Ordering information, Specifications, Test and measurement solutions for industry – Atec Mohr-CT100 Series User Manual
Page 2: General options, Standard accessories (included), Optional accessories
Test and Measurement Solutions for Industry
™
Copyright © MOHR Test and Measurement LLC 2009‐2012. All trademarks are the property of their
respective owners. Data contained herein is subject to change without notice.
For the most recent specifications, please visit http://www.mohrtm.com
SALES CONTACT:
ph: +1 (888) 852‐0408
fx: +1 (509) 946‐4395
TDR System Characteristics
Excitation Signal: Step‐rise, 300 mV into 50 Ω load
System Risetime (20‐80%, typ.): 60 ps, 100 ps (CT100HF, CT100)
Timebase Resolution: 760 fs
Timebase Random Jitter (typ.): < 1 ps rms
Timebase Non‐Linearity (typ.): < 1%
Sample Resolution: 16 bits
Sequential Sample Rate: 12.5 ‐ 250 kHz
TDR Framerate: 25 ‐ 500 waveforms/s
Horizontal System
Range: 0 ‐ 30,000 ft. (0 ‐ 9 km)
Scales: 0.003 ‐ 400 ft./div (0 ‐ 125 m/div)
Cursor Resolution: 0.003 in. (75 μm) at VoP 0.66
Accuracy (max, 0‐50°C): < 1% of measured distance, typ. < 1 mm
Velocity of Propagation (VoP)
Range: 0.250000 to 1.000000
Resolution: 0.000001
Vertical System
Range: < 1.0 Ω to > 1500.0 Ω
Resolution: ≤ 0.1 Ω, depending on scale
Accuracy (typ.): < 1% of measured value or < 1 Ω, 0 to 1000 Ω
Accuracy (max, 0‐50°C): < 10% of measured value, 0 to 1000 Ω
Measurements/Math
Measurements: Time‐to‐fault, distance‐to‐fault, Ohms‐at‐cursor,
reflection coefficient, return loss, ∆time, ∆distance, ∆Ohms,
∆reflection coefficient, relative return loss
Waveform Processing: smoothing, subtraction, 1st derivative, FFT,
S11 estimation, impedance
,
layer‐peeling
Special Features
Functions: AutoFit™, AutoScan™
Libraries: Waveform library, cable‐type library, configuration library
Data Storage
2+ GB flash memory, thousands of high‐resolution cable scans
Connectivity
Standard Features: USB host (front panel) and client (rear panel),
10/100 Mb Ethernet, optional 802.11b/g wireless networking
Special Features: Live streaming and remote control of any CT100
Series TDR over LAN/WAN/Internet using CT Viewer™
Display
Color LED‐BL 4.3 in. (10.9 cm) WQVGA TFT‐LCD, > 600 cd/m
2
Power System
AC Power: 90‐264 VAC, 50‐60 Hz using AC adapter
Battery Power: Internal 2500 mAh 14.4 VDC NiMH battery
Battery Life: >6h (typical use), unlimited with external battery packs
Battery Charging: <1 h low‐battery, <4 h fully‐discharged
Environmental and Mechanical
Operating / Non‐Operating Temp.: ‐10ºC to +55ºC / ‐20ºC to +60ºC
Dimensions: 4.3(H) x 11.5(W) x 6.9(L) in. (10.9 x 29.2 x 17.5 cm)
Weight: 5.1 lbs. (2.3 kg) with cover, 4.7 lbs. (2.2 kg) without cover
Regulatory
EMC: MIL‐PRF‐28800F. MIL‐STD‐461F RE102, CE102. IEC 61000
Shock/Vibration: MIL‐PRF‐28800F (Class 3)
Temperature/Humidity: MIL‐PRF‐28800F (Class 3)
Explosive Atm: MIL‐STD‐810G 511.5 Procedure 1 (+55ºC, 0‐4600 m)
Ordering Information
General Options
CT100, BNC test port (self‐grounding)
CT100S, SMA test port
CT100HF, SMA test port
Standard Accessories (Included)
One (1) License CT Viewer™ Software
Standard Calibration Kit
Operator's Manuals
Rugged Soft‐Sided Carrying Case
External AC Power Adapter
USB / Ethernet Cables
NIST‐Traceable Calibration / Certificate
12‐Month Standard Limited Warranty
Optional Accessories
General
Small Form‐Factor Keyboard CT100‐AC‐KBD
External Battery Pack (2700 mAh) CT100‐AC‐B270E
Hard Carrying Case CT100‐AC‐CH
Adapter Kits
SMA Adapter Kit CT100‐AK‐SMA
BNC Adapter Kit CT100‐AK‐BNC
Impedance Matching Kit CT100‐IK‐BNC
Impedance Matching Adapters
50 Ω to 75 Ω CT100‐AC‐I5075‐BNC
50 Ω to 93 Ω CT100‐AC‐I5093‐BNC
50 Ω to 125 Ω CT100‐AC‐I50125‐BNC
Specifications
Complies with all applicable EU directives, as specified
by the instrument's Declaration of Conformity.