Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano) - ITALY
Phone.: +39 02 269987-1
Fax: +39 02 269987-00
PMM 7010
SPECIFICATIONS
Frequency range
Resolution
Uncertainty
150 kHz ÷ 1 GHz
10 Hz
< 3 ppm
RF input
VSWR @ input attenuator = 10 dB
@ input attenuator = 0 dB
Attenuator
Pulse limiter
Z
in
50 Ω,
N fem.
< 1.2
< 1.2; < 1.4 above 30 MHz
0 ÷ 35 dB in 5 dB steps
0 ÷ 50 dB in 2 dB steps above 30 MHz
Internal, active for conducted measurements range (0,15 ÷ 30 MHz)
Maximum input level
(without damage)
Sinusoidal voltage
Pulsed spectral density
140 dB
µ
V (2 W)
100 dB
µ
V/MHz
IF bandwidth (RBW)
3, 10, 30, 100, 300 kHz:
-6 dB
9, 120 kHz:
compliant to CISPR 16-1-1
Noise level
(Hold time = 1 s)
0.15 ÷ 30 MHz (9 kHz RBW)
< 14 dB
u
V (QP) < 7 dB
u
V (AVG)
30 ÷ 1000 MHz (120 kHz RBW)
< 14 dB
u
V (QP) < 7 dB
u
V (AVG)
Spurious
(Peak det., Hold time = 10 ms)
< 20 dBµV
CISPR detectors
Simultaneous:
Peak, Quasi-Peak, Average, C-Average, RMS, RMS-Average
Smart Detector function
Hold time
According to CISPR 16-1-1
0.2 ms ÷ 120 s
PES – PMM Emission Suite PC software
Creation of scan tables, limits, factors
Reporting, export/import of measurements and settings
Trace comparison, marker, zoom, peaks finder
Operating modes
Sweep: automatic scanning receiver
Manual: manual tuning receiver
Analyzer: spectrum analyzer mode
Measuring units
dBm, dBµV, dBµA, dBpW, dBµV/m, dBµA/m, dBpT
Display dynamic: 80 ÷ 200 dB
Uncertainty
± 1.0 dB (150 kHz ÷ 1 GHz S/N > 20 dB)
Compliance to CISPR 16-1-1
Full compliance of Quasi-Peak detector for PRF ≥ 20 Hz
Demodulation
AM – FM with internal loudspeaker and volume control
I/O
RS-232, USB, remote control for other PMM LISN
Operating temperature
-5° ÷+ 45°C
Power supply
12 V DC, 0.8A; universal AC adapter
Internal LISN
Frequency range
Rated current
Maximum line voltage
Line frequency
CISPR-equivalent circuit
EUT socket
150 kHz ÷ 30 MHz
16A
250 V AC – 350 V DC
DC ÷ 60 Hz
50 Ω // (5 Ω + 50 uH)
SCHUKO 2P+E
Dimensions
235x105x335 mm
Weight
5.0 kg
08
/1
2 S
pe
cific
ati
on
s
su
bj
ect
to
c
ha
ng
e
s
w
ith
ou
t
pr
io
r
n
o
tice
Ordering information
7010
EMI receiver 150 kHz - 1 GHz
Includes:
- AC adapter mod. 9010/AC
- PC software PMM Emission Suite
- Calibration certificate
- USB, RS-232 cables
- User manual
OPTIONS
- 9010/RAV
CISPR RMS-AVG detector
- 9010/RMA
19” rack-mounting kit