Short circuit current
greater than 1 mA and less than 2 mA
Bar graph range
0 to 1 T
Ω
Insulation test voltage
-0 % to +10 % at 1 mA load current
accuracy
Induced ac mains
2 mA maximum
current rejection
Charging rate for
5 seconds per µF
capacitive load
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com/
©2005 Fluke Corporation. All rights reserved.
Trademarks are the property of their respective owners.
Printed in U.S.A. 7/2005 1629685 D-ENG-N Rev E
Fluke. Keeping your world
up and running.
Specifications
Electrical specifications
The meter’s accuracy is specified for one year
after calibration at operating temperatures of 0
°C to 35 °C. For operating temperatures outside
the range (-20 °C to 0 °C and 35 °C to 50 °C),
add +/- .25 % per °C, except on the 20 % bands
add +/- 1 % per °C.
Insulation resistance measurement
Test Voltage
Range
Accuracy
(dc)
(± reading)
250 V
200 k
Ω to 5 GΩ
5 %
5 G
Ω to 50 GΩ
20 %
500 V
200 k
Ω to 10 GΩ
5 %
10 G
Ω to 100 GΩ
20 %
1000 V
200 k
Ω to 20 GΩ
5 %
20 G
Ω to 200 GΩ
20 %
2500 V
200 k
Ω to 50 GΩ
5 %
50 G
Ω to 500 GΩ
20 %
5000 V
200 k
Ω to 100 GΩ
5 %
100 G
Ω to 1 TΩ
20 %
Timer
0 to 99 minutes; settable in 1 minute increments;
indicated to within 1 second
Ramp
0 % to 100 % of selected test voltage, or until
breakdown
Mechanical specifications
Dimensions (L x W x H)
170 mm x 242 mm x 330 mm
(6.7 in x 9.5 in x 13.0 in)
Weight (with battery)
4 kg (8.8 lbs)
Power
Battery type
12 volt, lead-acid, rechargeable
Charger input (AC)
85 to 250 V ac 50/60 Hz 50 VA
Battery life (typical)
Test voltages
Number of tests
Note: At higher temperatures, 250 V
4138
the battery requires more
500 V
3913
frequent charging.
1 kV
3462
2.5 kV
2043
5 kV
1000
Environmental specifications
Operating temperature
-20 °C to 50 °C (-4 °F to 122 °F)
Storage temperature
-20 °C to 65 °C (-4 °F to 149 °F)
Relative humidity
80 % at 31 °C decreasing linearly
to 50 % at 50 °C
Operating altitude
0 to 2000 meters
Enclosure sealing
IP-40
Shock, vibration
MIL-PRF-28800F, CLASS II
EMI immunity
EN 61326
Safety
EN 61010-1, CAT III, 600 V,
EN61557, PARTS 1 & 2
Input overload protection
600 V ac continuous
Software specifications
FlukeView
®
Forms Basic 1550B Software
requires a PC running Windows
®
95, 98, Me,
2000, NT4.0
Two-year warranty
Included accessories
Test leads and low leakage probes
Alligator clips
Interface adapter and cable
FlukeView
®
Forms Basic 1550B PC software
Line cord
Soft carrying case with
waterproof bottom
User’s manual CD with 12 languages
Getting Started Guide
Optional accessories
FVF-UG FlukeView
®
Forms Upgrade
to Version 3.0
TL1550EXT 25 foot extension leads
Range
Accuracy
Leakage current
1 nA to 2 mA
± (5 % + 2 nA)
Capacitance
0.01 uF to 15.00 µF
± (15 % rdg + 0.03 µF)
measurement
Live circuit
30 V to 600 V
± (5 % + 2 V)
indicator
ac/dc, 50/60 Hz