Technical specification, Ordering information – Atec Haefely-pim400-410-430 User Manual
Page 2
TECHNICAL
EXCELLENCE
EASY OPERATION
SAFETY & QUALITY
e.g. accurate, clean pulses
PIM400,410/PCD430
modules are fully compliant
to UL 1449 requirements.
n
System can be configured
for UL and ANSI testing.
n
Impulse parameters
guaranteed at CDN output.
n
Automatic impulse type
switching during testing.
n
Highly accurate and
reliable semiconductor
discharge switch places the
pulse at the selected angle on
the selected phase.
UL 1449 requires
synchronization of the surge
impulse with the test phase.
Haefelys’ Surge Platform
solution switches the
synchronization source
automatically between
phases.
UL 1449 test system can
be assembled into a rack or
used as separate bench
standing units. The
advantage of this
architecture is that the test
equipment can be tailored
to fit in your application,
not the other way around.
A built-in Micro processor is
the user interface and
controls:
n Test/Program setup
n System configuration
n Peak Voltage monitoring
n Test Log file printing
Operating the UL 1449
system, from WinFEAT&R
Software opens additional
possibilities such as:
n EUT failed limits control
n Automatic Test
reporting in MS Word™
format
n EUT monitoring during
testing
n
Pre programmed test
routines.
Designed as an
expandable system,
Modules can be used
either
n in a desktop
configuration or
n rack mounted with
other system modules.
All EUT connections are
on the unit front panels
providing a clean
separation between test
instrument and EUT
connections.
A full range of accessories
is available to provide
safe and easy connections
with the test object.
The unit is designed
according to EMC and LV
directives.
UL 1449 system is
supplied with an
additional high-visibility
warning lamp.
Access to all operational
high voltage and AC
power circuits is safety
interlock protected.
All Haefely EMC products
carry a two year
manufacturers warranty.
TECHNICAL SPECIFICATION
PIM 400/410
PCD 430
Impulse Voltage
0.50kV ... 7.0 kV
EUT AC Voltage
690V (phase-phase)
Impulse Repetition
5 ... 10 seconds @ 6kV
EUT AC Current
32A per phase
Voltage Waveform
Risetime 1.2
µ
s Backtime 50
µ
s
Overcurrent protection
32A in all phases
Current Waveform
Risetime 8
µ
s Backtime 20
µ
s
Synchronisation
with coupling phase
Impulse Impedance
2 ohm & 12 ohm
ORDERING INFORMATION
P
SURGE 8000
Controller
249 900
PIM 400
Impulse Module
249 946
PIM 410
Impulse Module
249 947
PCD 430
Coupler/Decoupler
249 948
PIM 110
Impulse Module
249 903
WinFEAT&R
Software
249 700
Issue 01/02
Haefely Test AG
Business Unit EMC
Lehenmattstrasse 353, P.O Box
CH-4028 Basel
Switzerland
Phone: ++ 41 61 373 41 11
Fax:
++ 41 61 373 45 99
email: