Tos6200, Earth continuity tester, Specifications – Atec Kikusui-TOS6200 User Manual
Page 2
Page2
Copyright © 2002 KIKUSUI ELECTRONICS CORP. All rights reserved.
http://www.kikusui.co.jp/
TOS6200
Earth Continuity Tester
■ Time
Test time
Setting range
0.3 to 999 s Timer ON/OFF function is available.
Accuracy
±
(100ppm of setting + 20ms)
■ Environment
Installation
Indoors and the altitude is less than 2,000 m
Warranty range
Temperature
5
°
to 35
°
C
Humidity
20% to 80% R.H (non condensing)
Operating range
Temperature
0
°
to 40
°
C
Humidity
20% to 80% R.H (non condensing)
Storage range
Temperature
-20
°
to 70
°
C
Humidity
90% or less R.H (non condensing)
■ Power requirement
Allowable voltage range
100 V model : 85 to 132 V AC
100 V/200 V model : 85 to 132 V AC/170 to 250 V AC
Power consumption
At no load (READY)
100 V model : 70 VA or less
100 V/200 V model : 45 VA or less
At rated load
100 V model : 450 VA max.
100 V/200 V model : 330 VA max.
Allowable frequency range
45 Hz to 65 Hz
■ Insulation resistance
30M
Ω
min. (500 V dc), between AC line and chassis
■ Withstanding voltage
1350 V AC (1 second), between AC line and chassis
■ Earth continuity
25 A AC/0.1
Ω
max.
■ EMC
Complied with the following standards
IEC61362-1: 1997-03/A1:1998-05
Electrical Equipment for Measurement, Control and Laboratory
Use - EMC requirements
Radiated Emissions Class A
Conducted Emissions Class A
IEC61000-4-2:1995-01
Electro-static Discharge
/A1:1998-01
IEC61000-4-3:1995-02
Radiated, radio-frequency, electromagnetic field
IEC61000-4-4:1995-01
Electrical fast transient / Burst
IEC61000-4-5:1995-02
Surge
IEC61000-4-6:1996-04
Conducted disturbances
IEC61000-4-11:1994-06
Voltage dips, short interruptions and voltage
variations
Under following conditions
1. Used test leadwires (TL11-TOS) which is supplied.
2. Used the shielded cable which length is less than three meters when the SIGNAL
I/O is used.
■ Safety
Complied with the following standards
IEC61010-1: 1990-09/A2:1995-07
Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use TOS6200 is designed so that it is connected to a power supply
of overvoltage category II as Class I equipment in environment of pollution
degree 2.
■ Physical dimensions (maximum)
430(450)W ✕ 88(140)H ✕
270(345)Dmm
■ Weight
Approx. 9kg
■ Accessories
AC power cord
1 piece
Test leadwire TL11-TOS
1 set
Short bar
2 pieces (These are inserted between the OUTPUT and
SAMPLING terminals.)
AC power fuse
2 pieces (2, including one spare in the fuse holder)
Operation manual
1 copy
*1: Time limitation with respect to output
The heat radiation capacity at the output block of the tester is designed to be half the rated output,
accounting for size, weight, cost, and other factors. Always use the tester within the limitation
values given below. Use of the tester beyond these limits will cause the temperature of the output
block to rise excessively, potentially tripping the internal protection circuit. In this case, suspend
testing for approximately 30 minutes, then press the STOP switch. When temperatures fall to
normal levels, the tester will revert to ready status.
Output time limitation
Ambient temperature t (
°
C) Test current I (A)
Quiescent time
Maximum test duration
t
≤
40
°
15
<
I
≤
30
Equal to or greater than test duration
30 minutes or less
I
≤
15
Not required
Continuous output possibl
■ Output block
Current setting range (*1)
3.0 to 30.0 A AC
(With respect to resistance resulting in output power of
the maximum rated Output or less and an output
terminal voltage of 5.4 V or less)
Resolution
0.1A
Accuracy
±
(1% of setting + 0.2A)
Maximum rated output
150 VA (at the output terminals)
Distortion factor
2% or less (with respect to 0.1
Ω
pure resistance load of
10 A or greater)
Frequency
50/60 Hz, sine wave (selectable)
Accuracy
±
200ppm
Open terminal voltage
6 Vrms or less
Output method
PWM switching method
■ Output ammeter
Measurement range
0.0 to 33.0 A AC
Resolution
0.1A
Accuracy
±
(1% of reading + 0.2A)
Response
Mean value response/rms value display
(response time: 200 ms)
Holding function
The current measured at the end of test is held during
the PASS or FAIL inteval
■ Output voltmeter
Measurement range
0.00 to 6.00 V AC
Resolution
0.01V
Accuracy
±
(1% of reading + 0.02A)
Response
Mean value response/rms value display
(response time: 200 ms)
Holding function
The voltage measured at the end of test is held during
the PASS or FAIL inteval
■ Ohmmeter
Measurement range
0.001 to 1.200
Ω
Resolution
0.001
Ω
Offset cancel function
0.000 to 1.200
Ω
(Offset ON/OFF function provided)
Accuracy
±
(2% of reading + 0.003
Ω
)
Holding function
The resistance measured at the end of test is held during
the PASS or FAIL inteval
■ Pass/fail judgement function
Judgement system
Window comparator system
•
If a resistance value equal to or greater than the upper
reference value is detected, a FAIL determination is returned.
•
If a resistance value equal to or less than the lower
reference value is detected, a FAIL determination is returned.
•
If a resistance value has been judged as FAIL, the
tester shuts off the output and generates a FAIL signal.
•
If the set time elapses without abnormalities, the
tester shuts off the output and generates a PASS signal.
Setting range for the upper reference value (UPPER)
0.001 to 1.200
Ω
Setting range for the lower reference value (LOWER)
0.001 to 1.200
Ω
Judgement accuracy
±
(2% of UPPER + 0.003
Ω
)
Calibration
Calibration is performed with the rms value of the sine
wave, using a pure resistance load.
LED
PASS
Lights for approximately 0.2 sec when the measured
value has been judged as PASS. It is lit continuously
when the PASS holding time is set to HOLD.
UPPER FAIL
Lights if a resistance value equal to or greater than the
upper reference value is detected and judged FAIL.
LOWER FAIL
Lights if the resistance value equal to or less than the
lower reference value is detected and judged FAIL.
Buzzer
•
The buzzer sounds for approximately 0.2 sec if the
measured value has been judged as PASS.
•
The buzzer sounds continuously under the following
condition:
The measured value has been judged as PASS when the
PASS holding time is set to HOLD.
The measured value has been judged as UPPER FAIL.
The measured value has been judged as LOWER FAIL.
The buzzer volume for FAIL or PASS judgment are
adjustable. Note that it cannot be adjusted individually
since setting is shared with the setting for PASS.
Specifications