Specifications, Ordering information – Atec Narda-HF3061 User Manual
Page 2
![background image](/manuals/319269/2/background.png)
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.
Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone: +49 (0) 7121-97 32-777
Fax: +49 (0) 7121-97 32-790
E-Mail: [email protected]
www.narda-sts.de
Narda Safety Test Solutions
435 Moreland Road
Hauppauge, NY 11788, USA
Phone: +1 631 231-1700
Fax: +1 631 231-1711
E-Mail: [email protected]
www.narda-sts.com
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano), Italy
Phone: +39 02 2699871
Fax: +39 02 26998700
E-mail: [email protected]
www.narda-sts.it
SPECIFICATIONS
a
Probe HF3061
Magnetic (H-)Field
Frequency range
(b)
300 kHz to 30 MHz
Type of frequency response
Flat
Measurement range
0.012 to 16 A/m (CW)
0.012 to 0.7 A/m (True RMS)
5.4 µW/cm² to 10 W/cm² (CW)
5.4 µW/cm² to 18 mW/cm² (True RMS)
Dynamic range
62 dB
CW damage level
35 A/m
46 W/cm²
Peak damage level
(c)
350 A/m
4.6 kW/cm²
Sensor type
Diode based system
Directivity
Isotropic (Tri-axial)
Readout mode / spatial assessment
3 separate axes
UNCERTAINTY
Flatness of frequency response
(d)
Calibration uncertainty not included
0/-1 dB (500 to 800 kHz)
+0.1/ -0.5 dB (800 kHz to 30 MHz)
Calibration uncertainty
(e)
@ 0.59 mW/cm² (0.125 A/m)
±0.6 dB
Linearity
Referred to 0.59 mW/cm² (0.125 A/m)
±3 dB (0.017 to 0.033 A/m)
±1 dB (0.033 to 0.068 A/m)
±0.5 dB (0.068 to 3 A/m)
±1 dB (3 to 16 A/m)
±3 dB (10 to 40 µW/cm²)
±1 dB (40 to 175 µW/cm²)
±0.5 dB (175 µW/cm² to 340 mW/cm²)
±1 dB (0.34 to 10 W/cm²)
Isotropic response
(f)
±1 dB
Temperature response
+0.2/ -0.8 dB (±0.025 dB/K @ 10 to 50 °C )
GENERAL SPECIFICATIONS
Calibration frequencies
0.1/ 0.15/ 0.2/ 0.3/ 0.4/ 0.5/ 0.6/ 0.7/ 0.8/ 0.9 MHz
1/ 1.2/ 1.5/ 2/ 3/ 4/ 5/ 10/ 15/ 20/ 25/ 27.12/ 30 MHz
Recommended calibration interval
24 months
Temperature range
Operating
Non-operating (transport)
0 °C to +50 °C
-40 °C to +70 °C
Humidity
5 to 95 % RH @ ≤28 °C
≤26 g/m³ absolute humidity
Size
300 mm x 120 mm Ø
Weight
190 g
Compatibility NBM-500
series
meters
Country of origin
Germany
(a)
Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal
(b)
Cutoff frequency at approx. -3 dB
(c)
Pulse length 1µsec, duty cycle 1:100
(d)
Frequency response can be compensated for by the use of correction factors stored in the probe memory
(e)
Accuracy of the fields generated to calibrate the probes
(f)
Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe
ORDERING INFORMATION
Part number
Probe HF3061, H-field for NBM, 300 kHz - 30 MHz, isotropic
2402/05
NSTS 0510-E0237C
2 / 2
Subject to change