beautypg.com

Ordering information – Atec Fluke-772-773 User Manual

Page 2

background image

2 Fluke Corporation 772 and 773 Milliamp Process Clamp Meters

Functions

mA measure

w/jaw

mA measure

In circuit

mA source

mA sim

Loop power

24 V

DCV source

0-10 V

DCV

measure

0-30 V

Scaled mA

output to mA

input

mA in/out

771

772

773

Functional specifications

Function

Resolution and range

Accuracy

Notes

771, 772, 773

mA measurement

0 to 20.99 mA

0.2 % + 5 counts

Measured by clamp

21.0 mA to 100.0 mA

1 % + 5 counts

772 and 773

mA measurement

0 to 24.00 mA

0.2 % + 2 counts

Measured in series with test

jacks

772 and 773

mA source

0 to 24.00 mA

0.2 % + 2 counts

Maximum mA drive: 24 mA into

1,000 ohms

772 and 773

mA simulate

0 to 24.00 mA

0.2 % + 2 counts

Maximum voltage 50 V dc

773

Voltage source

0 to 10.00 V dc

0.2 % + 2 counts

2 mA maximum drive current

773

Voltage measure

0 to 30.00 V dc

0.2 % + 2 counts

General specifications 772 and 773

(see www.fluke.com/771 for more information on the Fluke 771 specifications)

Influence of earth’s field

< 0.12 mA

Battery

(4) 1.5 V, Alkaline, IEC LR6

Working hours

12 hours @ 12 mA source into 500 ohms

Size (HxWxL)

772, 773: 41.3 mm x 76 mm x 248 mm (1.625 in x 3 in x 9.75 in)

771: 59 mm x 38 mm x 212 mm, (2.32 in x 1.5 in x 8.35 in)

Weight

772, 773: 415 g (14 oz) 771: 260 g, (9.1 oz)

Operating temperature

-10 °C to 50 ºC

Storage temperature

-25 °C to 60 °C

Operating humidity

< 90 % @ < 30 °C; < 75 % @ 30 ~55 °C

Operating altitude

0 ~ 2,000 m

Storage altitude

None

IP rating

IP 40

Vibration requirements

Random 2 g, 5 Hz to 500 Hz

Drop test

Passes 1 meter drop test (except the jaw)

EMI, RFI, EMC

Meets applicable requirements in EN61326-1 Note: For current

measurement w/jaw, add 1 mA to specification for EMC field strengths

of 1 V/m up to 3 V/m

Temperature coefficients

0.1(/°C x specified accuracy for temperature < 18 °C or > 28 °C)

Warranty

Three-years, one-year on mA clamp assembly and cable

Ordering information

Fluke-772

Milliamp Process Clamp Meter

Fluke-773

Milliamp Process Clamp Meter

Included accessories

Soft carrying case, test leads, alligator clips, hanging

strap and user’s manual.

772

773

Fluke Corporation

PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.

PO Box 1186, 5602 BD

Eindhoven, The Netherlands
For more information call:

In the U.S.A. (800) 443-5853 or

Fax (425) 446-5116

In Europe/M-East/Africa +31 (0) 40 2675 200 or

Fax +31 (0) 40 2675 222

In Canada (800)-36-FLUKE or

Fax (905) 890-6866

From other countries +1 (425) 446-5500 or

Fax +1 (425) 446-5116

Web access: http://www.fluke.com

©2009 Fluke Corporation. Specifications subject to

change without notice. Printed in U.S.A.

5/2009 3442293 D-EN-N Rev B

Modification of this document is not permitted

without written permission from Fluke Corporation.

Fluke. Keeping your world

up and running.

®