Ordering information, Insulation resistance, G round bond – Atec Quadtech-G32 User Manual
Page 2: Line leakage current, Common feature s, Guardian 6100
Guardian 6100
AC Output Vo l t a g e :
Range: 50V to 5000V AC, 1V re s o l u t i o n
Frequency: 50 or 60 Hz Programmable
Waveform: Sinusoidal
Regulation: <(1% +5V) at Rated Load
Voltage Display:
Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
AC Current Display:
Total current
Range: 1µA to 40mA AC
Resolution: 0.001mA
Accuracy: +/-(1% + 5cnt)
High/Low Limit Test:
1µA to 40mA AC
Low limit can be turned OFF
Arc Detection:
Programmable Level and OFF, 0.001mA/step
DC Output Vo l t a g e :
Range: 50V to 6000V DC, 1V re s o l u t i o n
Regulation: <(1% +5V) at Rated Load
Voltage Display:
Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
DC Current Display:
Range: 0.1µA to 20mA DC
Resolution: 0.0001mA
Accuracy: +/-(1% or reading + 5cnt)
High/Low Limit Test:
0.1µA to 20mA DC
Low limit can be turned OFF
Arc Detection:
Programmable Level and OFF, 0.001mA/step
Insulation Resistance:
Range: 100k
Ω
- 50G
Ω
Accuracy: +/- 5% to +/- 15% depending upon volt-
age and resistance
Voltage Range:
50V to 1000V DC
Voltage Accuracy:
+/- (1% of setting + 5V)
High/Low Limit Test:
100k
Ω
- 50G
Ω
Low limit can be turned OFF
IR Test Delay:
0.3 to 99.9 seconds Programmable in 0.1 second
steps
G round Bond:
Output Curre n t :
Range: 1.0 to 30.0A AC, setting 0.1A/step
Accuracy: +/- (1% of setting + 0.3A)
Display:
+/- (1% of reading + 3 counts)
Frequency:
50 or 60Hz Selectable
No Load Voltage:
6 to 15 V Programmable
Resistance:
Range: 0.1m
Ω
- 510.0m
Ω
, 4 digits
Accuracy: +/-(1% of reading + 3 counts)
Resolution: 0.1m
Ω
Hi Limit:
0.1m
Ω
to 510m
Ω
Offset Function:
0 to 100m
Ω
offset, user selectable
Test Time:
0.5 - 999sec (+/-20ms)
Line Leakage Current:
Input Voltage:
Range: 0V to 300V AC, 50 or 60Hz.
Line Voltage Meter:
0 - 300VAC, + / -(1% of reading + 5cnts)
Line Current Meter:
0 - 10A, + / -(5% of re a d i n g )
AC Current Display:
R a n g e
R e s o l u t i o n
A c c u r a c y
0.0001mA - 0.59mA
0.0002mA
+/-(2% + 5cnts)
0.6mA - 9.999mA
0.003mA
+/-(2% + 5cnts)
Current Trip Limits:
0.1µA to 9.999mA, 1µA Resolution
Measuring Circuit:
5 Types of Human Body Models in accordance with
UL544 NP, UL544P, UL1563, UL2601-1,IEC601-1,
IEC 950, UL1950, UL3101 Standards
Measurement Modes:
Normal, Reverse and Single Fault
Max. DUT Current:
10A
Common Feature s :
AC/DC Test Time:
Ramp: 0.1 to 99.9s (+/-20ms)
Test: 0.1 to 999s (+/-20ms) and Continuous
Remote Control:
Inputs: Start,Stop
Outputs: Pass/Fail/Under Test
Connector: Terminal Strip and 9 pin D Series
Test Setups:
50 Test Setups with 10 Steps each
Rear Connectors:
Input: Binding Posts (line & neutral)
Isolation transformer (not included)
is required for proper operation
(10A, 300V max)
Output: High Voltage Sockets for connection
to 115V corded adapter, G30 (included)
Front Panel Lockout:
Password
Safety Feature s :
Fast Cutoff (<0.4ms) and Fast Discharg e
Miscellaneous:
Continuous Voltage on Fail
Scanner Delay: 0.1 to 99.9, 0.1s/step
Indication:
Pass/fail lights, audible sound
Buzzer Level:
1,2,3 and Off
Interface:
IEEE-488 Standard, RS-232 Optional
Dimensions:
(w x h x d):16.875x5.625x20.50in
(421.9x140.6x512.5mm)
Weight:
53 lbs (24kg) - Net, 60 lbs (27kg) Shipping
Environmental:
Operating: 0 to + 40° C,
Humidity: <75%
Storage: - 20 to + 70° C
Warm-up Time: 1minute
Power:
• 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 500W max
For more detailed specifications, visit
www.quadtech.com •
For more information about special purchase, rent & lease options, call
1-800-253-1230
•
Fax 1-508-485-0295
•
Intl. 1-508-229-0806
5 Clock Tower Place, 210 East, Maynard, MA 01754
PN030089/A1
G u a rdian 6100 Production Safety Analyzer
Includes:
Guardian 6000 Electrical Safety Analyzer
6000-04 Line Leakage Current Scan Unit
150528
Instruction Manual
S02
HV Lead Set, 1m
G15
Ground Continuity Lead Set
G30
Corded Product
Adapter (115V)
Power Cable
Calibration Certificate Traceable to NIST
Optional Accessories
Calibration Data
S05
Foot Switch
G26
RS232 Interface
G27
Rack Mount Flanges
G28
Printer Interface
(replaces IEEE 488)
G31
Isolation Transformer
(500VA)
G32
Isolation Transformer
(1000VA)
Ordering Information