2 Fluke Corporation Fluke 1555 and 1550C Insulation Resistance Testers

Fluke Corporation

PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.

PO Box 1186, 5602 BD

Eindhoven, The Netherlands
For more information call:

In the U.S.A. (800) 443-5853 or

Fax (425) 446-5116

In Europe/M-East/Africa +31 (0) 40 2675 200 or

Fax +31 (0) 40 2675 222

In Canada (800)-36-FLUKE or

Fax (905) 890-6866

From other countries +1 (425) 446-5500 or

Fax +1 (425) 446-5116

Web access: http://www.fluke.com

©2005-2010 Fluke Corporation.

Specifications subject to change without notice.

Printed in U.S.A. 10/2010 1629685G D-EN-N

Modification of this document is not permitted

without written permission from Fluke Corporation.

Fluke.

Keeping your world

up and running.

Ordering

information

1550C

5 kV Insulation Tester

1555

10 kV Insulation Tester

1550C/Kit 5 kV Insulation Tester Kit

1555/Kit 10 kV Insulation Tester Kit

Optional accessories

TL1550EXT 25 foot extended

test lead set

Included accessories

Test Cables with Alligator Clips

(red, black, green)
Infrared adapter with interface cable
FlukeView Forms Basic CD-ROM
AC Power Cord
Soft Carrying Case (base models only)
English Manual
Users Manual on CD-ROM
Quick Reference Card
Software License Agreement
Registration Card
FlukeView Forms Installation Guide
USB-IR Cable Installation Guide
IP67 Hard Case (kit only)
Certificate of Calibration (kit only)
Ruggedized Alligator Clips

(kit and 1555 only)

Software specification

s

Fluke ViewForms basic software

requires a PC running Windows 2000,

Windows XP and Windows Vista.

Specifications

Electrical specifications

The tester’s accuracy is specified for one year after calibration at operating

temperatures of 0 °C to 35 °C. For operating temperatures outside the range

(-20 °C to 0 °C and 35 °C to 50 °C), add ± .25 % per °C, except on the 20 %

bands add ± 1 % per °C.

Insulation resistance measurement

Test voltage (dc)

Range

Accuracy (± reading)

250 V

< 200 kΩ

200 kΩ to 5 GΩ

5 GΩ to 50 GΩ

> 50 GΩ

unspecified

5 %

20 %

unspecified

500 V

< 200 kΩ

200 kΩ to 10 GΩ

10 GΩ to 100 GΩ

> 100 GΩ

unspecified

5 %

20 %

unspecified

1000 V

< 200 kΩ

200 kΩ to 20 GΩ

20 GΩ to 200 GΩ

> 200 GΩ

unspecified

5 %

20 %

unspecified

2500 V

< 200 kΩ

200 kΩ to 50 GΩ

50 GΩ to 500 GΩ

> 500 GΩ

unspecified

5 %

20 %

unspecified

5000 V

< 200 kΩ

200 kΩ to 100 GΩ

100 GΩ to 1 TΩ

> 1 TΩ

unspecified

5 %

20 %

unspecified

10000 V (1555 Only)

< 200 kΩ

200 kΩ to 200 GΩ

200 GΩ to 2 TΩ

> 2 TΩ

unspecified

5 %

20 %

unspecified

Bar graph range

0 to 1 TΩ

Insulation test voltage accuracy

-0 %, +10 % at 1 mA load current

Induced ac mains current rejection

2 mA maximum

Charging rate for capacitive load

5 seconds per μF

Discharge rate for capacitive load

1.5 s/μF

Range

Accuracy

Leakage current

measurement

1 nA to 2 mA

± (5 % + 2 nA)

Capacitance measurement

0.01 uF to 15.00 μF

± (15 % rdg + 0.03 μF)

Timer

Range

Resolution

0 to 99 minutes

Setting: 1 minute

Indication: 1 second

Live circuit warning

Warning range

Voltage accuracy

30 V to 660 V ac/dc, 50/60 Hz

± (15 % + 2 V)

General specifications

Display

75 mm x 105 mm

Power

12 V lead-acid rechargeable battery, Yuasa NP2.8-12

Charger Input (AC)

85 V to 250 V ac 50/60 Hz 20 VA

This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal

(ground pin) is not connected internally. The extra pin is for added plug retention only.

Dimensions

269 mm x 277 mm x 160 mm (10.6 in x 10.9 in x 6.3 in)

Weight

3.7 kg (8.2 lb)

Temperature (operating)

-20 ºC to 50 ºC (-4 ºF to 122 ºF)

Temperature (storage)

-20 ºC to 65 ºC (-4 ºF to 149 ºF)

Humidity

80 % to 31 ºC decreasing linearly to 50 % at 50 ºC

Altitude

2000 m

Enclosure sealing

IP40

Input overload protection

1000 V ac

Electromagnetic compatibility

EN 61326

Certifications

Safety compliance

EN 61010, EN 61557 Parts 1 and 2, IEC 61010-1,

CAT III 1000 V, CAT IV 600 V

Polution degree

2