Atec Montena-EMP25K-2-23 User Manual
Nemp radiated susceptibility test setup
montena technology - 1728 Rossens - Switzerland - phone + 41 26 411 84 84 - fax + 41 26 411 17 79
www.montena.com
e-mail: [email protected]
NEMP radiated susceptibility test setup
This test set-up can be used for the susceptibility testing of small pieces of equipment such as
printed boards, small subsystems, etc., according to MIL-STD-461 / RS105 (NEMP: nuclear
electromagnetic pulse). Associated with the generator EMP25K-2-23, the minimum field under
the line RL50-50 is 50 kV/m at full output. Even higher fields can be achieved for smaller
objects placed close to the beginning of the line. The line can also be used for other pulse or
CW tests.
The antenna is fixed on a stable and solid table which can be moved easily. The generator can
be controlled by a computer through RS232 and USB interface. We also provide all the
necessary accessories like electromagnetic field sensors, interconnecting cables, voltage
dividers for the measurement of the output of the generator, etc.
SPECIFICATIONS - Generator
Type
EMP25K-2-23
Standard
MIL-STD461-E / F, paragraph RS105
Peak voltage
typically ≤ +0.2 kV to ≥ +25 kV
Impedance
100 ohm
Rise time (10 – 90 %)
2.3 ns ± 0.5 ns
Pulse length on 100 ohm (50 – 50 %)
23 ns ± 5 ns
Output connector
HVM50K
Power rating
85 - 264 V
ac
/ 50 - 60 Hz / 150 VA
Dimensions
56 x 19 x 45 cm (L x H x W)
Weight
15 kg
SPECIFICATIONS - Line
Type
RL50-50
Structure
bounded wave line / TEM mode
Impedance
100 ohm
Maximum height of the EUT
17 cm (according to MIL-STD461 / RS105)
Input connector
HVM50K
Dimensions
200 x 55 x 80 cm (L x H x W)
Weight
25 kg