beautypg.com

Atec Instek-PEL-300 User Manual

Page 2

background image

Setting Accuracy

±(0.5% + 100mA) for 6.02A ~ 60A
±(0.1% + 10mA) for 0.602A ~ 6A
±(0.1% + 1mA) for 6mA ~ 0.6A

Setting Resolution

20mA for 6.02A ~ 60A
2mA for 0.602A ~ 6A
0.2mA for 6mA ~ 0.6A

CR MODE

Operating Range

50m ~ 1k

Setting Accuracy

±(5% + 1 ) for 100.33

~ 1k

±(5% + 100m ) for 10.033 ~ 100

±(5% + 10m

) for 1.0033 ~ 10

±(5% + 1m ) for 50m ~ 1

Setting Resolution

330m for 100.33 ~ 1k

33m for 10.033 ~ 100

3.3m for 1.0033 ~ 10

0.33m for 50m ~ 1

DYNAMIC OPERATION CC MODE ONLY

Frequency Range

1Hz ~1kHz

Resolution

5Hz for 100Hz ~1kHz
0.5Hz for 10Hz ~100Hz
0.05Hz for 1Hz ~10Hz

Duty Range

10% ~ 90%

Resolution

1%

VOLTAGE READBACK

Accuracy

±(0.1% + 60mV)

Resolution

20mV

CURRENT READBACK

Accuracy

±(0.5% + 100mA) for 6.02A ~ 60A
±(0.1% + 10mA) for 0.602A ~ 6A
±(0.1% + 1mA) for 6mA ~ 0.6A

Resolution

20mA for 6.02A ~ 60A
2mA for 0.602A ~ 6A
0.2mA for 6mA ~ 0.6A

MEMORY

100 sets

TIMER

Range

1 second ~ 999 minutes 59 seconds

Resolution

1 second

Function

for load on

Page 2 of 3