beautypg.com

Specifications, Ordering information – Atec Narda-EF0391 User Manual

Page 2

background image



® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.

NSTS 0209-E0233B-0.0

2 / 2

Subject to change

Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone: +49 (0) 7121-97 32-777
Fax: +49 (0) 7121-97 32-790
E-Mail: [email protected]
www.narda-sts.de

Narda Safety Test Solutions
435 Moreland Road
Hauppauge, NY 11788, USA
Phone: +1 631 231-1700
Fax: +1 631 231-1711
E-Mail: [email protected]
www.narda-sts.com

Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano), Italy
Phone: +39 02 2699871
Fax: +39 02 26998700
E-mail: [email protected]
www.narda-sts.it

SPECIFICATIONS

a

Probe EF0391

Electric (E-)Field

Frequency range

(b)

100 kHz to 3 GHz

Type of frequency response

Flat

Measurement range

0.2 to 320 V/m (CW)
0.2 to 10 V/m (True RMS)

10 nW/cm² to 27 mW/cm² (CW)
10 nW/cm² to 0.027 mW/cm² (True RMS)

Dynamic range

64 dB

CW damage level

800 V/m

170 mW/cm²

Peak damage level

(c)

8 kV/m

17 W/cm²

Sensor type

Diode based system

Directivity

Isotropic (Tri-axial)

Readout mode / spatial assessment

3 separate axes

UNCERTAINTY
Flatness of frequency response

(d)

Calibration uncertainty not included

±1 dB (1 MHz to 1 GHz)
±1.25 dB (1 GHz to 2.45 GHz)

Calibration uncertainty

(e)

@ 0.01 mW/cm² (6.14 V/m)

±1 dB (<400 MHz)
±1.5 dB (400 MHz to 1.8 GHz)
±1 dB (≥ 1.8 GHz)

Linearity

Referred to 0.01 mW/cm² (6.14 V/m)

±0.5 dB (1.2 to 200 V/m)
±0.7 dB (200 to 320 V/m)

±0.5 dB (0.00038 to 10.6 mW/cm²)
±0.7 dB (10.6 to 27 mW/cm²)

Isotropic response

(f)

±1 dB

Temperature response

+0.2/ -1 dB (±0.025 dB/K)

GENERAL SPECIFICATIONS

Calibration frequencies

0.1/ 0.2/ 0.3/ 1/ 3/ 10/ 27.12 MHz
0.1/ 0.2/ 0.3/ 0.5/ 0.75/ 1/ 1.8/ 2.45/ 2.7/ 3 GHz

Recommended calibration interval

24 months

Temperature range
Operating
Non-operating (transport)


0 °C to +50 °C
-40 °C to +70 °C

Humidity

5 to 95 % RH @ ≤25 °C

≤23 g/m³ absolute humidity

Size

318 mm x 66 mm Ø

Weight

90 g

Compatibility NBM-500

series

meters

Country of origin

Germany


(a)

Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal

(b)

Cutoff frequency at approx. -3 dB

(c)

Pulse length 1µsec, duty cycle 1:100

(d)

Frequency response can be compensated for by the use of correction factors stored in the probe memory

(e)

Accuracy of the fields generated to calibrate the probes

(f)

Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe


ORDERING INFORMATION

Part number

Probe EF0391, E-field for NBM, 100 kHz – 3 GHz, isotropic

2402/01