Atec HP-Agilent-5347A User Manual
Atec Equipment
Agilent 5347A 20 GHz Microwave Counter /
Power Meter
Data Sheet
Product Specifications
Frequency Range: 500 MHz to 20 GHz Sensitivity 500 MHz - 12.4 GHz: -32 dBm (-35 dBm typical) 12.4 GHz - 20.0
GHz: -27 dBm (-32 dBm typical)
Maximum Input: +7 dBm Damage Level: +25 dBm, peak SWR 500 MHz - 12.4 GHz: <2:1 typical 12.4 GHz - 20.0
GHz: <3:1 typical Accuracy: ±1 LSD ±time base error x frequency
Resolution: 1 Hz or 10 kHz, selectable
Optional Increased Damage Level Opt 006 Protects Input 1 from damage by limiting high level signals. All
specifications are the same except Input 1.
Damage Level 500 MHz to 6 GHz: +39 dBm (8 watts) 6 GHz to 18 GHz: +36 dBm (4 watts) Sensitivity reduced by 3
dBm, 500 MHz to 12.4 GHz 4 dBm, 12.4 GHz to 20.0 GHz
Input 2
Frequency Range: 10 Hz - 525 MHz
Sensitivity: 25 mV rms (15 mV rms typical)
Impedance: 1 Mohm nominal shunted by <70 pF (10 Hz to 80 MHz) or 50 Ù nomimnal (10 MHz to 525 MHz)
Damage Level: 50 Ù or 1 MÙ dc - 5 kHz: 250V (dc + ac peak) >5 KHz: 5.5V rms (+28 dBm) + 1.25 x 10
6
V
rms/FREQ
Accuracy: ±1 LSD ±[(1.4 x Trigger Error/Gate Time)±Time Base Error] x frequency
Resolution: 1 Hz or 10 kHz, selectable
Automatic Amplitude Discrimination Automatically measures the largest of all signals present, provided that signal is
>6 dB (typical) above any signal within 500 MHz; >20 dB (typical) above any signal within 500 MHz to 20 GHz.
Tracking Speed Resolution = 1 Hz, Speed = 1 MHz/sec Resolution = 10 kHz, Speed = 1 GHz/sec Acquisition Time
Resolution = 1 Hz, Time = <125 ms Resolution = 10 kHz, Time = <60 ms Maximum Deviation 20 MHz p-p, Automatic
mode 60 MHz p-p, Manual mode (via HP-IB only)
Maximum FM Rate: 10 MHz
AM Tolerance: Any modulation index provided the minimum signal level is not less than the sensitivity specification.
Power Meter
Frequency Range: 0 MHz to 20 GHz, sensor dependent Power Range: -70 dBm to +20 dBm (100 pW to 100 mW),
sensor dependent
Dynamic Range: 50 dB in 10 dB steps Resolution: 0.01 dB in logarithmic mode, 0.1% of full scale in linear mode.
Auto Filter: The meter automatically selects the required number of averages for the selected range.
Accuracy Instrumentation: ±0.02 dB or ±5%
Zero Set (digital settability of zero): ±5% of full scale on most sensitive range. Decrease percentage by a factor of 10
for each higher range, ±1 display count.