TOS6200

Earth Continuity Tester

■ Time

Test time

Setting range

0.3 to 999 s Timer ON/OFF function is available.

Accuracy

±

(100ppm of setting + 20ms)

■ Environment

Installation

Indoors and the altitude is less than 2,000 m

Warranty range

Temperature

5

°

to 35

°

C

Humidity

20% to 80% R.H (non condensing)

Operating range

Temperature

0

°

to 40

°

C

Humidity

20% to 80% R.H (non condensing)

Storage range

Temperature

-20

°

to 70

°

C

Humidity

90% or less R.H (non condensing)

■ Power requirement

Allowable voltage range

100 V model : 85 to 132 V AC
100 V/200 V model : 85 to 132 V AC/170 to 250 V AC

Power consumption

At no load (READY)

100 V model : 70 VA or less
100 V/200 V model : 45 VA or less

At rated load

100 V model : 450 VA max.
100 V/200 V model : 330 VA max.

Allowable frequency range

45 Hz to 65 Hz

■ Insulation resistance

30M

min. (500 V dc), between AC line and chassis

■ Withstanding voltage

1350 V AC (1 second), between AC line and chassis

■ Earth continuity

25 A AC/0.1

max.

■ EMC

Complied with the following standards

IEC61362-1: 1997-03/A1:1998-05

Electrical Equipment for Measurement, Control and Laboratory
Use - EMC requirements
Radiated Emissions Class A
Conducted Emissions Class A

IEC61000-4-2:1995-01

Electro-static Discharge

/A1:1998-01
IEC61000-4-3:1995-02

Radiated, radio-frequency, electromagnetic field

IEC61000-4-4:1995-01

Electrical fast transient / Burst

IEC61000-4-5:1995-02

Surge

IEC61000-4-6:1996-04

Conducted disturbances

IEC61000-4-11:1994-06

Voltage dips, short interruptions and voltage
variations

Under following conditions

1. Used test leadwires (TL11-TOS) which is supplied.
2. Used the shielded cable which length is less than three meters when the SIGNAL
I/O is used.

■ Safety

Complied with the following standards

IEC61010-1: 1990-09/A2:1995-07

Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use TOS6200 is designed so that it is connected to a power supply
of overvoltage category II as Class I equipment in environment of pollution
degree 2.

■ Physical dimensions (maximum)

430(450)W ✕ 88(140)H ✕

270(345)Dmm

■ Weight

Approx. 9kg

■ Accessories

AC power cord

1 piece

Test leadwire TL11-TOS

1 set

Short bar

2 pieces (These are inserted between the OUTPUT and
SAMPLING terminals.)

AC power fuse

2 pieces (2, including one spare in the fuse holder)

Operation manual

1 copy

*1: Time limitation with respect to output

The heat radiation capacity at the output block of the tester is designed to be half the rated output,

accounting for size, weight, cost, and other factors. Always use the tester within the limitation
values given below. Use of the tester beyond these limits will cause the temperature of the output
block to rise excessively, potentially tripping the internal protection circuit. In this case, suspend

testing for approximately 30 minutes, then press the STOP switch. When temperatures fall to
normal levels, the tester will revert to ready status.

Output time limitation

Ambient temperature t (

°

C) Test current I (A)

Quiescent time

Maximum test duration

t

40

°

15

<

I

30

Equal to or greater than test duration

30 minutes or less

I

15

Not required

Continuous output possibl

■ Output block

Current setting range (*1)

3.0 to 30.0 A AC
(With respect to resistance resulting in output power of
the maximum rated Output or less and an output
terminal voltage of 5.4 V or less)

Resolution

0.1A

Accuracy

±

(1% of setting + 0.2A)

Maximum rated output

150 VA (at the output terminals)

Distortion factor

2% or less (with respect to 0.1

pure resistance load of

10 A or greater)

Frequency

50/60 Hz, sine wave (selectable)

Accuracy

±

200ppm

Open terminal voltage

6 Vrms or less

Output method

PWM switching method

■ Output ammeter

Measurement range

0.0 to 33.0 A AC

Resolution

0.1A

Accuracy

±

(1% of reading + 0.2A)

Response

Mean value response/rms value display
(response time: 200 ms)

Holding function

The current measured at the end of test is held during
the PASS or FAIL inteval

■ Output voltmeter

Measurement range

0.00 to 6.00 V AC

Resolution

0.01V

Accuracy

±

(1% of reading + 0.02A)

Response

Mean value response/rms value display
(response time: 200 ms)

Holding function

The voltage measured at the end of test is held during
the PASS or FAIL inteval

■ Ohmmeter

Measurement range

0.001 to 1.200

Resolution

0.001

Offset cancel function

0.000 to 1.200

(Offset ON/OFF function provided)

Accuracy

±

(2% of reading + 0.003

)

Holding function

The resistance measured at the end of test is held during
the PASS or FAIL inteval

■ Pass/fail judgement function

Judgement system

Window comparator system

If a resistance value equal to or greater than the upper

reference value is detected, a FAIL determination is returned.

If a resistance value equal to or less than the lower

reference value is detected, a FAIL determination is returned.

If a resistance value has been judged as FAIL, the
tester shuts off the output and generates a FAIL signal.

If the set time elapses without abnormalities, the

tester shuts off the output and generates a PASS signal.

Setting range for the upper reference value (UPPER)

0.001 to 1.200

Setting range for the lower reference value (LOWER)

0.001 to 1.200

Judgement accuracy

±

(2% of UPPER + 0.003

)

Calibration

Calibration is performed with the rms value of the sine
wave, using a pure resistance load.

LED

PASS

Lights for approximately 0.2 sec when the measured
value has been judged as PASS. It is lit continuously
when the PASS holding time is set to HOLD.

UPPER FAIL

Lights if a resistance value equal to or greater than the
upper reference value is detected and judged FAIL.

LOWER FAIL

Lights if the resistance value equal to or less than the
lower reference value is detected and judged FAIL.

Buzzer

The buzzer sounds for approximately 0.2 sec if the
measured value has been judged as PASS.

The buzzer sounds continuously under the following
condition:
The measured value has been judged as PASS when the
PASS holding time is set to HOLD.
The measured value has been judged as UPPER FAIL.
The measured value has been judged as LOWER FAIL.

The buzzer volume for FAIL or PASS judgment are
adjustable. Note that it cannot be adjusted individually
since setting is shared with the setting for PASS.

Specifications