ASM 310, universal voltage

Oil free pumping system, roughing capacity 1.7 m

3

/h

Ultralight, only 21 kg (46 lbs) and easy to move

Large bright color touchscreen with graphics functionality

Intuitive and customizable menu

Integrated SD memory card for recording and downloading capacities

Dimensions

Technical Data

ASM 310, universal voltage

Backing pump

with oil free backing pump

Backing pump capacity

1.7 m

3

/h

Detectable gases

4

He,

3

He, H

2

Dimensions (L x W x H)

350 mm

|

13.78 inch

Flange (in)

DN 25 ISO-KF

I/O interface

Digital, Analog (mantissa, exponent)

Interface

RS-232

Maximum inlet test pressure

15 hPa

|

11.25 Torr

|

15 mbar

Minimum detectable leak rate for helium
(sniffing leak detection)

1 · 10

-8

Pa m

3

/s

|

7.5 · 10

-8

Torr l/s

|

1 · 10

-7

mbar l/s

Minimum detectable leak rate for helium
(vacuum leak detection)

5 · 10

-13

Pa m

3

/s

|

3.75 · 10

-12

Torr l/s

|

5 · 10

-12

mbar l/s

Noise level

< 45 dB (A)

Operating temperature

10-40 °C

|

50-104 °F

|

283.15-313.15 °K

Pumping speed for He

1.1 l/s

Response time (Sniffing leak detection)

< 1 s

Start-up time (20°C) with calibration

<3.5 min

Start-up time (20°C) without calibration

< 2 min

Supply

100-240 V, 50/60 Hz

Test method

Vacuum and sniffing leak detection

Type

Mobile

User interface

Graphic color touch screen (detachable)

Weight

21 kg

|

46.3 lb

Order number

ASM 310

BSAA0000MM9A

Leak Detection & Contamination Management / Leak Detectors / Portable

www.pfeiffer-vacuum.com