Specifications

Radiation detected

Beta

> 1 MeV

Gamma

> 25 keV

Operating ranges,

response time

0 μR/h to 500 μR/h (5 sec)

0 mR/h to 5 mR/h (2 sec)

0 mR/h to 50 mR/h (1.8 sec)

0 mR/h to 500 mR/h (1.8 sec)

0 R/h to 5 R/h (1.8 sec)

Accuracy

Within 10 % of readings between 10 % and 100 % of full scale indication on any range,

exclusive of energy response

Detector

Chamber

230 CC Pressurized ionization chamber to 8 atmospheres or 125 psi

Controls

ON/OFF and MODE

Automatic features

Auto-zeroing, auto-ranging, and auto-backlight

Warm-up time

Less than one minute for initial operation when the instrument is in temperature equilibrium

with the surrounding area, typical. (About four minutes for readings if less than 20 μR/h in a

10 μR/h or less background)

Display LCD analog/

digital with backlight

Analog

100 element bar graph 6.4 cm long. Bar graph is divided into 5 major

segments, each labeled with the appropriate value for the range of

the instrument

Digital

2.5 digit display is followed by a significant zero digit depending on

the operating range of the instrument. The units of measurement are

indicated on the display at all times. Digits are 6.4 mm (0.25 in) high.

Low battery and freeze indicators are also provided on the display

Modes

Integrate mode

Operates continuously 30 seconds after the instrument has been

turned on. Integration is performed even if the instrument is

displaying in mR/h or R/h

Freeze mode

Will place a tick mark on the bar graph display to hold on the peak

displayed value. The unit will continue to read and display current

radiation values

Environmental

Temperature range

-20 °C to 50 °C (-4 °F to 122 °F)

Relative humidity

0 % to 100 %

Geotropism

Negligible

Typical energy

dependence

16

Nitrogen gamma rays are 110 % to 120 % of indicated readings as determined at the

University of Lowell

Power requirements

Two 9 V alkaline, 200 hours operation

Dimensions (WxDxH)

10 cm x 20 cm x 15 cm (4 in x 8 in x 6 in)

Weight

1.07 kg (2.4 lb)

2 Fluke Biomedical 451P Pressurized µR Ion Chamber Survey Meter