beautypg.com

Signal sources, High-performance rf signal generators (cont.) – Atec HP-Agilent-8662A-8663A User Manual

Page 3

background image

8662A
8663A
11721A

High-Performance RF Signal Generators (cont.)

Signal Sources

Phase Modulation (Option 002)

Maximum Peak Phase Deviation: From ± 25˚ for f

c

between

120 and 160 MHz to ± 400˚ for f

c

between 1280 and 2560 MHz

Maximum Rate: From 10 kHz for f

c

between 0.15 and 10 MHz to

10 MHz for f

c

between 640 and 2560 MHz

Phase Deviation Resolution: 1˚ (0.1

f

c

<640 MHz);

2˚ (640

f

c

<1280 MHz); 4˚ (1280

f

c

<2560 MHz)

Phase Modulation Distortion: 10% at maximum rate

Biphase Modulation (BPSK)

Biphase modulation is available on the standard 8663A for f

c

less than 640

MHz and available for all f

c

with Option 002.

Deviation: ± 90˚
Carrier Null when Modulated with 1 MHz, 50% Duty Cycle

Square Wave: >25 dBc

Modulation Input Required: TTL positive true. The internal modulation
oscillator can be used for 50% duty-cycle modulation. External input is
on rear panel.

Pulse Modulation

1

Pulse On/Off Ratio: >80 dB (50 to 2560 MHz)
Pulse Rise/Fall Time: <250 ns (50 to 120 MHz); < 800 ns (120 to 640 MHz);
<100 ns (f

c

640 MHz)

Pulse Repetition Frequency (50% duty cycle):

Internal: 10 Hz to 99.9 kHz
External: 10 Hz to 2 MHz, 50 MHz < f

c

<640 MHz;

10 Hz to 5 MHz, f

c

>640 MHz

Internal Modulation Oscillator

Rates: 10 Hz to 99.9 kHz
Frequency Resolution: 3 digits
Frequency Accuracy: Same as reference oscillator
Output Level (available on rear panel): 1 V peak into 600

Output Impedance: 600

Flatness (referenced to 1 kHz): < ± 1%
Distortion: < 1%

Other 8662A and 8663A Information

Remote Programming: The GPIB interface is standard on the 8662A and
8663A signal generators. All functions controlled from the front panel,
with the exception of the line switch, are programmable with the same
accuracy and resolution as in manual mode.
Operating Temperature Range: 0˚ C to + 55˚ C
Leakage: Meets radiated and conducted limits of MIL-STD-461A
methods RE02 and CE03 as well as BVDE 0871
Power Requirements: 115 (90 to 126) V or 230 (198 to 252) V;
48 to 66 Hz; 450 VA max
Size:

8662A: 178 mm H x 425 mm W x 572 mm D (7 in x 16.75 in x 22.5 in)
8663A: 178 mm H x 425 mm W x 642 mm D (7 in x 16.75 in x 25.3 in)

Note: depth includes front panel depth of 45 mm (1.75 in).

Weight: 8662A: net, 30 kg (65.5 lb); shipping, 36 kg (80 lb)
8663A: net, 33.8 (74 lb); shipping, 40 kg (88 lb)

Key Literature

Synthesized Signal Generator 10 kHz to 1280 MHz

Technical Data, p/n 5953-8402

Synthesized Signal Generator 100 kHz to 2.56 GHz

Technical Data, p/n 5953-8376

Ordering Information

8662A 1280 MHz Signal Generator

2

Opt 001 RF Connectors on Rear Panel Only
Opt 003 Specified SSB Phase Noise for 640 MHz Output
Opt 907 Front Handle Kit (5062-3990)

4

Opt 908 Rack Flange Kit (5062-3978)

4

Opt 909 Rack Flange Kit w/Front Handles(5062-3984)

4

Opt 910 Two Sets of Operating and Service Manuals

4

(08662-90069)
Opt W30 Extended Repair Service (see page 70)
Opt W32 Calibration Service (see page 70

11721A

8663A 2560 MHz Signal Generator

2

Opt 001 RF Connectors on Rear Panel Only
Opt 002 Wideband Linear Phase Modulation
Opt 003 Specified SSB Phase Noise for 640 MHz Output
Opt 907 Front Handle Kit (5061-9690)

4

Opt 908 Rack Flange Kit (5061-9678)

4

Opt 909 Rack Flange Kit w/Front Handles (5061-9684)

4

Opt 910 Additional Operation and Calibration Manual

4

(08663-90069) and Service Manuals (08663-90071)
Opt 915 Add Service Manual (08663-90071)

4

Opt W30 Extended Repair Service (see page 70)
Opt W32 Calibration Service (see page 70)

11714A Service Support Kit (required for servicing
8662A/8663A)

1

Pulse modulation is available for f

c

< 50 MHz but is unspecified.

2

GPIB cables not supplied. For description and price, see page 568.

4 Indicates QuickShip availability.

11721A Frequency Doubler

The 11721A doubler is an ideal accessory for extending the usable
frequency range of signal generators, frequency synthesizers, or
other signal sources. Operating on input frequencies of 50 MHz to
1300 MHz, it provides a doubled output in the range of 100 MHz to
2600 MHz. The 11721A will work well with any RF source with an
output in the range of 50 to 1300 MHz.

The 50

passive circuit of the 11721A offers low conversion

loss, low spurious, and excellent flatness over its entire frequency
range when operated above +10 dBm.

11721A Specifications

Input Frequency Range: 50 to 1300 MHz
Output Frequency Range: 100 to 2600 MHz
Conversion Loss (+13 dBm input, 50 to 1280 MHz): < 15 dB
Spurious Referenced to Desired Output Frequency f:
(+ 13 dBm input with harmonics <–50 dBc, 50 to 1280 MHz):
f/2, –15 dB; 3f/2, –15 dB
Input SWR: 1.5 typical
Input/Output Impedance: 50

nominal

Operating Temperature Range: 0˚ to 50˚ C
Connectors: Input, type-N male; output, type-N female
Size: 20.5 mm H x 30 mm W x 161 mm D (.19 in x 1.19 in x 6.38 in)
Weight: Net, .02 kg (0.5 lb); shipping, 0.4 kg (1 lb)

Ordering Information

11721A Frequency Doubler

Opt W30 Extended Repair Service (see page 567)

5

213

Internet URL www.agilent.com/find/tmc

Product & Order Info See page 604

01SEC023_IN.QXD 9/6/00 11:38 AM Page 213