MODEL 8282-1 TRANSIENT PULSE GENERATOR

s

Adjustable pulse position on a.c. power lines
relates the transient susceptibility to the real
time aspects of digital circuitry served by
a.c. power.

s

Transients can be injected in synchronism with
repetitive circuit functions as required by
Method CS06 of MIL-STD-462.

s

Remote triggering of single or repetitive pulses
in terms of particular system characteristics.

s

The upper terminals of the PARALLEL pair and
the SERIES pair provide a positive-going spike
on the 5

ȖS and the 10 ȖS modes. These termi-

nals deliver a negative-going spike in the 0.15
ȖS mode. When the test plan requires both a
positive and a negative spike, it is necessary to
reverse the connections to the output termi-
nals of the Model 8282-1 Spike Generator.

SPECIFICATIONS

Spike Durations: Pushbutton selectable dura-
tions of 0.15

ȖS, 5.0 ȖS and 10.0 ȖS (DŽ20%)

to zero crossover, into 5.0 ohm resistive load.

Adjustable Peak Amplitude: Up to 600 volts
for 0.15

ȖS, 5.0 ȖS and 10.0 ȖS durations into five

ohm non-inductive load.

Internal Impedance: Less than 5.0 ohms for
0.15

ȖS, less than 2.0 ohms for 5.0 ȖS, less than

1.0 ohm for 10.0

ȖS.

Pulse Repetition Rate: Manually adjustable up to
50 p.p.s. for all pulse durations.

Pulse Shape: Ringing characteristic similar to
Figure 19 in MIL-STD-462 when connected to
non-inductive load.

Pulse Position: Adjustable from 0° to 360° on
50 Hz, 60 Hz or 400 Hz power lines.

External Sync Operation: Remotely triggerable
up to 50 p.p.s. for 0.15

ȖS, up to 1000 p.p.s. for 5.0

and 10.0

ȖS.

Amplitude Display: Panel meter is analog LED
display of peak amplitude as it would be into
a five ohm resistive load.

Power Current in Series Injection Mode:
Handles up to 50 amperes of current at power
frequencies.

Power Requirements: 115 volts 60 Hz, 3.0
amperes. (230 volts 50 Hz, 1.5 ampere available.)

Size: 12.25" wide , 8.7" high 13" deep. (311 mm x
211 mm x 330 mm.)

Weight: 30 pounds.

USEFUL ACCESSORIES

Type 7115-2 High Voltage Pulse Transformer.
Plugs into SERIES output terminals to provide
transient levels up to 15 KV, peak, into Type
7510-1
Spark Gap assembly for static discharge tests.

Type 7512-1 Spike Injection Probe*

Type 7519-1 Pulse Shaping Network*

Type 7541-1 Spike Receptor Probe*

Type 8282-150 Transient Pulse Transformer.
Plugs into SERIES output terminals. Handles up
to 150 amperes through the secondary for high
current test samples.

Type 8525-1 Non-Inductive Five Ohm Load

Type 8527-2 Transient Pulse Transformer. Plugs
into SERIES output terminals to provide spike
levels up to 2 KV, peak, into 50 ohms when using
the ten microsecond function.

Type 8908-1 Transient Pulse Transformer. Plugs
into series output to provide up to 600 V spike
into 50 ohms when using 5

ȖS or 10 ȖS function.

Type 9007-1 Transient Pulse Transformer. Plugs
into SERIES output terminals to provide spike
levels up to 1200 V into 50 ohms when using the
0.15 microsecond function.

* See Application Note on Cable Induced

Transients

PARALLEL INJECTION ON D.C. LINE

SERIES INJECTION ON A.C. LINE