Specifications, Ordering information – Atec Narda-HF0191 User Manual

Page 2

background image



® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.

NSTS 0209-E0238B-0.0

2 / 2

Subject to change

Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone: +49 (0) 7121-97 32-777
Fax: +49 (0) 7121-97 32-790
E-Mail: [email protected]
www.narda-sts.de

Narda Safety Test Solutions
435 Moreland Road
Hauppauge, NY 11788, USA
Phone: +1 631 231-1700
Fax: +1 631 231-1711
E-Mail: [email protected]
www.narda-sts.com

Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano), Italy
Phone: +39 02 2699871
Fax: +39 02 26998700
E-mail: [email protected]
www.narda-sts.it

SPECIFICATIONS

a

Probe HF0191

Magnetic (H-)Field

Frequency range

(b)

27 MHz to 1 GHz

Type of frequency response

Flat

Measurement range

0.018 to 16 A/m (CW)
0.018 to 1 A/m (True RMS)

12 µW/cm² to 10 W/cm² (CW)
12 µW/cm² to 38 mW/cm² (True RMS)

Dynamic range

59 dB

CW damage level

20 A/m

15 W/cm²

Peak damage level

(c)

200 A/m

1.5 kW/cm²

Sensor type

Diode based system

Directivity

Isotropic (Tri-axial)

Readout mode / spatial assessment

3 separate axes

UNCERTAINTY

Flatness of frequency response

(d)

Calibration uncertainty not included

±0.7 dB (50 to 80 MHz)
±0.5 dB (80 to 250 MHz)
±0.8 dB ( 250 to 1000 MHz)

Calibration uncertainty

(e)

@ 2 mW/cm² (0.23 A/m)

±0.5 dB (27 to 700 MHz)
±1.5 dB (> 700 MHz)

Linearity

Referred to 2 mW/cm² (0.23 A/m)

±3 dB (0.026 to 0.05 A/m)
±1 dB (0.05 to 0.1 A/m)
±0.5 dB (0.1 to 3 A/m)
±1 dB (3 to 16 A/m)

±3 dB (25 to 100 µW/cm²)
±1 dB (100 to 380 µW/cm²)
±0.5 dB (0.38 to 340 mW/cm²)
±1 dB (0.34 to 10 W/cm²)

Isotropic response

(f)

±1 dB

Temperature response

+0.5/-0.8 dB (±0.025 dB/K @ 10 to 50 °C)

GENERAL SPECIFICATIONS

Calibration frequencies

10/ 15/ 20/ 27.12/ 30/ 35/ 40/ 50/ 60/ 70/ 80/ 90/ 100/ 120/ 150/ 180/ 200/ 250/ 300/ 400/ 433/
500/ 600/ 700/ 800/ 900/ 1000 MHz

Recommended calibration interval

24 months

Temperature range
Operating
Non-operating (transport)


-10 °C to +50 °C
-40 °C to +70 °C

Humidity

5 to 95 % RH @ ≤28 °C

≤26 g/m³ absolute humidity

Size

318 mm x 66 mm Ø

Weight

90 g

Compatibility NBM-500

series

meters

Country of origin

Germany


(a)

Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal

(b)

Cutoff frequency at approx. -3 dB

(c)

Pulse length 1µsec, duty cycle 1:100

(d)

Frequency response can be compensated for by the use of correction factors stored in the probe memory

(e)

Accuracy of the fields generated to calibrate the probes

(f)

Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe

ORDERING INFORMATION

Part number

Probe HF0191, H-field for NBM, 27 MHz – 1 GHz, isotropic

2402/06