beautypg.com

Rf vector signal analyzer, Si de t ex t, Mod el 28 20 spe ci fic at io ns – Atec Keithley-2820 User Manual

Page 11: Trigger and synchronization inputs and outputs

background image

www.keithley.com

1.888.KEITHLEY

(U.S. only)

Si

de

T

ex

t

R

F/

M

Ic

R

o

w

a

v

E

M

od

el

28

20

spe

ci

fic

at

io

ns

Trigger and Synchronization Inputs and outputs

trigger SourCeS

1

:

Free run

External

Video

Bus

External arm using video trigger

Bus arm using external or video trigger

Latched External

trigger delAy:

–30 to +30 seconds.

trigger ModeS:

On measurement.

On acquire.

externAl trigger:

Selectable on rising or falling edge of external input.

Input level TTL.

Minimum input pulse width required 50ns (characteristic).

Video trigger ModeS: Selectable on rising or falling signal edge.

Video level.

Pre-qualification mode level and time settings

SynC output ModeS: Generate a sync pulse:

• Off

• Begin measurement

• Start tune

• Ready acquire

• Start acquire

• End acquire

• End measurement

SynC output polArity SeleCt: Sync out is on the falling or rising edge.
SynC output: TTL level. Minimum pulse width 200ns (characteristic).
eVen SeCond CloCk input: External even second clock (TTL).
eVen SeCond CloCk output: External even second clock (TTL).

1. Bus trigger and bus arm available only in channel power list mode.

GENERaL SPEcIFIcaTIoNS

power: 100VAC to 240VAC; 50/60 Hz (automatically detected); 150VA max.
Ce eMC CoMpliAnCe: EU Directive 89/336/EEC; EN 61326-1.
Ce SAFety CoMpliAnCe: CE; EU Directive 73/23/EEC, EN 61010-1.
CAlibrAtion: Annual calibration cycle in system.
enVironMent (for indoor use only):

18° to 23°C specified operating, unless otherwise noted.
0° to 50°C operating survival, non-specified operation.
–25°C to 65°C non-operating (AC power off) storage.
Altitude: Maximum 2000 meters above sea level.
Cooling: Forced air top, bottom, and side intakes and rear exhaust. For proper cooling

in a rack, use Keithley Instruments 2910-RMK Rack Mount Kit.

digitAl inputS/outputS: 4 bits, TTL-compatible.
interFACeS: IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status

model topology.
lAn: 10/100BT Ethernet, RJ45, LXI Class C, no auto MDIX.
iVi-CoM.
uSb: USB full speed.
rF in/tg out: Type N connector.

MeChAniCAl VibrAtion And ShoCk:

MIL-PRF-2880 CL3 random vibration, 3 axes.
Sine-Sweep test for resonances, 3 axes.
MIL-STD-810F 516.5 paragraph, 4.5.7 procedure VI bench drop MIL-PRF-2880 CL3

random vibration, 3 axes.

generAl MeChAniCAl ChArACteriStiCS:

height: 3U (133mm) (5.25˝).
width: Half-rack (213mm) (8.4˝).
depth: 464mm (18.25˝).
weight: 7.5kg (16.5 lbs.).

wArrAnty: 1 year.

2820

RF Vector Signal Analyzer

400MHz to 4GHz or 6GHz

SPEcIFIcaTIoN NoTES
Specifications describe the instrument’s warranted performance. Typical and characteristic values
are not warranted, but provide additional information regarding performance that you should
expect from the Model 2820 and are provided to assist in application of the Model 2820.

SpeCiFiCAtionS (warranted performance):
Specifications indicate performance that is warranted. All units are warranted to meet these
performance specifications under the following conditions:
• Ambient operating temperature of 18° to 28°C, unless otherwise noted.
• After a warm-up time of 30 minutes and self calibration at ambient temperature.

typiCAl (mean + 3 standard deviations):
Typical

indicates performance that units will meet under the following conditions:

• Ambient operating temperature of 23°C, unless otherwise noted.
• After a warm-up time of 30 minutes and self calibration at ambient temperature.
This performance is not warranted.

ChArACteriStiC (mean or expected value):
Characteristic

indicates performance that a unit would be expected to exhibit under the

following conditions:
• Ambient operating temperature of 23°C, unless otherwise noted.
• After a warm-up time of 30 minutes and self calibration at ambient temperature.
This performance is not warranted.