Atec Hipotronics-OC60D-A-TCDE-TCVDE User Manual
Page 2
European Contact
Locate your local
USA Contact
Haefely Test AG
Lehenmattstrasse 353
CH-4028 Basel
Switzerland
+ 41 61 373 4111
+ 41 61 373 4912
|
sales representative at
www.high-voltage-hubbell.com
Hipotronics Inc.
1650 Route 22
PO Box 414
Brewster, NY 10509 USA
+ 1 845 279 8091
+ 1 845 279 2467
|
Subject to technical modifications without notice.
OC60E-DS / 08.2003
Technical Specifications
Control Panel
General
Output Voltage
0 - 60 kV
Input Voltage - A
120 V, 60 Hz
Meter Accuracy
+ 2% of full scale
Input Voltage - B
220 V, 50/60 Hz
Weights and Dimensions (W x H x D, net weight, ship weight)
OC60D
16" x 13" x 15" (41 x 33 x 38 cm) 69 lbs (31 kg)
74 lbs (86 kg)
OC90D
30" x 12" x 17" (76 x 30 x 43 cm) 122 lbs (55 kg)
190 lbs (86 kg)
TCDE
6" x 4" x 3" (15 x 10 x 8 cm)
2 lbs (1 kg)
5 lbs (2 kg)
TCVDE
6" x 6" x 6" (15 x 15 x 15 cm)
5 lbs (2 kg)
10 lbs (5 kg)
TCVDE90 13" x 6" x 6" (33 x 15 x 15 cm)
11 lbs (5 kg)
20 lbs (9 kg)
OCCM-E
6" x 6" x 6" (15 x 15 x 15 cm)
4 lbs (2 kg)
8 lbs (4 kg)
Scope of Supply
Qty. 1
OC60D or OC90D Oil Tester
Qty. 1
Standard Line Cord, 7.5 ft (2.3 m)
Qty. 1
Operations Manual & Calibration Certificate
Ordering Information
System
Oil Tester, Digital 60 kV
OC60D-A, OC60D-B
Oil Tester, Digital 90 kV
OC90D-A, OC90D-B
Accessories
Digital Calibration kV meter and ramp rate meter. A 0.5
inch digital display, molded epoxy case, with 2% accuracy
at full scale. The adapter for the OC90 is included.
OCCM-E
Spare parts kit
SPK1-OC60D
SPK1-OC90D
The TCDE test cell is equipped with two one-inch diameter flat disc
electrodes; one is adjustable to provide the desired gap via a set-
screw. A standard 0.1 inch gap gauge is supplied. The TCDE is used
to test at 3000 v/s rate of rise according to ASTM D877. For use with
OC60’s.
TCDE
The TCVDE test cell is equipped with a motor-driven circulating system
0.08 inch and 0.04 inch gap gauges, and two VDE electrodes, one is
adjustable to provide the desired gap. The TCVDE is used to test
according to ASTM D1816. For use with OC60’s.
TCVDE
The TCIEC test cell is similar to the Model TCVDE cell but without a
motor. For use with OC60’s.
TCIEC
The TCCE90 test cell has VDE (mushroom) electrodes and a motor
driven circulating system for testing to ASTM D1816. For testing to
ASTM D877, disc electrodes (included) with 0.1 inch gap gauge may
be substituted. Only for use with OC90’s.
TCCE90
Test Cells
OCCM-E
TCDE ASTM D877
TCVDE ASTM D1816
TCCE90 ASTM D1816 and ASTM D877