Atec Fischer-F-16-1 User Manual
F-16-1 current probe
The F-16-1 is used to measure conducted emissions from 10kHz to
50MHz. Its useable frequency range is 10Hz to 70MHz. The basic
F-16 probe has a transfer impedance of 4
Ω from 30kHz to 50MHz.
The F-16-1 with the power absorbing terminator has a constant
transfer impedance of .5
Ω from 20kHz to 50MHz. All
configurations of the F-16 probe can handle 400 amperes of
primary power line currents from DC to 60Hz without saturating or
impacting the transfer impedance of the probe.
F-16-1
Current Probe
Fischer Custom Communications, Inc.
F-16-1 Specifications
Frequency:
100 Hz – 70 MHz
Connector:
Type-N
Internal diameter:
32mm DC to 60 Hz: 400 amperes
External diameter
: 94mm
400 MHz:
400
amperes
Height:
64mm RF(CW): 50
amperes
Z
t
Ω
1
:
0.5
Pulse
2
:
300
amperes
dB
Ω
1
:
−6
1: Probe calibrated with 50Ω+j0Ω Load Impedance
2:
Depends
upon
the
pulse width and pulse rep. Rate.
20603 Earl St., Torrance, CA 90503 · Phone: 310-303-3300 Fax 310-371-6268 · email: [email protected]
© Fischer Custom Communications, Inc.
072803DS-F-16-1
All rights reserved
• Printed in USA