beautypg.com

Fluke-1550_datasheet, Specifications, Electrical specifications – Atec Fluke-1550 User Manual

Page 2: Mechanical specifications, Power, Environmental specifications, Software specifications, Included accessories, Optional accessories, Flukeview

background image

Short circuit current

greater than 1 mA and less than 2 mA

Bar graph range

0 to 1 T

Insulation test voltage

-0 % to +10 % at 1 mA load current

accuracy

Induced ac mains

2 mA maximum

current rejection

Charging rate for

5 seconds per µF

capacitive load

Fluke Corporation
PO Box 9090, Everett, WA USA 98206

Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands

For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com/

©2005 Fluke Corporation. All rights reserved.
Trademarks are the property of their respective owners.
Printed in U.S.A. 7/2005 1629685 D-ENG-N Rev E

Fluke. Keeping your world

up and running.

Specifications

Electrical specifications

The meter’s accuracy is specified for one year
after calibration at operating temperatures of 0
°C to 35 °C. For operating temperatures outside
the range (-20 °C to 0 °C and 35 °C to 50 °C),
add +/- .25 % per °C, except on the 20 % bands
add +/- 1 % per °C.

Insulation resistance measurement

Test Voltage

Range

Accuracy

(dc)

(± reading)

250 V

200 k

Ω to 5 GΩ

5 %

5 G

Ω to 50 GΩ

20 %

500 V

200 k

Ω to 10 GΩ

5 %

10 G

Ω to 100 GΩ

20 %

1000 V

200 k

Ω to 20 GΩ

5 %

20 G

Ω to 200 GΩ

20 %

2500 V

200 k

Ω to 50 GΩ

5 %

50 G

Ω to 500 GΩ

20 %

5000 V

200 k

Ω to 100 GΩ

5 %

100 G

Ω to 1 TΩ

20 %

Timer

0 to 99 minutes; settable in 1 minute increments;
indicated to within 1 second

Ramp

0 % to 100 % of selected test voltage, or until
breakdown

Mechanical specifications

Dimensions (L x W x H)

170 mm x 242 mm x 330 mm
(6.7 in x 9.5 in x 13.0 in)

Weight (with battery)

4 kg (8.8 lbs)

Power

Battery type

12 volt, lead-acid, rechargeable

Charger input (AC)

85 to 250 V ac 50/60 Hz 50 VA

Battery life (typical)

Test voltages

Number of tests

Note: At higher temperatures, 250 V

4138

the battery requires more

500 V

3913

frequent charging.

1 kV

3462

2.5 kV

2043

5 kV

1000

Environmental specifications

Operating temperature

-20 °C to 50 °C (-4 °F to 122 °F)

Storage temperature

-20 °C to 65 °C (-4 °F to 149 °F)

Relative humidity

80 % at 31 °C decreasing linearly
to 50 % at 50 °C

Operating altitude

0 to 2000 meters

Enclosure sealing

IP-40

Shock, vibration

MIL-PRF-28800F, CLASS II

EMI immunity

EN 61326

Safety

EN 61010-1, CAT III, 600 V,
EN61557, PARTS 1 & 2

Input overload protection

600 V ac continuous

Software specifications

FlukeView

®

Forms Basic 1550B Software

requires a PC running Windows

®

95, 98, Me,

2000, NT4.0
Two-year warranty

Included accessories

Test leads and low leakage probes
Alligator clips
Interface adapter and cable
FlukeView

®

Forms Basic 1550B PC software

Line cord
Soft carrying case with
waterproof bottom
User’s manual CD with 12 languages
Getting Started Guide

Optional accessories

FVF-UG FlukeView

®

Forms Upgrade

to Version 3.0
TL1550EXT 25 foot extension leads

Range

Accuracy

Leakage current

1 nA to 2 mA

± (5 % + 2 nA)

Capacitance

0.01 uF to 15.00 µF

± (15 % rdg + 0.03 µF)

measurement

Live circuit

30 V to 600 V

± (5 % + 2 V)

indicator

ac/dc, 50/60 Hz