beautypg.com

Atec OMICRON-CMC-156 User Manual

Page 2

background image

Ordering Information

CMC 156 EP with Test Universe Software

VE001616

CMC 156 EP Basic

VE001617

CMC 156 EP Protection

VE001618

CMC 156 EP Advanced Protection

VE001619

CMC 156 EP Meter

VE001621

CMC 156 EP Measurement

VE001622

CMC 156 EP Universal

1

Guaranteed values valid over one year within 23 °C ±5 °C (73 °F ±10 °F), in the

frequency range of 10 ... 100 Hz at nominal value, analog measurement inputs at

full-scale value. Specifications for three-phase systems under symmetrical conditions

(0°, 120°, 240°)

2

R

load

: 0 ... 0.5 W

3

rd. = reading, rg. = range

4

THD+N: Values at 50/60 Hz with 20 kHz bandwidth

5

R

load

: > 250 W

6

Data are valid for set value from 0.1 ... 12.5 A (current amplifier group A or B) at

50/60 Hz

Permissible load current outputs:

Range 1.25 A: 0 … 1 W and max. 1 VA, cos j = 0.5 … 1

Range 12.5 A: 0 … 0.5 W and max. 6 VA, cos j = 0.5 … 1

Permissible load voltage outputs:

max. 10 VA at 50 V … full scale voltage, cos j = 0.5 … 1

7

For the control of CMC 156 EP via the USB port of the PC, a CMUSB-P USB to

Parallel Converter is included in delivery

Binary outputs, relays
Type

potential-free relay contacts, software
controlled

Number

4

Break capacity AC

Vmax: 250 VAC / Imax: 8 A / Pmax: 2000 VA

Break capacity DC

Vmax: 300 VDC / Imax: 8 A / Pmax: 50 W

Connection

4 mm (0.16 in) banana sockets

Binary outputs, transistor
Type

open collector transistor outputs

Number

4

Update rate

10 kHz

Imax

5 mA

Connection

16 pin combination socket (rear side)

DC voltage measuring input
Measuring range

0 ... ±10 V

Accuracy

error < 0.01 % rg.

3

typ., < 0.05 % rg. guar.

Input impedance

1 MW

Connection

4 mm (0.16 in) banana sockets or measure-
ment combination socket

DC current measuring input
Measuring range

0 ... ±20 mA

Accuracy

error < 0.01 % rg.

3

typ.,

< 0.05 % rg. guar.

Input impedance

15 W

Connection

4 mm (0.16 in) banana sockets or measure-
ment combination socket

Power supply
Nominal input voltage

110 – 240 VAC, 1-phase

Permissible input voltage

99 ... 264 VAC

Nominal frequency

50/60 Hz

Permissible frequency range

47 ... 63 Hz

Power consumption

< 600 VA

Rated current

6 A

Connection

Standard AC socket (IEC 60320)

Environmental conditions
Operation temperature

0 ... +50 °C (+32 ... +122 °F)

Storage temperature

-25 ... +70 °C (-13 ... +158 °F)

Humidity range

Relative humidity 5 ... 95 %,
non-condensing

Vibration

IEC 60068-2-6 (20 m/s

2

at 10 ... 150 Hz)

Shock

IEC 60068-2-27 (15 g/11 ms half-sine)

Safety Standards, Electromagnetic Compatibility
EMC

The product adheres to the electromagnetic
compatibility (EMC) Directive 2004/108/EC
(CE conform).

Emission

Europe EN 61326-1; EN 61000-6-4; EN 61000-3-2/3

International IEC 61326-1;IEC 61000-6-4; IEC 61000-3-2/3

USA FCC Subpart B of Part 15 Class A

Immunity

Europe EN 61326-1; EN 61000-6-2;

EN 61000-4-2/3/4/5/6/11

International IEC 61326-1;IEC 61000-6-2;

IEC 61000-4-2/3/4/5/6/11

Safety

The product adheres to the low voltage
Directive 2006/95/EC (CE conform).

Europe EN 61010-1 Insulation of PC and SELV

interfaces complies with EN 60950-1

International IEC 61010-1

USA UL 61010-1

Canada CAN/CSA-C22.2 No 61010-1-04

Miscellaneous
Weight

9.8 kg (21.6 lbs)

Dimensions (W x H x D,
(without handle)

343 x 145 x 268 mm
(13.5 x 5.7 x 10.6 in)

PC connection

7

parallel port (D-Sub 25 connector)

Signal indication (LED)

> 42 V for voltage outputs

Hardware diagnostics

Self diagnostics upon each start up

Galvanic separated groups

The following groups are galvanically
separated from each other: mains, voltage/
current amplifier output, binary input

Protection

All current and voltage outputs are
fully overload and short circuit proof and
protected against external high-voltage
transient signals and over temperature