ORDERING INFORMATION

Item (Qty)

Order No.

10kV Insulation Tester

MEG10-01

Complete with

3m Lead set

8101-181

AVO Download Manager

6111-422

User Guide on CD-ROM

6172-785

Item (Qty)

Order No.

RS232 download cable

25955-025

Quick Start Guide (English)

6172-765

Quick Start Guide (French)

6172-784

Warranty card

SPECifications

Test Voltages (d.c.)
500 V, 1000 V, 2500 V, 5000 V, 10000 V

Test Voltage Accuracy
±5% on 100 M

load and above

Insulation resistance ranges
Digital:
10 k

to 50 G

@ 500 V

10 k

to 100 G

@ 1000 V

10 k

to 250 G

@ 2500 V

10 k

to 500 G

@ 5000V

10 k

to 500 G

@ 10000 V

Analogue:
100 k

to 1 T

@ all voltages

Accuracy (at 23

°C)

±5% on 100 M

load

Basic Accuracy
±5% of reading 10 M

to 100 G

, all voltages

Current Range
0,01 nA to 999 µA

Accuracy (at 23

°C)

±5% +/- 1nA on 100 M

load.

Basic Accuracy
±5% of reading 10 M

to 100 G

all voltages

Guard
2% error guarding 500 k

leakage with 100 M

load.

Short circuit current
5mA nominal.

Voltmeter range
50 to 1000 V d.c. or a.c.

Accuracy (23

°C)

±2% ±1V.

Interference rejection
1 mA r.m.s. per kV to a maximum of 2 mA.

Capacitor discharge time
<2 s per µF to discharge from 10 kV to 50 V.

Timer
0:00 to 99:59 minutes. Instrument can be set to stop a test after a
time limit, or to run indefinitely.

Instrument Display
3 digit LCD Analogue/Digital

Power supply
Rechargeable, sealed lead-acid batteries (12 V, 4 Ah), or 95 V to
265 V 50/60 Hz supply.

Battery life
Typically 4 hrs continuous testing at 10 kV, increasing to
8 hours at 5 kV and lower. Remaining battery life indicated on
display.

Battery charging
Charging takes approximately 12 hours from ac supply in the
range 95 to 265 V.

Temperature range
Operating:-

-20

°C to +50°C (4°F to 122°F)

Storage:-

-25

°C to +65°C (-12°F to 149°F)

Humidity: 90% R.H. @ 40

°C (104°F)max.

Temperature coefficient
±0,1% per

°C over the test voltage over the temperature range 0°C

to +30

°C. (±0.06% per °F)

Instrument protection.
IP54 with case closed.

Safety
The instrument meets the requirements of EN61010-1:2001 for
operation at an altitude of 2000 m above sea level or less.

EMC
The instrument meets the requirements of EN61326-1:1998 for use
in areas other than domestic.

Dimensions
385mm x 265mm x 250mm (15 x 10 x 9 inches)

Weight
7 kg (15 lb) approx

MEG10-01

10 kV Insulation Tester

Designed in accordance with the requirements of IEEE43:
2000 the MEG10-01 is the ideal instrument for measuring
the insulation resistance of armature and field windings in
rotating machines rated 1hp (750 W) or greater. The
standard applies to synchronous, induction and d.c.
machines as well as synchronous condensers.

The Megger MEG10-01 also performs automatic
Polarisation Index tests. This test is routinely performed
on a variety of equipment and cables to assess the general
condition of the insulation before applying a potentially
destructive applied voltage test at voltages greater than
rated voltage.

UK

Archcliffe Road Dover
CT17 9EN England
T +44 (0) 1304 502101
F +44 (0) 1304 207342

UNITED STATES

4271 Bronze Way
Dallas TX 75237-1019 USA
T 800 723 2861 (USA only)
T +1 214 333 3201
F +1 214 331 7399

OTHER TECHNICAL SALES OFFICES

Norristown USA, Toronto CANADA,
Mumbai INDIA, Trappes FRANCE,
Sydney AUSTRALIA, Madrid SPAIN
and the Kingdom of BAHRAIN.

Registered to ISO 9001:2000 Reg no. Q 09290

Registered to ISO 14001 Reg no. EMS 61597

MEG10_01_DS_en_V13

www.megger.com
Megger is a registered trademark