Specifications, Ordering information – Atec Narda-HF0191 User Manual
Page 2
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.
NSTS 0209-E0238B-0.0
2 / 2
Subject to change
Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone: +49 (0) 7121-97 32-777
Fax: +49 (0) 7121-97 32-790
E-Mail: [email protected]
www.narda-sts.de
Narda Safety Test Solutions
435 Moreland Road
Hauppauge, NY 11788, USA
Phone: +1 631 231-1700
Fax: +1 631 231-1711
E-Mail: [email protected]
www.narda-sts.com
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano), Italy
Phone: +39 02 2699871
Fax: +39 02 26998700
E-mail: [email protected]
www.narda-sts.it
SPECIFICATIONS
a
Probe HF0191
Magnetic (H-)Field
Frequency range
(b)
27 MHz to 1 GHz
Type of frequency response
Flat
Measurement range
0.018 to 16 A/m (CW)
0.018 to 1 A/m (True RMS)
12 µW/cm² to 10 W/cm² (CW)
12 µW/cm² to 38 mW/cm² (True RMS)
Dynamic range
59 dB
CW damage level
20 A/m
15 W/cm²
Peak damage level
(c)
200 A/m
1.5 kW/cm²
Sensor type
Diode based system
Directivity
Isotropic (Tri-axial)
Readout mode / spatial assessment
3 separate axes
UNCERTAINTY
Flatness of frequency response
(d)
Calibration uncertainty not included
±0.7 dB (50 to 80 MHz)
±0.5 dB (80 to 250 MHz)
±0.8 dB ( 250 to 1000 MHz)
Calibration uncertainty
(e)
@ 2 mW/cm² (0.23 A/m)
±0.5 dB (27 to 700 MHz)
±1.5 dB (> 700 MHz)
Linearity
Referred to 2 mW/cm² (0.23 A/m)
±3 dB (0.026 to 0.05 A/m)
±1 dB (0.05 to 0.1 A/m)
±0.5 dB (0.1 to 3 A/m)
±1 dB (3 to 16 A/m)
±3 dB (25 to 100 µW/cm²)
±1 dB (100 to 380 µW/cm²)
±0.5 dB (0.38 to 340 mW/cm²)
±1 dB (0.34 to 10 W/cm²)
Isotropic response
(f)
±1 dB
Temperature response
+0.5/-0.8 dB (±0.025 dB/K @ 10 to 50 °C)
GENERAL SPECIFICATIONS
Calibration frequencies
10/ 15/ 20/ 27.12/ 30/ 35/ 40/ 50/ 60/ 70/ 80/ 90/ 100/ 120/ 150/ 180/ 200/ 250/ 300/ 400/ 433/
500/ 600/ 700/ 800/ 900/ 1000 MHz
Recommended calibration interval
24 months
Temperature range
Operating
Non-operating (transport)
-10 °C to +50 °C
-40 °C to +70 °C
Humidity
5 to 95 % RH @ ≤28 °C
≤26 g/m³ absolute humidity
Size
318 mm x 66 mm Ø
Weight
90 g
Compatibility NBM-500
series
meters
Country of origin
Germany
(a)
Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal
(b)
Cutoff frequency at approx. -3 dB
(c)
Pulse length 1µsec, duty cycle 1:100
(d)
Frequency response can be compensated for by the use of correction factors stored in the probe memory
(e)
Accuracy of the fields generated to calibrate the probes
(f)
Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe
ORDERING INFORMATION
Part number
Probe HF0191, H-field for NBM, 27 MHz – 1 GHz, isotropic
2402/06