beautypg.com

Technical data, Electrical, Portaflow 220 technical datasheet – Atec Micronics-PF220 User Manual

Page 6: Technical data electrical

background image

Portaflow 220 Technical Datasheet

6

Portaflow 220 Technical Datasheet

(Issue 1.0)

Technical data

Electrical

General specification:

DSP measurement technique –

The PF220 uses ‘transit time’ measurement with a sampling resolution of 50 pico-
second, with a continuous signal level indication to the display

Flow velocity range (bi-directional) –

Minimum velocity 0.1m/s
Max velocity 20m/s

Flow measurement accuracy –

±0.5% to ±2% of flow reading for flow rate >0.2m/s and Pipe ID >75mm
±3% of flow reading for flow rate >0.2m/s and Pipe ID in range 13mm - 75mm
±6% of flow reading for flow rate < 0.2m/s

Flow velocity corrected for Reynolds number over entire velocity range

Repeatability –

±0.5% of measured value or ±0.02m/s whichever is the greater

Response time –

< 500ms depending on pipe diameter

Selectable flow units –

VELOCITY: m/sec, ft/sec

VOLUME: l/s, l/min, l/h, gal/min, gal/h, USgals/min, USgals/h, Barrel/h,

Barrel/day, m³/s, m³/min, m³/h

Selectable volume units –

l, gal, USgals, Barrel, m³

Total volume –

12 Digits - forward and reverse

Display languages –

The following operator languages can be selected from the PF220 set-up menu

English
French
German

Italian
Spanish
Portuguese

Russian
Norwegian
Dutch

Supply voltage:

Power supply charger:

Input voltage range

9–24Vdc

Manufacturer –

Model ECO-181WP12

Power consumption

10.5W

Input voltage range –

90–264Vac

Battery:

Input frequency range –

47–63Hz

Technology –

5-cell NiMH

Output voltage –

12Vdc

Capacity –

3.8AHr

Max. Output current –

1.5A

Operating time –

Typically 20 hours continuous with
backlight and 4-20mA output OFF

Approvals –

UL, CUL, TUV, CB & CE

Recharge time –

6.5 Hours

Service life –

>500 charge/discharge cycles