Atec Haefely-pim110 User Manual
Page 2
Headquarters
Haefely Test AG
Lehenmattstrasse 353
CH-4052, Basel
Switzerland
+ 41 61 373 41 11
+ 41 61 373 45 99
Locate your local
sales representative at
www.haefelyEMC.com
North American Office
Hipotronics Inc.
Haefely EMC Division
1650 Route 22
Brewster, NY 10509
++1 845 279 3644 x264
++1 845 279 2467
Haefely has a policy of continuous product improvement. Therefore we reserve the right to change design and specification without notice.
PIM110_2006-03-13.doc
Technical Specifications
Impulse Voltage
0.20 – 7.8kV
±
10%
Impulse rise time
V= 0.5us
±
20% & I= < 1.0us
Source Impedance
12, 30 & 200 Ohms
Impulse frequency
100kHz
±
10% ( V & I)
Impulse Polarity
Positive and Negative
Voltage Monitor
BNC, (750/1)
Repetition @ Umax
5 seconds / 12 per minute
Current Monitor
BNC, (12 & 30
Ω
80A/V) 200
Ω
5A/V
Line Synchronisation
1° steps
±
1° @50Hz
Impulse Output
HV connector with safety circuit
Weights and Dimensions (W x H x D, net weight)
PIM 110
45 x 17 x 57 cm
44 lbs (25 kg)
PIM 110 Art. No. 249903 Scope of Supply
Qty. 1
PIM 110 Impulse Module
Qty. 1
HV DC Bus cable 1m
Qty. 1
Haefely Bus cable 0.5m
Qty. 1
Earth bonding cable 1m
Qty. 1
Earth bonding cable 0.25m
Qty. 1
Users Manual
Options and Accessories
PCD 100
Single phase CDN 16Aac & dc.
Automatic coupling path switching.
Two multiplexed high voltage
inputs.
PCD 130
Three phase CDN 32Aac & dc
Automatic coupling path switching.
Three multiplexed high voltage
inputs.
PCD 126
Coupling network for control and
data lines
DEC 7
Decoupling network for control and
data lines
PIM 100
Combination wave generator
according to
ANSI C62.41 & IEC 61000-4-5
PEX 400
Case for PIM modules.
ADAPTERS
Single phase input & output
adapters enable PCD 130 to be
used for single phase applications
WinFEAT&R
Control and reporting software.
Runs under windows 98, NT, ME,
2000, XP
Rack Mounting
Modules can be rack mounted for
greater mechanical stability and
mobility.
WinFEAT&R Control Window
PCD 130 Three phase power line CDN