specifications
There are three tables of uncer-
tainties included for the 792A.
The first two are absolute and
traceable to the National Insti-
tute of Standards and Technol-
ogy (NIST). Absolute uncertainty
includes time stability, tempera-
ture, line and load regulation,
and the traceability provided by
NIST (see Fluke application note
B0205A, Establishing Traceabil-
ity for a High Performance
AC/DC Transfer Standard). The
third table, Relative Uncertainty,
includes time stability, tempera-
ture, line and load regulation,
but not the traceability to
national or external standards.
These uncertainties apply
when the calibration report,
provided with the 792A, is used
to correct the response of the
transfer standard.
2 Fluke Corporation 792A AC/DC Transfer Standard
The RMS Sensor is designed
to be rugged and reliable. Each
is built to exacting standards by
the Fluke Microelectronics Op-
eration to maintain quality and
consistency, part after part.
Fully traceable
performance
Each 792A is shipped from
Fluke with documented trace-
ability to the US National Insti-
tute of Standards and Technol-
ogy (NIST). Included is a table
of correction factors and uncer-
tainties for measured ac/dc
differences.
To support the traceability
requirements for an instrument
as accurate as the 792A, the
Fluke Metrology Department
developed a unique system to
transfer NIST values to the 792A
production environment. This
system is based on proven tech-
niques developed to maintain a
732A direct voltage standard to
within a few parts in 100 million.
Statistical treatment of data
resulting from successive
intercomparisons of virtually
identical 792As is used to mini-
mize transfer uncertainties. To
achieve even higher perfor-
mance, you can have the trans-
fer uncertainties of your 792A
assigned directly by NIST, or
other national standards organi-
zations. For more information,
request the application note
titled Establishing Traceability
for a High Performance AC/DC
Transfer Standard.
Periodic calibration and
recertification is available from
Fluke. This service repeats the
original calibration procedure.
A new table of correction factors
and undertainties for measured
ac/dc differences is returned
with your instrument. Order
792A-000.
Description
The 792A consists of four units.
The Transfer Unit is the main
analog component of the 792A
for the 20 mV to 220V ranges.
Stainless Steel Type-N connec-
tors assure low signal loss and
high measurement repeatability.
The separate Power Pack unit
permits the Transfer Unit and
1000V Range Resistor to be
shipped independently for
calibration. The 1000V Range
Resistor isolates the heat gener-
ated at high voltages outside
the main Transfer Unit. The
Range Resistor has a low tem-
perature coefficient, stabilizes
quickly and exhibits small ac/dc
differences. The Transfer Switch
provides for switching between
the ac and dc inputs.
Each 792A is shipped from
the factory traceable to NIST
standards. Each 792A also
includes a table of correction
factors and uncertainties for
measured ac/dc differences.
The 792A consists of four main components: The Transfer Unit, Power Pack, 1000V Range
Resistor and Transfer Switch.
Each 792A includes a table of correction factors and uncertainties for measured ac/dc differences.