3 dbm, At 0 dbm, Ordering information – Atec Rohde-Schwarz-SML03 User Manual
Page 10
www.rohde-schwarz.com
R&S® is a registered trademark of Rohde&Schwarz GmbH&Co. KG · Trade names are trademarks of the owners
PD 0758.1664.32 · ¸SML · Version 06.00 · August 2004 · Data without tolerance limits is not binding · Subject to change
ARI/RDS
ARI identification
ARI BK
RDS traffic program
RDS traffic announcement
RDS data set
Maximum data length
functions (directly selectable by menu
or remote control)
selection of traffic announcement
identification (DK) or area identifica-
tion (BK),
OFF, DK, BK, DK + BK
selection of standardized area
identification A to F
traffic program off/on
traffic announcement off/on
selection of RDS data set 1 to 5
64 kByte, can be loaded via IEC60625
or RS-232-C interface
Analog modulation inputs L, R
Input impedance
Input voltage V
p
for selected
deviation (nominal value)
2 × BNC
600
Ω or 100 kΩ
1 V
Digital modulation input S/P DIF
Input impedance
Input voltage V
pp
BNC
75
Ω
1 V (400 mV to 5 V)
Sweep
digital in discrete steps
RF sweep, AF sweep
Operating modes
Sweep range
Step width (lin)
Step width (log)
automatic, single-shot, manually or ex-
ternally triggered, linear or logarithmic
user-selectable
user-selectable
0.01% to 100%
Level sweep
Operating modes
Sweep range
Step width (log)
automatic, single-shot, manually or
externally
triggered, logarithmic
user-selectable
user-selectable
Step time
Resolution
10 ms to 1 s
0.1 ms
Memory for device settings
Storable settings
100
Remote control
System
IEC60625 (IEEE 488) and RS-232-C
Command set
SCPI 1995.0
Connector
Amphenol, 24-pin and 9-pin
IEC/IEEE-bus address
0 to 30
Interface functions
SH1, AH1, T6, L4, SR1, RL1, PP1, DC1,
DT1, C0
General data
Temperature loading
Specs complied with between
0
°C and 55°C;
meets EN60068-2-1: 1995-03 and
EN60068-2-2: 1994-08
Storage temperature range
−40°C to +70°C
Climatic resistance
Damp heat
95% relative humidity at
+25
°C/+40°C cyclically;
meets EN60068-2-30: 2000-02
Mechanical resistance
Vibration, sinusoidal
5 Hz to 150 Hz, max. 2 g at 55 Hz,
max. 0.5 g between 55 Hz and
150 Hz, meets EN60068-2-6: 1996-05,
EN61010-1 and MIL-T-28800D, class 5
Vibration, random
10 Hz to 300 Hz,
acceleration 1.2 g (rms)
Shock
40 g shock spectrum,
meets MIL-STD-810E and
MIL-T-28800D, class 3/5
Electromagnetic compatibility
meets EN61000-6-3 and EN61000-6-4
(EMC directive of EU)
Susceptibility to radiated interference 10 V/m
Power supply
100 V to 120 V (AC), 50 Hz to 400 Hz,
200 V to 240 V (AC), 50 Hz to 60 Hz,
autoranging, max. 200 VA
Safety
meets EN61010-1, UL 3111-1,
CSA 22.2 No. 1010-1
Dimensions (W × H × D)
427 mm × 88 mm × 450 mm
Weight
8.5 kg when fully equipped
1)
With option ¸SML-B3 only for f >20 MHz.
2)
With Attenuator Mode Auto.
3)
–140 dBm to +11 dBm at f
≤5 MHz, f >3 GHz for ¸SML02 and ¸SML03.
4)
Temperature range 20°C to 30°C.
5)
With Attenuator Mode Fixed.
6)
With Attenuator Mode Auto, f
≥100 kHz.
7)
+5 dBm to +11 dBm at f
≤5 MHz, f >3 GHz.
8)
With option ¸SML-B3 only for f >10 MHz.
9)
Values in brackets apply to wide modulation bandwidth.
10)
Generator without preemphasis, receiver with deemphasis.
Ordering information
Signal Generator
¸SML01
¸SML02
¸SML03
1090.3000.11
1090.3000.12
1090.3000.13
Accessories supplied
Power cable, user manual
Options
Reference Oscillator OCXO
Pulse Modulator
Stereo/RDS Coder
Rear Connectors for AF, RF
¸SML-B1
¸SML-B3
¸SML-B5
¸SML-B19
1090.5790.02
1090.5403.02
1)
1147.8805.02
1090.5303.02
Recommended extras
Service Kit
19“ Rack Adapter
Transport Bag
Service Manual, Modules
¸SML-Z2
¸ZZA-211
¸ZZT-214
1090.5203.02
1096.3260.00
1109.5119.00
1090.3123.24
1)
Factory-fitted only.