Field dry-wells – Atec Fluke-9141 User Manual
Page 2

diameter for doing comparison calibra-
tions. Insert C has six 1/4-inch-diameter
wells for multiple probe calibrations, and
Insert D has three pairs of metric sized
wells.
9140
The 9140 has a temperature range of
35 °C to 350 °C, and it reaches its maxi-
mum temperature in 12 minutes. At six
pounds, it’s small enough to easily carry
in one hand. It’s truly a unique innovation
in dry-wells.
The unit has a stability of ±0.05 °C or
better and a uniformity of at least 0.4 °C
in the largest-diameter wells and 0.1 °C
in the smaller wells. Despite its small
size, this unit performs.
Use the display, calibrated to ±0.5 °C, as
your reference, or use an external ther-
mometer for maximum calibration accuracy.
With three removable inserts to choose
from, the 9140 is as versatile as it is fast.
9141
Here’s an upright unit you’re going to
love. It does calibrations up to 650 °C,
weighs only eight pounds, and heats up
to 650 °C in only 12 minutes—12! This
dry-well does everything but get legs
and walk to the job for you. (And we’re
working on one that does that too.)
This four-inch-wide dry-well is amaz-
ing. You can control all functions from the
front panel or hook it up to your PC with
its built-in RS-232 port. And just like the
9140, it works with all of our software
described on page 82.
It has three removable well inserts
available, an optional carrying case, a
NIST-traceable calibration, and the best
price in the industry.
Specifications
9103
9140
9141
Range
–25 °C to 140 °C (–13 °F to 284 °F)
at 23 °C ambient
35 °C to 350 °C (95 °F to 662 °F)
50 °C to 650 °C (122 °F to 1202 °F)
Accuracy
±0.25 °C
±0.5 °C (holes greater than 1/4"
[6.35 mm]: ±1 °C)
±0.5 °C to 400 °C; ±1.0 °C to 650 °C
(holes greater than 1/4": ±2 °C)
Stability
±0.02 °C at –25 °C
±0.04 °C at 140 °C
±0.03 °C at 50 °C
±0.05 °C at 350 °C
±0.05 °C at 100 °C
±0.12 °C at 500 °C
±0.12 °C at 650 °C
Well-to-Well
Uniformity
±0.1 °C between similarly sized wells
±0.1 °C with similarly sized wells
±0.1 °C below 400 °C, ±0.5 °C above
400 °C with similarly sized wells
Heating Times
18 minutes from ambient to 140 °C
12 minutes from ambient to 350 °C
12 minutes from ambient to 650 °C
Cooling Times
20 minutes from ambient to –25 °C
15 minutes from 350 °C to 100 °C
25 minutes from 650 °C to 100 °C
Stabilization Time
7 minutes
Immersion Depth
124 mm (4.875")
Inserts
Insert A, B, C, or D included (specify when ordering)
Outside Insert
Dimensions
31.8 mm dia. x 124 mm (1.25 x 4.88 in)
28.5 mm dia. x 124 mm (1.12 x 4.88 in)
Computer Interface
RS-232 included with free Interface-it software (Model 9930)
Power
115 VAC (±10 %), 1.3 A or 230 VAC
(±10 %), 0.7 A, switchable, 50/60 Hz,
150 W
115 VAC (±10 %), 4.4 A or 230 VAC
(±10 %), 2.2 A, switchable, 50/60 Hz,
500 W
115 VAC (±10 %), 8.8 A or 230 VAC
(±10 %), 4.4 A, switchable, 50/60 Hz,
1000 W
Size (WxHxD)
143 x 261 x 245 mm
(5.63 x 10.25 x 9.63 in)
152 x 86 x 197 mm
(6 x 3.375 x 7.75 in)
109 x 236 x 185 mm
(4.3 x 9.3 x 7.3 in)
Weight
5.7 kg (12 lb.)
2.7 kg (6 lb.)
3.6 kg (8 lb.)
NIST-Traceable
Certificate
Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C,
100 °C, and 140 °
Data at 50 °C, 100 °C, 150 °C, 200 °C,
250 °C, 300 °C, and 350 °C
Data at 100 °C, 200 °C, 300 °C, 400 °C,
500 °C, and 600 °C
Field Dry-Wells
Ordering Information
9103-X Dry-Well (specify X, X = A, B, C, or D
included insert)
3103-1 Insert,
blank
3103-2 Insert A
3103-3 Insert B
3103-4 Insert C
3103-6 Insert D
9316
Rugged Carrying Case
9140-X Dry-Well (specify X, X = A, B, C, or D
included insert)
3140-1 Insert,
blank
3140-2 Insert A
3140-3 Insert B
3140-4 Insert C
3140-6 Insert D
9308
Rugged Carrying Case
9141-X Dry-Well (specify X, X = A, B, C, or D
included insert)
3141-1 Insert,
blank
3141-2 Insert A
3141-3 Insert B
3141-4 Insert C
3141-6 Insert D
9309