beautypg.com

Atec NHResearch-4100 User Manual

Page 2

background image

13

NH Research, Incorporated

16601 Hale Avenue, Irvine, California 92606

Tel: 949-474-3900 • Fax: 949-474-7062

E-mail: [email protected]

MODEL 4100/4110 SPECIFICATIONS

PROTECTION CIRCUITS

Over-power: Maximum power is limited through monitoring of heatsink

temperature. Programmable to lower limits.

Over-current: All modes limited to current set in constant current mode
Over-voltage: Output protected for transients over 450 V
Reverse-voltage: Reverse polarity diodes on outputs to short UUT
Over-temperature: Monitor of heatsink temperature

SUPPLEMENTAL CHARACTERISTICS

Size (HWD): Single-slot (six available) of S300 mainframe chassis

(5

1

/

4

x 19 x 22-inch)

Weight: Module: 6.6 lbs, Chassis: 21 lb
Operating Temperature: 0 to 50° C

All specifications apply for 27° C ± 5° C

Power (loads only): 115 VAC ± 10%, 208 to 264 VAC, 47 to 63 Hz

ADDITIONAL FEATURES

Mode Switching: Glitch-free transition when changing between current,

voltage, power, and resistance modes

Isolation: ± 500 VDC between input and chassis ground
Remote Sense: 2 VDC max drop between sense and load input
OVPS Relay: DPST, 5A, isolated control
Self-test: Complete built-in hardware self-test of all major functions

including each output transistor

Calibration: Closed cover, all adjustments done in software and stored in

an on-board EEPROM

MODAL OPERATION

4100

4110

Constant Current

Ranges:

0 to 60 A, 6 A, and 660 mA

0 to 60A, 6A, and 660 mA

Accuracy:

60 mA, 30 mA, 660 µA + 2 µA/V 60 mA, 30 mA, 660 µA + 2 µA/V

Resolution:

0.025% FS

0.025% FS

Constant Voltage

Ranges:

0.5 to 120 V, to 450 V

0.2 to 8V, 0.75 to 80 V

Accuracy:

0.1% FS

0.1% FS

Resolution:

0.025% FS

0.025% FS

Constant Resistance

Ranges:

0.035 to 10 k

0.12 to 10 k

Accuracy:

0.5% Set

0.5% Set

Resolution:

0.15% Set

0.15% Set

Constant Power

Ranges:

0 to 400 W, 40 W

0 to 400 W, 40 W

Accuracy:

1% Set

1% Set

Resolution:

0.1% FS

0.1% FS

Transient Generator

Pulse

Current Settings:

3

3

Total Period:

40 µsec to 1 sec (25 kHz to 1 Hz) 40 µsec to 1 sec (25 kHz to 1 Hz)

Delay Between

Settings:

20 µsec to 1 sec

20 µsec to 1 sec

Resolution:

10 µsec

10 µsec

Accuracy:

1% + 5 µsec

1% + 5 µsec

Modes:

Single burst, continuous

Single burst, continuous

Rise/Fall Time

Range:

10 µsec to 1 sec (10% to 90%)

10 µsec to 1 sec (10% to 90%)

Resolution:

2 µsec

2 µsec

Accuracy:

1% of setting + 3 µsec

1% of setting + 3 µsec

External Modulation

Bandwidth:

DC to 25 kHz

DC to 25 kHz

Programming Voltage:

0 to 5V

0 to 5V

Accuracy:

5% FS

5% FS

Short-circuit

Resistance:

0.035

@ 60 A

0.012

@ 60 A

READBACK
INSTRUMENTATION

4100

4110

DC Current

Ranges:

0 to 60 A, 6 A, 660 mA

0 to 60 A, 6 A, 660 mA

Accuracy:

15 mA, 5m A, 0.3 mA, + 2 µA/V 15 mA, 5 mA, 0.3 mA, + 2µA/V

Resolution:

250, 25, 2.5 µA

250, 25, 2.5 µA

DC Voltage

Ranges:

0 to 120, 450 V

0 to 8, 80 V

Accuracy:

0.025% FS

0.025% FS

Resolution:

500, 1.8 µV

33, 330 µV

Power

Ranges:

0 to 500, 50 W

0 to 500, 50 W

Accuracy:

1%

1%

Resolution:

± 30 mW

± 30 mW

CURRENT

S C A L E C H A N G E

0

10

20

30

40

50

60

LOG VOLTS

VOLTS

0

5

10

50

100

500

MODEL 4100

MODEL 4110

PROGRAMMABLE FEATURES

Current: 0 to 60A
Voltage:

4100

4110

2.1* to 450 VDC

0.7* to 80 VDC

* Full current operation. See graph below for current derating at lower voltages.

Power: 300 W
Modes: Constant current, constant voltage, constant resistance, constant power,

transient generator, short-circuit and external analog modulation

Control: GPIB (IEEE 488.2) or RS485