Asm 310, universal voltage – Atec Pfeiffer-ASM-310 User Manual
Page 2
ASM 310, universal voltage
Oil free pumping system, roughing capacity 1.7 m
3
/h
Ultralight, only 21 kg (46 lbs) and easy to move
Large bright color touchscreen with graphics functionality
Intuitive and customizable menu
Integrated SD memory card for recording and downloading capacities
Dimensions
Technical Data
ASM 310, universal voltage
Backing pump
with oil free backing pump
Backing pump capacity
1.7 m
3
/h
Detectable gases
4
He,
3
He, H
2
Dimensions (L x W x H)
350 mm
|
13.78 inch
Flange (in)
DN 25 ISO-KF
I/O interface
Digital, Analog (mantissa, exponent)
Interface
RS-232
Maximum inlet test pressure
15 hPa
|
11.25 Torr
|
15 mbar
Minimum detectable leak rate for helium
(sniffing leak detection)
1 · 10
-8
Pa m
3
/s
|
7.5 · 10
-8
Torr l/s
|
1 · 10
-7
mbar l/s
Minimum detectable leak rate for helium
(vacuum leak detection)
5 · 10
-13
Pa m
3
/s
|
3.75 · 10
-12
Torr l/s
|
5 · 10
-12
mbar l/s
Noise level
< 45 dB (A)
Operating temperature
10-40 °C
|
50-104 °F
|
283.15-313.15 °K
Pumping speed for He
1.1 l/s
Response time (Sniffing leak detection)
< 1 s
Start-up time (20°C) with calibration
<3.5 min
Start-up time (20°C) without calibration
< 2 min
Supply
100-240 V, 50/60 Hz
Test method
Vacuum and sniffing leak detection
Type
Mobile
User interface
Graphic color touch screen (detachable)
Weight
21 kg
|
46.3 lb
Order number
ASM 310
BSAA0000MM9A
Leak Detection & Contamination Management / Leak Detectors / Portable
www.pfeiffer-vacuum.com