background image

Resolution
Sensors
Measurement mode
Sampling rate

Digital filtering
Internal battery
Operation time

Recharging time

Internal data memory

Communication
Fiber optic connector
Fiber optic length

Fiber optic to PC

PC software

Operating temperature
Temperature reading resolution
Battery voltage reading resolution
Dimensions

Weight
Probe holder tip

EP-600

EP-601

EP-602

EP-603

Frequency Range

100 kHz - 9,25 GHz

10 kHz - 9,25 GHz

5 kHz - 9,25 GHz

300 kHz - 18 GHz

Flatness

Frequency correction OFF
1 - 150 MHz: 0,8 dB
0,5 - 6000 MHz: 1,6 dB
0,3 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,3 - 7500 MHz: 0,4 dB

Frequency correction OFF
0,1 - 150 MHz: 0,4 dB
0,05 - 6000 MHz: 1,6 dB
0,03 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,05 - 7500 MHz: 0,4 dB

Frequency correction OFF
0,1 - 150 MHz: 0,4 dB
0,05 - 6000 MHz: 1,6 dB
0,03 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,05 - 7500 MHz: 0,4 dB

Frequency correction OFF
3 - 8200 MHz: 1,4 dB
1 - 12000 MHz: 2,4 dB
0,6 - 1800 MHz: 3,8 dB
Frequency correction ON, typical
0,3 - 18000 MHz: 0,4 dB

Dynamic Range

0,14 - 140 V/m
60 dB single range

0,5 - 500 V/m
60 dB single range

1,5 - 1500 V/m
60 dB single range

0,17 - 170 V/m
60 dB single range

Overload

300 V/m

1000 V/m

3000 V/m

350 V/m

0,01 V/m
Six monopoles
X-Y-Z axis simultaneous sampling
22 to 0,03 Samples/s depending on
filter settings
Low-pass 2,3 to 28 Hz; 80 dB notch 50/60 Hz
3V - 5 mA/h, Li-Mn rechargeable
80 hours @ 0,4 S/s, 28 Hz filter
60 hours @ 5 S/s, 28Hz filter
48 hours for full operation time
Partial charging allowed, e.g.:
8 hrs rech. = 12 hrs operation @ 5 S/s
1 hr rech. = 5 hrs operation @ 0,4 S/s
Serial nr. - Calibration factors & date
Firmware version
Bidirectional fiber optic link
HFBR-0500
10 m, standard
20/40 m options
Fiber optic to RS232 converter
RS232 to USB converter
Displays field, temperature, battery
Sampling, filters, averaging, frequency setting
Manual data logging as text file
-10 °C + 50 °C
0,1 °C
10 mV
53 mm overall
(body: 17 mm dia.; sensor, 17 mm each)
25 g, incl. 1 m fiber optic pigtail
1/4 20 UNC female

EP-600, EP-601, EP-602, EP-603

USB/RS232 Electric Field Probes

Ordering Information:

EP-600 Field probe 100 kHz - 9,25 GHz 0,14 - 140 V/m

EP-601 Field probe 10 kHz - 9,25 GHz 0,5 - 500 V/m

EP-602 Field probe 5 kHz - 9,25 GHz 1,5 - 1500 V/m

EP-603 Field probe 300 kHz - 18 GHz 0,17 - 170 V/m

Supplied with:

• 10 m fiber optic cable

• Optical/RS232 adapter + RS232/USB adapter

• PC Utility WIN-EP600

• Charging fixture, A/C charger, tripod mounting tip

• Standard calibration certificate

Options:

• FO-EP600/20 Fiber optic cable, 20 m

• FO-EP600/40 Fiber optic cable, 40 m

• Accreditated Certificate

Narda Safety Test Solutions Srl

Via Leonardo da Vinci, 21/23

20090 Segrate (Milano) - ITALY

Phone: +39 02 2699871

Fax: +39 02 26998700

E-Mail: [email protected]

Internet: www.narda-sts.it

EP600-FEN-30200 - Specifications subject to changes without prior notice

COMMON SPECIFICATIONS

Main specifications

(for full information please refer to the User Manual)