beautypg.com

Epoch-ii high-current output unit – Atec Megger-Multi-Amp-EPOCH-II User Manual

Page 2

background image

EPOCH-II with EPOCH-10 control unit

EPOCH-II

High-Current Output Unit

SPECIFICATIONS

Input

Input Voltage (specify one)

115 V ±10%, 50/60 Hz, 30 A (at full rated output)

OR

230 V ±10%, 50/60 Hz, 20 A

Output

Output Current
To provide a variety of test circuit impedances, six output taps
with two ranges are provided.

High Range
2.00 to 10.00 A at 100 V max
3.00 to 15.00 A at 66.6 V max
8.00 to 40.00 A at 25 V max
10.00 to 50.00 A at 20 V max
20.00 to 100.0 A at 10 V max
34.00 to 170.0 A at 5.9 V max
Output Power: 1000 VA

Low Range
2.00 to 10.00 A at 50 V max
3.00 to 1 5.00 A at 33.3 V max
8.00 to 40.00 A at 12.5 V max
10.00 to 50.00 A at 10 V max
20.00 to 100.0 A at 5 V max
34.00 to 170.0 A at 2.95 V max
Output Power: 500 VA

Accuracy

Typical: ±0.5% of setting
Maximum: ±1.0% of setting
Alarm will indicate when amplitude, phase angle, or waveform is
in error.

Resolution

Four ranges: 0.01 A
Two ranges: 0.l A

Duty Cycle

Five minutes at full rated VA output. Fifteen minutes recovery
time.

Overrange Capability

The EPOCH-II has an overrange capability of +10% for each tap
with a maximum output current of 187 A on the 170 A tap.

Distortion

Less than 1% typical, 3% max.
Output of EPOCH-10/EPOCH-II Combination

Frequency

Synchronized to input power source

60 Hz crystal-controlled

50 Hz crystal-controlled

Accuracy

Synchronized, tracks input frequency

±0.006 Hz for 60-Hz crystal control (±001%)

±0.005 Hz for 50-Hz crystal control (±0.01%)

Output of PULSAR and Interface Module Connected to EPOCH-II
The High-Current Interface Module provides a variable frequency
signal to the Epoch-II High-Current Output Unit.
Output frequency is continuously displayed for each channel with
large, high-intensity LEDs with the following ranges:
5.000 to 99.999 Hz
100.01 to 999.99 Hz
Frequency Accuracy: ±10 ppm at 23° C, ±2° C)

Current Phase Angle Control

Angle is adjusted on the EPOCH-10 control unit by 4-digit,
pushbutton control with large LED display of setting.
Range: 0.0 to 359.9°
Resolution: 0.1°
Accuracy: less than ±0.3° typical, ±1.0° max

Control Section

The control unit for the EPOCH-II high-current section is PULSAR
(with the High-Current Interface Module), an EPOCH-I or EPOCH-
10. Thus, the excellent operating and control features of PULSAR
and the EPOCH-10 are used to control the EPOCH-II.
Note: When the EPOCH-II is in use, the current output of the
EPOCH-10 control unit is inoperative. All EPOCH-10s can control
an EPOCH-II high-current section. When the high-current section
of the EPOCH-II is not needed, the EPOCH-10 can be used
independently or slaved together with other EPOCH units.

Protection

The input line circuit is breaker-protected. The dc power supply is
overcurrent-protected. In addition, overvoltage protection is
provided on the input line circuit.
The power amplifiers are forced-air cooled and are protected by
thermal-overload sensors.
Audio and visual alarms on the PULSAR and EPOCH-10 control
units indicate whenever the current or potential outputs are
overloaded.

Temperature

Operating
32 to 122° F (0 to 50° C)
Reduced duty cycle above 113° F (45° C)

Storage
–13 to +158° F (–25° to +70° C)

Dimensions

7.6 H x 19.75 W x 21.6 D in.
(193 H x 502 W x 549 D mm)

Weight

115 lb (52.3 kg)